• Nie Znaleziono Wyników

A successful transcatheter aortic valve implantation of a balloon-expandable valve for paravalvular leak in a patient with bicuspid aortic valve and horizontal aorta

N/A
N/A
Protected

Academic year: 2022

Share "A successful transcatheter aortic valve implantation of a balloon-expandable valve for paravalvular leak in a patient with bicuspid aortic valve and horizontal aorta"

Copied!
2
0
0

Pełen tekst

(1)

C L I N I C A L V I G N E T T E TAVI in the horizontal aorta 1187 United States) was introduced via the right femoral artery with a 16F integrated sheath.

As aortography showed significant paraval‑

vular leakage (PVL) after valve implantation, prosthesis was optimized with a 26‑mm bal‑

loon during rapid right ventricular pacing (160/

min). Periprocedural TTE confirmed proper but relatively low valve position with the maxi‑

mum gradient of 16 mm Hg and moderate PVL (Figure 1B and 1C).

The patient was transferred to intensive care unit where pulmonary edema occurred on the second day after the procedure. On TTE, PVL was assessed as hemodynamically signif‑

icant and the patient was referred for valvu‑

loplasty with a 30‑mm balloon. Despite that, the residual leak did not decrease. On the 14th day following Evolut R implantation, a second prosthesis, 29‑mm Sapien 3 (Edwards Life‑

sciences, Irvine, California, United States), was implanted using a 16F sheath via the left femo‑

ral access (Figure 1D). Final aortography and TTE confirmed proper position of the prosthesis with trace PVL and the maximum aortic gra‑

dient of 10 mm Hg.

Further recovery in the intensive care unit was uneventful. Prior to discharge, TTE showed trace PVL and maximum aortic gradient of 10 mm Hg, and the patient was discharged home with significant clinical improvement.

Certain anatomical features, such as a hor‑

izontal aorta (in most studies defined as pres‑

ence of >48° angle between the aortic annulus Transcatheter aortic valve implantation (TAVI)

is a well ‑established modern treatment of severe symptomatic aortic stenosis. Some anatomical features, such as bicuspid aortic valve or severe aortic angulation, often referred to as the “hor‑

izontal aorta,” increase the rates of unsuccess‑

ful procedures.1‑4

We present the case of an 83‑year ‑old man with a history of severe symptomatic aortic stenosis, type 2 diabetes mellitus on insulin, and chronic kidney disease who was referred to our institution due to deteriorating exer‑

cise tolerance (New York Heart Association class III) and chest pain. Transthoracic echo‑

cardiography (TTE) confirmed severely calci‑

fied aortic valve with the valve area of 0.9 cm2 and mean gradient of 43 mm Hg. The patient was disqualified from surgical valve replace‑

ment and referred for TAVI due to high surgi‑

cal risk (EuroScore II, 9.33%).

Multislice computed tomography revealed femoral arteries suitable for transfemoral ac‑

cess, the aortic annulus with a maximum diam‑

eter of 31 mm and minimum diameter of 25 mm, perimeter of 88 mm, and area of 5.6 cm2, func‑

tionally bicuspid aortic valve, and a horizontal course of the ascending aorta (angle between the horizontal plane and the aortic annulus of 65°; Figure 1A).

Transcatheter aortic valve implantation was performed under local anesthesia with con‑

scious sedation. The Evolut R 34 mm pros‑

thesis (Medtronic, Minneapolis, Minnesota,

Correspondence to:

Anna Pyłko, MD,  Department of interventional  Cardiology and Angiology,  institute of Cardiology,  ul. Alpejska 42, 04‑628, Warszawa,  Poland, phone: +48 22 343 41 27,  email: aniapylko@gmail.com Received: June 16, 2020.

Revision accepted: August 3, 2020.

Published online: August 5, 2020.

Kardiol Pol. 2020; 78 (11): 1187‑1188 doi:10.33963/KP.15544 Copyright by the Author(s), 2020

C L I N I C A L V I G N E T T E

A successful transcatheter aortic valve

implantation of a balloon ‑expandable valve for paravalvular leak in a patient with bicuspid aortic valve and horizontal aorta

Anna Pyłko1,2, Maciej Dąbrowski1, Zbigniew Chmielak1, Krzysztof Kukuła1, Adam Witkowski1 1  Department of interventional Cardiology and Angiology, institute of Cardiology, Warsaw, Poland

2  Department of Medicine, Medical university of Warsaw, Warsaw, Poland

(2)

KARDIOLOGIA POLSKA 2020; 78 (11) 1188

How to cite Pyłko A, Dąbrowski M, Chmielak Z, et al. A successful transcath‑

eter aortic valve implantation of a balloon ‑expandable valve for paravalvular leak  in a patient with bicuspid aortic valve and horizontal aorta. Kardiol Pol. 2020; 78: 

1187‑1188. doi:10.33963/KP.15544

RefeRences

1  Stefano DD, Colombo A, Mangieri A, et al. impact of horizontal aorta on pro‑

cedural and clinical outcomes in second ‑generation transcatheter aortic valve im‑

plantation. eurointervention. 2019; 15: e749‑e756.

