www.kardiologiapolska.pl

Kardiologia Polska 2018; 76, 7: 1115; DOI: 10.5603/KP.2018.0138 ISSN 0022–9032

CLINICAL VIGNETTE

Address for correspondence:

Maciej Wójcik, MD, PhD, Department of Cardiology, Medical University of Lublin, SPSK No. 4, ul. Jaczewskiego 8, 20–954 Lublin, Poland, tel/fax: +48 81 72 44 151, e-mail: m.wojcik@umlub.pl

Conflict of interest: Przemysław Zając is an Abbott employee.

Kardiologia Polska Copyright © Polish Cardiac Society 2018

High-density bipolar voltage mapping for substrate-guided ablation of atrial fibrillation

Maciej Wójcik

1

, Łukasz Konarski

1

, Robert Błaszczyk

1

, Piotr Aljabali

1

, Przemysław Zając

2

1Department of Cardiology, Medical University of Lublin, Lublin, Poland

2Abbott Poland

A 64-year-old diabetic man (body mass index 28 kg/m2; enlarged left atrial diameter of 45 mm and left ventricular ejection fraction of 67% on echocardiography) with an eight-year history of symptomatic (score 2b according to the modified European Heart Rhythm Association symptom classification for atrial fibrillation [AF]), drug-resistant, long-standing persistent AF was referred for elective AF ablation. After performing transseptal puncture, sinus rhythm was restored with direct current cardioversion, and high-density three-dimensional mapping of the left atrium was performed with a circular diagnostic catheter (AdvisorTM FL Sensor EnabledTM, Abbott, St. Paul, MN, USA). We collected 8571 points, and 2207 points were used (EnSite PrecisionTM, Abbott; best-duplicate algorithm marked by small yellow dots, Fig. 1) for a final bipolar voltage map, which revealed low-voltage areas (LVAs; defined as 0.2–0.5 mV) within the left atrium. The diagnostic catheter was exchanged with the ablation catheter (TactiCathTM, Abbott), and AF was reinduced with burst pacing. We used our standard radio-frequency energy settings (45oC, 30 W, and 25 W on the posterior wall only). Following our standard stepwise ablation protocol at that time (pulmonary veins isolation [PVI] > roof line > mitral isthmus line) we could not terminate AF after completing all procedural steps. After reconfirming LVA on the anterior wall with a contact-force catheter (Fig. 2) we ablated there, reaching a successful endpoint with sinus rhythm restoration (Fig. 3). The next day the patient was discharged home (on a b-blocker and non-vitamin K antagonist oral anticoagulant only) with sinus rhythm documented in electrocardiogram (ECG). He has been followed-up for 18 months and there have been no symptoms of AF or recurrence of left atrial arrhythmia since the procedure. No arrhythmia was documented in seven-day Holter ECG monitoring (at the 3rd, 6th, 9th, and 12th month) or in ECG recordings presented by the patient at follow-up visits. Previous reports showed that LVA substrate modification in addition to PVI was more effective than PVI alone or PVI plus additional empirical lesion abla- tion in patients with non-paroxysmal AF. Moreover, LVA ablation has lower pro-arrhythmic potential (demonstrated by reduced occurrence of post-ablation atrial tachycardia) as compared to PVI plus conventional wide empirical ablation. Nevertheless, the detailed identification of the substrate, i.e. ultra-high-density substrate mapping, is crucial in LVA ablation.

Figure 1. High-density bipolar voltage mapping of the left atrium:

anterior (upper panel) and posterior (lower panel) view; LAA — left atrium appendage; LIPV — left inferior pulmonary vein; LSPV — left superior pulmonary vein;

RIPV — right inferior pulmonary vein; RSPV — right superior pulmonary vein

Figure 3. Restoration of sinus rhythm during ablation of low-voltage area; abbreviations — see Figure 1

Figure 2. The confirmation of a low-voltage area (the anterior wall of the left atrium) with contact force ablation catheter (stable catheter — tissue contact of 13 g); low- -voltage areas (0.2–0.5 mV) marked in yellow/red/green/blue, healthy tissue (> 0.5 mV) in purple; abbreviations — see Figure 1

Cytaty

Powiązane dokumenty

6‑8 However, about two ‑thirds of the analyzed procedures were performed for paroxysmal AF, while two ‑thirds of patients had normal or Catheter ablation (CA) is

In the group of patients undergoing AI -based PVI, the procedure was slightly longer, but the fluoroscopy time and the radiofrequency ablation time were shorter.. Furthermore, in

Background: The aim of this study was to assess the frequency of all-cause rehospitaliza- tion and due to atrial fibrillation/flutter (AF/AFl), repeat ablation of AF/AFl,

Circumferential pul- monary vein ablation with additional linear ablation results in an increased incidence of left atrial flutter compared with segmental pulmonary vein isolation

Methods and Results: One hundred and ten patients were treated with pulmonary vein isolation with 4 different technologies: open-irrigated tip radiofrequency (RF) catheter

Pul- monary vein isolation (PVI)-focused ablation, which eliminates triggers, and partially elimi- nates substrate located near LA-PVs junction, results in high success rate in

There are no randomized prospective trials comparing the safety and efficacy of catheter abla- tion for paroxysmal or persistent AF in the elderly to best medical therapy or

The complication was observed in a female patient who had been using oral contraceptives for many years and consisted in sud- den dyspnoea which developed on the first night af-