Incomplete apposition of a transcatheter aortic valve replacement frame associated with late subclinical leaflet thrombosis

N/A
N/A
Protected

Academic year: 2022

Share "Incomplete apposition of a transcatheter aortic valve replacement frame associated with late subclinical leaflet thrombosis"

Copied!
2
0
0

Pełen tekst

(1)

C L I N I C A L V I G N E T T E  Transcatheter aortic valve apposition and subclinical leaflet thrombosis 1289 A 92‑year ‑old woman with symptomatic bicuspid

aortic valve stenosis (mean gradient, 80 mm Hg;

valve area, 0.7 cm2) and a left ventricle of nor‑

mal size and function was referred for trans‑

catheter aortic valve replacement (TAVR) due to the EuroSCORE II of 7.96 and Society of Tho‑

racic Surgeons score of 10.16. The aortic annu‑

lus assessed by multislice computed tomogra‑

phy (MSCT) angiography was oval and had a di‑

ameter of 21.3 × 26.4 mm, a perimeter of 73 mm, and an area of 398 mm2. There were a nonpro‑

truding calcium nodule at the annulus, severe calcifications within the base of the noncoro‑

nary cusp and its free margins, and severely

calcified raphe between the right and left cusps.

A 23‑mm Edwards SAPIEN 3 (ES3) valve was im‑

planted via the transfemoral route. Angiography showed no aortic regurgitation with the ES3 mid diameter of 21.8 mm. Echocardiography dem‑

onstrated a mean gradient of 6 mm Hg and no aortic regurgitation. Intravascular ultrasound (IVUS) was performed using a 20‑MHz Visions PV0.018” digital catheter (Philips Volcano, San Diego, California, United States) equipped with the ChromaFlo software, which differentiates blood from the vessel wall or stent struts to as‑

sess the apposition.1 The IVUS catheter was de‑

livered using a 3.5 Judkins Right guiding cathe‑

ter and slightly extended outside its tip to sweep the transducer around the entire valve frame perimeter, keeping it coaxial to the valve long axis. Although IVUS transducer location was af‑

fected by valve leaflet motion, it demonstrated a ChromaFlo signal in the space between the aor‑

tic wall and the outer ‑valve frame surface, in‑

dicating the incomplete apposition of the valve frame at 2 locations: 1) during diastole (Figure 1A)

Correspondence to:

Łukasz Kalińczuk, MD,  National institute of Cardiology,  ul. Alpejska 42, 

04-628 Warszawa, Poland,  phone: +48 22 343 43 42, email: 

lukasz.kalinczuk@gmail.com Received: August 21, 2020.

Revision accepted:

September 11, 2020.

Published online:

September 24, 2020.

Kardiol Pol. 2020; 

78 (12): 1289-1290 doi:10.33963/KP.15618 Copyright by the Author(s), 2020

C L I N I C A L V I G N E T T E

Incomplete apposition of a transcatheter aortic valve replacement frame associated with late

subclinical leaflet thrombosis

Łukasz Kalińczuk1, Kamil Zieliński1, Adam Witkowski1, Gary S. Mintz2 1  National institute of Cardiology, Warsaw, Poland

2  Cardiovascular research Foundation, New York, united States

Figure 1 Corresponding postprocedural intravascular ultrasound (A and B) and late multislice computed  tomography findings (C) in a patient after transcatheter aortic valve replacement. White arrows indicate the ChromaFlo  signal, yellow arrows—the valve frame, blue arrows—the  valve leaflets, the black arrow—calcium deposits,  asterisks—hypoattenuated leaflet thickening, and green  arrows—the sites of incomplete valve frame apposition, filled  with either hypoattenuated tissue or contrast

Abbreviations: IAS, interatrial septum; LA, left atrium; 

RVOT, right ventricular outflow tract

IAS RVOT LA

*

*

C

A B

(2)

KARDIOLOGIA POLSKA  2020; 78 (12) 1290

and 2) during systole (Figure 1B). The outer frame area measured using IVUS constituted 89.5%

of the nominal and 91.3% of the baseline annu‑

lus. Four months later, MSCT angiography re‑

vealed hypoattenuated thickening on the sur‑

faces of the left and right bioprosthesis leaflets (measuring 2.4 mm and 4.5 mm, respectively) with reduced leaflet motion (78.67% and 59.65%, respectively), features of hypoattenuated leaf‑

let thickening (Figure 1C) at the sites of extensive native valve calcification and ES3 underexpan‑

sion (oval outer frame area of 87.5% of the nom‑

inal value). Comparing IVUS and MSCT images, the sites of incomplete ES3‑frame apposition on postprocedural IVUS closely corresponded with the sites of hypoattenuated leaflet thickening at follow ‑up. Hypoattenuated leaflet thicken‑

ing following successful TAVR could be associ‑

ated with regional incomplete valve frame ap‑

position,2 a potential nidus for thrombus for‑

mation, particularly at the site of an underex‑

panded valve frame.

ArtiCle informAtion

ConfliCt of interest gSM received honoraria from Boston Scientific,  Philips, Medtronic, and Terumo. Other authors declare no conflict of interest with  regard to this manuscript.

open ACCess This is an Open Access article distributed under the terms  of  the  Creative  Commons  Attribution -NonCommercial -NoDerivatives  4.0  in- ternational License (CC BY -NC -ND 4.0), allowing third parties to download ar- ticles and share them with others, provided the original work is properly cited,  not changed in any way, distributed under the same license, and used for non- commercial purposes only. For commercial use, please contact the journal office  at kardiologiapolska@ptkardio.pl.

How to Cite Kalińczuk Ł, Zieliński K, Witkowski A, Mintz gS. incomplete ap- position of a transcatheter aortic valve replacement frame associated with late  subclinical leaflet thrombosis. Kardiol Pol. 2020; 78: 1289-1290. doi:10.33963/

KP.15618

referenCes

1 Kalińczuk Ł, Chmielak Z, Zieliński K, et al. intravascular ultrasound online guid- ance during transcatheter valve replacement for native aortic stenosis or failed bio- prosthesis. Kardiol Pol. 2020; 78: 762-765.

2 Bobrowski r, Świerczewski M, Kalińczuk Ł, et al. Late subclinical leaflet throm- bosis after transcatheter replacement of a stenotic bicuspid aortic valve. Kardiol  Pol. 2020; 78: 1169-1170.

Cytaty

Powiązane dokumenty

Macroscopic findings showing the anterior mitral leaflet (AML) was injured by mechanical stimulation from the bottom of implanted valve (arrow);D. Microscopic findings showing

It has been shown that new-generation valves reduced the rates of paravalvular leak (PVL), annular rupture and second valve implantation compared with the early-generation ones

There was no difference between the early discharge and late discharge group in terms of Valve Academic Research Consortium-2 (VARC-2) criteria out- comes, all-cause re-admission

Computed tomography (CT) before transcatheter aortic valve replacement (TAVR): CT showed calcific atheromatous changes in the aortic valve and calcification in the ostial segment

figure 1 A – multislice contrast computed tomography of implanted bioprosthesis and the ascending aorta, the short axis view; B – multislice contrast computed tomography of

tic stenosis (mean gradient, 80 mm Hg; aortic valve area, 0.7 cm 2 ) and left ventricular ejec‑A. tion fraction

However, a higher risk of THVT may be expected in patients with chronic obstructive pulmonary disease, a smaller aortic valve area, a higher mean postprocedural

In a sta- ble patient, the approach is to control the posi- tion of a stiff guidewire, verify the valve shape and re-balloon in the case of distortion, as well as increase