Kardiologia Polska 2014; 72, 5: 466; DOI: 10.5603/KP.2014.0095 ISSN 0022–9032

Studium przypadku / CliniCal Vignette

Heart perforation by pro-MRI right ventricular lead in a 26-year-old woman

Perforacja serca przez elektrodę komorową pro-MRI u 26-letniej kobiety

Małgorzata Hardzina

1

, Andrzej Ząbek

1

, Bartosz Laskowicz

2

, Anna Rydlewska

1

, Krzysztof Boczar

1

, Barbara Małecka

1, 3

1Department of Electrocardiology, John Paul II Hospital in Krakow, Poland

2Department of Radiology, John Paul II Hospital in Krakow, Poland

3Instytute of Cardiology, Jagiellonian University Medical College in Krakow, Poland

A 26-year-old patient had a VVI pacemaker implanted at the age of nine due to congenital complete atrio-ventricular block. Uni- polar passive-fixation Pacessetter Membrane lead was implanted to the right ventricle apex (Fig. 1A). For the following 17 years, the patient felt well. She underwent an elective device replacement and gave birth naturally. Recently, the patient had transient ischaemic attacks twice with symptoms of syncope accompanied by transient motor aphasia, disorder of vision and right-side hypoesthesia. To give a final neurological diagnosis, magnetic resonance imaging (MRI) of the brain was demanded. Additionally, routine pacemaker follow-up showed an episode of miopotential oversensing. For these two reasons, the patient was qualified for transvenous lead extraction (TLE) with simultaneous implantation of DDD pro-MRI stimulation system. Using a mechanical Cook system, the old ventricular lead was removed and new active-fixation, bipolar Medtronic CapSureFixMRI was initially implanted to the right ventricular outflow tract. However, due to intraoperative dislocation of the lead, it was finally placed in the periapical region of the right ventricle. Atrial

lead was implanted to the right atrial appendage.

A Medtronic Ensura DR pacemaker was placed in the previous pocket (Fig. 1B). Beginning from the day following the procedure, the patient felt recurring pricking pain in the apical region radiating to the right shoulder. These symptoms aggravated significantly over the following few weeks and were the reason for the patient’s readmission. During the device follow-up, a high pacing threshold was noted (unipolar [UP]: 4.25 V/1 ms; bipolar [BP]:

2.0 V/1 ms). Sudden increase of the threshold value was recorded in the device’s stored data (from 1.3 V/0.4 ms to 2.5 V/0.4 ms in BP configuration).

Variation of R wave sensing in BP measurements (from 5 mV to 14.5 mV) synchronised with heart cycle was also detected. Moreover, R wave sensing values were two times higher in BP than in UP mea- surements. Heart apex perforation by ventricular lead was confirmed in transthoracic echocardio- graphy (TTE) (Fig. 2) and computed tomography (CT) (Fig. 3). Another urgent TLE of the ventricular lead with cardio-surgical back-up was performed.

Subclavian vein obstruction revealed in venography before the procedure became a significant difficulty during the TLE procedure (Fig. 4). The perforating lead was removed using a yellow Byrd dilator from Cook Medical. A new lead was implanted to intra- ventricular septum through regained vessel lumen (Fig. 1C). The whole procedure was uneventful.

TTE on the third day after the procedure showed an increased amount of fluid in the pericardial space (from 4 mm to 11 mm), which decreased over the following days. The patient was discharged in good condition. The probable cause of heart perforation was the displacement of lead tip into the channel remaining after removal of the 17-year-old lead.

Address for correspondence:

Andrzej Ząbek, MD, MSc, Department of Electrocardiology, John Paul II Hospital in Krakow, ul. Prądnicka 80, 31–202 Kraków, Poland, e-mail: andrzej_j_z@poczta.onet.pl Conflict of interest: Anna Rydlewska — faculty fee for Medtronic

Figure 4. X-ray: vein obstruction with colla- teral circulation Figure 1. Chest X-ray in PA projection; A. VVI pacemaker; B. Perforating lead; C. Final pacemaker

Figure 2. TTE (arrow) the lead tip in pericardial space

Figure 3. CT (arrow) perforating lead tip;

A. Multi planar reconstruction; 1 — right heart cavity with contrast medium; 2 — pericardial sac space; B. Volume rendering technique

A B C

A B

Cytaty

Powiązane dokumenty

After five months of LV support, due to significant improvement of the patient’s cardiac efficiency including myocardial contractility with increase of LVEF +35% and

Maciej Janiszewski, MD, PhD, Division of Heart Failure and Cardiac Rehabilitation, Department of Cardiology, Hypertension, and Internal Diseases, 2 nd Medical Faculty,

Dawid Lipski, MD, Department of Hypertension, Angiology and Internal Medicine, Poznan University of Medical Sciences, ul. Origin of LCA

Computed tomography angiography of aorta showed stenosis of left subclavian artery (Fig. 1A) and severe descending aortic stenosis (Figs. 1B, C), with extensive

Patent ductus arteriosus and an accessory structure in left ventricular outflow tract in a 68-year-old woman.. Przetrwały przewód tętniczy i dodatkowa struktura w drodze odpływu

A 46-year-old Caucasian male patient with brady-tachycardia syndrome and a presentation of sinus bradycardia with oc- casional, atypical atrial flutter episodes had received the DDD

Chest computed tomography demonstrated thoracic aorta dissection, which began 15 mm above the aortic valve, expansion of ascending aorta to 44 mm, expansion of arch of aorta to 28

Transvenous extraction of a five year−old ventricular lead inadvertently implanted in the left ventricle.. Przezżylne usunięcie elektrody przypadkowo wszczepionej do lewej komory 5