2  Chan PH, Alegria ‑Barrero e, Mario CD. Difficulties with horizontal aortic root  in transcatheter aortic valve implantation. Cather Cardiovasc interv. 2013; 81: 

630‑635.

3  Bauer T, Linke A, Sievert H, et al. Comparison of the effectiveness of transcath‑

eter aortic valve implantation in patients with stenotic bicuspid versus tricuspid  aortic valves (from the german TAVi registry). Am J Cardiol. 2014; 113: 518‑521.

4  Wijesinghe N, Ye J, rodés ‑Cabau J, et al. Transcatheter aortic valve implanta‑

tion in patients with bicuspid aortic valve stenosis. JACC Cardiovasc interv. 2010; 

3: 1122‑1125.

plane and the horizontal plane in the coro‑

nal view) or bicuspid aortic valve, increase the risk of periprocedural complications and pose a particular technical challenge for prop‑

er valve sizing and positioning.1‑4 Additional‑

ly, presence of bicuspid aortic valve may re‑

sult in a decrease of postdilatation efficacy due to heavy calcifications and noncircularity of such a valve.4

It remains an open question whether or not this patient would have benefited from implan‑

tation of a balloon ‑expandable prosthesis as the first choice. In this particular case, a bail‑

‑out procedure of valve ‑in ‑valve implantation of a balloon ‑expandable prosthesis resolved the problem of PVL and caused significant clin‑

ical improvement.

ARticle infoRmAtion

conflict of inteRest MD is a member of Medical Advisory Board of  Medtronic. AW is a proctor of Medtronic. Other authors declare no conflict of  interest.

open Access This is an Open Access article distributed under the terms  of  the  Creative  Commons  Attribution ‑Non  Commercial ‑No  Derivatives  4.0  in‑

ternational License (CC BY ‑NC ‑ND 4.0), allowing third parties to download ar‑

ticles and share them with others, provided the original work is properly cited,  not changed in any way, distributed under the same license, and used for non‑

commercial purposes only. For commercial use, please contact the journal office  at kardiologiapolska@ptkardio.pl.

Figure 1 A – preprocedural computed tomography showing a 65º angle between the horizontal plane and aortic annulus; B – postdilatation of initially implanted self­

­expandable prosthesis (Evolut R, 34 mm) with a 26­mm balloon; c – transthoracic echocardiography at day 1 after the first procedure showing moderate perivalvular leak (arrow); 

D – the second balloon ­expandable prosthesis (Sapien 3, 29 mm) implanted on day 14 after the initial procedure

D c

B A

Cytaty

Powiązane dokumenty

This echocardiographic evaluation performed in a real-world setting in European patients with severe AS who received a transcatheter SAPIEN 3 demonstrated good valve performance,

1) Sustained hemodynamic changes are observed as peak velocity and transvalvular gradient dif- ferences between discharge and follow-up within 1 year after the procedure; 2) BAV

As a result it was noted that TAVI with a CoreValve (Medtronic, Inc, Minneapolis, MN, USA) was technically possible in high-risk patients with mechanical mitral valves or

Transfemoral transcatheter aortic valve implantation in a patient with a severe aortic stenosis and cardiogenic shock requiring intra-aortic balloon pump support.. Piotr Chodór 1

tion in mean (SD) transvalvular gradients after the placement of transcatheter implanted valve in surgical implanted valve of 19.0 (8.8) mm Hg (P <0.001) and a further mean

short ‑axis view; C – echocardiography imaging of implanted valvular prostheses after severe paravalvular regurgitation: long ‑axis view; D – removed infected valvular

In a sta- ble patient, the approach is to control the posi- tion of a stiff guidewire, verify the valve shape and re-balloon in the case of distortion, as well as increase

Prosthetic valve endocarditis and acute heart failure in a patient after transcatheter aortic valve implantation procedure.. Tomasz Skowerski, Patryk Grzywocz, Mariusz Bałys,