J. Aerosol Sci., Vol. 24, Suppl. I, pp. $201-$202, 1993 0021-8502/93 $6.00 + 0.00

Printed in Great Britain. Pergamon Press Ltd

19 P 07

O n - L i n e M e a s u r e m e n t o f P a r t i c l e S i z e a n d C o m p o s i t i o n M. Weiss, J.C.M. Marijnissen, P.J.T. Verheijen and B. Scarlett

Delft University of Technology, Faculty of Chemical Engineering and Materials Science Particle Technology Group, P.O. Box 5054, 2600 GA Delft, The Netherlands.

A b s t r a c t

The Particle Technology Group at Delft University of Technology is developing an instrument for on-line, real-time measurement of size and chemical compostion of individual aerosol particles. Particles are sampled from an aerosol by using aerosol beam techniques. Collimated by the beam generator system the particles enter the ionization chamber of a time- of-flight mass spectrometer. There they are detected by the use of a photomultiplier when travelling through the beam of a small He-Ne laser and thereby scattering light. The photomultiplier signal is also used to trigger a Q-switched YAG laser which is vaporizing and ionizing the same particle. Analysis of the generated ions provides information on the chemical composition of the particles and the photomultiplier signal contains the particle size information.

K e y w o r d s

Chemical; instrumentation; single particle analysis; time-of-flight mass spectrometry I n t r o d u c t i o n

Aerosols continue to play an important part in numerous fields of science. For example, since substructures are still getting smaller in the semiconductor industry, even sub-micron particles can cause an electronic circuit to fail. It is therefore of the utmost importance to prevent the emission of particles and to reduce their concentration as much as possible. One method to determine the origin of a particle is to measure its size and chemical compostition. Another application for such a measurement device would be atmosperic processes. Natural and antropogenic particles can play a major role in atmosperic chemistry, acting as catalyst or reactant. To gain insight in the processes occurring in our atmosphere it would be very helpful to have an instrument capable of measuring particle size and chemical composition in real-time. Until now no commercially available instrument is developed. At the Particle Technology Group at Delft University of Technology a program whereby such a instrument is developed is in progress. The size of the particles is measured through light scattering. The chemical composition is determined with laser-induced time-of-flight mass spectrometry.

E x p e r i m e n t a l

set-up

The instrument as described by Marijnissen et al. (1988) consists of a number of components

(figure 1). By expanding an aerosol through a nozzle into a vacuum chamber an aerosol beam is formed. Because the particles have a much higher inertia than the gas molecuies they continue to travel at straight lines, while the gas is removed by the vacuum pumps. For

further details on the beam generator see Kievit et al., 1990.

Collimated by the beam system the particles enter the ionization chamber of a time-of-flight mass spectrometer. The particles are travelling through the beam of a He-Ne laser. The scattered light is measured and gives an indication of the particles size. Furthermore the scattering signal is used to trigger a second laser within 100 ns. This second laser is a Q- switched Nd:YAG laser, which generates pulses of l0 ns length containing 1200 mJ of energy (wavelength 1064 nm). By using frequency doubling crystals the output can be

(2)

converted into a multi-wavelenth pulse, containing green light (wavelength 532 nm) and UV light (wavelength 266 nm).

The described laser pulse vaporizes the particle and ionizes the fragments. An electric field accelerates the ions into the flight tube of a time-of-flight mass spectrometer. The mass spectrometer detects the arrival of ions at the end of the flight tube. The signal is recorded by a digital oscilloscope and transfered on-line into the memory of a connected computer. Because the laser can fire with a maximum rate of 5 Hz, the data of 5 mass spectra per second has to be transfered, processed and stored. In addition the signal of the photomultiplier has to be recorded and transfered to the computer simultaniously to get the desired information of the particle size.

II

b e a m g e n e r a t o r o o Q .,, l a s e r f o r s i z e / t r i g g e r l a s e r f o r p y r o l y s i ~ l i o n / s a t / o n T O F m a s s s p e c t r o m e t e r d i g i t a l s c o p e

I

P C

Figure 1 set-up in principal

Planed data processing

It is planned to use the instrument for classification of different kinds of particles. The idea is to implement a learning algorithm. Such algorithms have to be trained. In this case several spectras of for example two classes of particles have to be classified manually. Those known spectra are used as a training data set. After the training phase the algorithm should be able to classify particles into the trained classes.

Several learning algorithms will be implemented and the results will be compared.

If possible classes are not known a priori a cluster analysis shall be used to find groups/classes of particles (for examples from the same origin).

References

Kievit, O., J.C.M. Marijnissen, P.J.T. Verheijen and B. Scarlett (1990). Some

improvements of the particle beam generator. J. Aerosol Sci.21. Suppl. 1,685-688.

Kievit, O., J.C.M. Marijnissen, P.J.T. Verheijen and B. Scarlett (1992). On-line

measurement of particle size and composition. J, Aerosol Sci. 23 Suppl.: 1, $301-$3041

Marijnissen, J., B. Scarlett and P, Verheijen (1988). Proposed on-line aerosol analysis combining size determination, laser induced fragmentation and time-of-flight mass

Cytaty

Powiązane dokumenty

Since geophones are commonly used in SHM systems to monitor vertical deformations and to calculate modal properties, the first natural frequency of both beams

W przeciwieństwie do okresu wczesnej republiki w schyłkowym jej etapie obserwuje się upadek moralności, rodziny i religii, co wiązało się między innymi z przyjmowaniem

Wówczas to zbliży! się do środowiska li­ beralnych profesorów rosyjskich grupują­ cych się wokół wybitnego naukowca i poli­ tyka - K.D. Kawielina; poglądom liberal­

Przy ustalaniu czasu pracy narzędzia przyjmuje się standardowo 15 minut trwałości pracy ostrza w materiale, jest to wartość zalecana przez więk- szość producentów narzędzi i

This article, introducing the legal regulation related to the introduced time limits, aims to draw attention to interpretation doubts regarding the issue of linking the time limit

Organizatorami Sympozjum było wiele podmiotów międzynarodo- wych: Słoweńska Akademia Nauk, Uniwersytet w Ljubljanie (Wydział Teologiczny i Wydział Sztuki), Instytut

Obecnie w Bibliotece Politechniki Lubelskiej przygotowuje się opracowa- nie strategii szkoleń użytkowników oraz podnoszenia kwalifi kacji pracowników, która zakłada

Najstarszym zachowanym dokumentem z pierwszych lat istnienia Cukrowni, znajdującym się w archiwum zakładowym, jest pisany ręcznie w języku rosyj- skim „Opis główny