Implantation of a leadless pacemaker in a young adult patient with repaired tetralogy of Fallot

Marta Jagosz

1

, Wiktoria Kowalska

1

, Aleksandra Woźniak

2

, Bartosz Bichalski

1

, Zbigniew Kalarus

2, 3

, Beata Średniawa

2, 3

, Ewa Jędrzejczyk-Patej

2

1Students Scientific Society at the Department of Cardiology, Division of Medical Sciences in Zabrze, Medical University of Silesia, Katowice, Poland

2Department of Cardiology, Congenital Heart Diseases and Electrotherapy, Silesian Center for Heart Diseases, Zabrze, Poland

3Division of Medical Sciences in Zabrze, Medical University of Silesia, Katowice, Poland

Reported herein, is the case of 21-year-old patient with repaired tetralogy of Fallot within her first year of life. Due to sick sinus syndrome and II grade atrioventricular block endocardial single- chamber permanent pacemaker was inserted transvenously during her third year of life.

The patient was admitted in generally good condition aiming at pacemaker replacement. Right ventricle lead dysfunction was observed and thus, percutaneously removing and implantation of a new electrode was planned. Angio-computed tomography showed an obstruction of the superior vena cava. Pacemaker generator and lead were removed percutaneously but it was not possible to implant a new lead. The patient was managed with temporary transvenous pacing. The Heart Team qualified the patient for implantation of a leadless pacemaker (the MicraTM Transcatheter Pacing Sys- tem, Medtronic, Minneapolis, MN, USA).

Implantation of the MicraTM was performed under general anesthesia. Access via the right femoral vein was obtained with 23 Fr sheaths.

MicraTM was fixed into upper part of the inter- ventricular septum (Fig. 1). Implant parameters were optimal with 8.2 mV sensing, 650 Ohm electrode impedance and pacing threshold 0.5 V/

/0.24 ms. There were no complications. The time of fluoroscopy was 13 min, and exposition dose was 864 mGy.

The patient was discharged receiving beta- -blocker. The pacing threshold increased, reaching a maximum of 3.5 V at 0.24 ms during the first, and second month post-implantation. At 3-month follow-up pacing threshold decreased below 3.0 V/

/0.24 ms and was stabile until the 21-month follow- up. Other parameters were stable during follow- up. The stimulation percentage ranged from 1%

to 12.8%.

Conflict of interest: Beata Średniawa — con- sultant: Medtronic, Zoll, Bayer, lectures fee for: Boehringer-Ingelheim, Bayer, Pfizer; Ewa Jędrzejczyk-Patej — consultant fees from Medtron- ic, Biotronik, Abbott, Boston Scientific; Zbigniew Kalarus — company sponsored speaker’s bureau from Pfizer, Eli Lilly, Boehringer-Ingelheim, Ab- bott, Bayer; travel expenses to cardiology con- gresses from St. Jude Medical and Adamed; ad- visory committee: Boehringer-Ingelheim, Amgen, AstraZeneca.

CLINICAL CARDIOLOGY

Cardiology Journal 2020, Vol. 27, No. 5, 652–653

DOI: 10.5603/CJ.2020.0123 Copyright © 2020 Via Medica

ISSN 1897–5593

652 www.cardiologyjournal.org

IMAGE IN CARDIOVASCULAR MEDICINE

(2)

Months of follow up

Months of follow up

Months of follow up Procedure

Procedure

Pacing treshold [V/0, 24 ms]RV sensing [mV]Stimulation percentage [%]

1

1

1 0

0

0 0.5

1

2 1.0

2

4 1.5

3

6 2.0

4

8 2.5

5 6

10 3.0

7 3.5

8

12 4.0

9

14

2

2

2

3

3

3

5

5

5 21

21

21

Figure 1. Location of the MicraTM device in fluoroscopy during the implantation procedure (A, B). Electrical parameters of the device: right ventricle (RV) pacing threshold (C), RV sensing (D), stimulation percentage (E).

C

D

E A

B

www.cardiologyjournal.org 653

Marta Jagosz et al., Implantation of the leadless pacemaker in the young adult patient with repaired tetralogy of Fallot

Cytaty

Powiązane dokumenty

We conclude that successful stent implantation into the RVOT in patients with ToF and hypoplastic pulmonary arteries improves their clinical condition, increases pulmonary blood flow

On April 28, the implantation of a Micra TM VR Transcatheter Pacing System (Medtronic, Dublin, Ireland) was carried out in the Electrophysiological Laboratory under

In our study, we used only the Sorin Crown bioprosthetic aortic valve and the postoperative peak and mean transpulmo‑. nary gradients

nous lead extraction of the abandoned lead with surgical treatment of the fistula in the axillary fossa, leaving the pacing system on the right side of the chest.. First,

tricular septal defect, significant subpulmonary stenosis with suprasystemic RV pressure, right Patients with tetralogy of Fallot (TOF) have.. impaired pulmonary flow due

Left BCV occlusion is a rare condition, detected during diagnostic or therapeutic procedures involving cardiovascular catheterisation or lead insertion via veins located in the

RVOT aneurysm/akinesia was similar irrespective of ventricular systolic functions); (2) in patients with late correction and normal systolic functions, higher median peak

Etiologia choroby niedokrwiennej serca w tym przypadku może być złożona: oprócz klasycznych czynników ryzyka, udział może mieć przewlekła hipoksja wynikająca z