Advances in Interventional Cardiology 2016; 12, 2 (44)

Images in intervention

Corresponding author:

Emin Evren Ozcan MD, Cardiology Department, Sifa University, Fevzipaşa Bulvarı No: 172/2, 35240 Basmane İzmir, Turkey, phone: +90 2324460880, e-mail: eminevrenozcan@gmail.com

Received: 11.05.2015, accepted: 30.05.2015.

Percutaneous right atrial pacemaker lead repositioning using a regular deflectable ablation catheter

Erdem Ozel1, István Osztheimer2, Ali Ozturk3, László Gellér2,Emin Evren Ozcan3

1Cardiology Department, Tepecik Training and Research Hospital, İzmir, Turkey

2Heart Center, Semmelweis University, Budapest, Hungary

3Cardiology Department, Sifa University, İzmir, Turkey

Adv Interv Cardiol 2016; 12, 2 (44): 183–184 DOI: 10.5114/aic.2016.59374

Lead dislocation is one of the most undesired com- plications of permanent pacemaker implantation. The rate of dislodgement of atrial pacing leads is nearly 3%

[1]. Reopening the pacemaker pocket and repositioning the lead is the most commonly used technique for cor- recting dislocations which occur in the early period after implantation. However, this increases the risk of pocket infection and increases lead insulation problems. Snare systems and urological baskets have also been used for repositioning atrial leads [2, 3]. We report a patient who experienced dislodgement of the atrial pacemaker lead of a cardiac resynchronization therapy (CRT) device, which was corrected using a regular deflectable ablation catheter by the transfemoral route.

A  77-year-old female patient with severe ischemic left ventricular dysfunction and wide QRS had received a  CRT device. At the follow-up 1 week after the proce- dure, atrial sensing and pacing dysfunctions were de- tected. Fluoroscopy revealed dislodgement of a passive

fixation J-shaped atrial lead. Lead repositioning via a per- cutaneous transfemoral approach was planned to avoid surgical revision. A regular deflectable ablation catheter (steerable 4-mm-tip ablation catheter, Mariner, Medtron- ic, Minneapolis, USA) was advanced to the right atrium through a 7 Fr introducer from the right femoral vein. We propagated the ablation catheter above the level of the atrial lead and made an anteflexion to the distal portion of the catheter. We captured the distal loop of the atri- al lead, then with a slight counterclockwise rotation and pull back we oriented the atrial lead towards the right atrial appendage (Figure 1). The lead functions were nor- mal after the intervention. The patient was discharged the next morning and had normal pacemaker functions at follow-up.

Early lead displacements after pacemaker implanta- tions are the most frequent cause of reintervention in the majority of patients. We performed successful atrial lead repositioning with a deflectable ablation catheter by

Figure 1. Successful atrial lead reposition by deflectable ablation catheter

(2)

Erdem Ozel et al. Percutaneous lead repositioning using a deflectable ablation catheter

184 Advances in Interventional Cardiology 2016; 12, 2 (44)

the percutaneous approach without the need for a surgi- cal operation. However, this technique has a number of contraindications. Percutaneous lead repositioning can only be performed on patients who have passive fixation electrodes and leads with suitable loop anatomy. Con- ditions which necessitate the removal of the dislocated lead such as serious insulation problems are another contraindication for percutaneous lead repositioning. Re- positioning of active fixation leads should be performed surgically. Since our case had a  passive fixation atrial lead and a suitable loop anatomy, it was possible to per- form the procedure safely. However, operators need to be cautious about dislodgement of other stable leads while performing the procedure, especially in patients with an implanted CRT. Favale et al. reported successful percutaneous atrial lead repositioning in 2 patients with permanent pacemakers and in 1 patient with an implant- able cardioverter defibrillator [4]. We believe that our paper is the first one to report percutaneous atrial lead repositioning with a regular deflection ablation catheter in a patient with a CRT device.

Although it needs some experience, percutaneous transfemoral lead repositioning is a  much less invasive technique and may be preferred as a first-step approach if the patient has acceptable loop anatomy of the atrial lead.

Conflict of interest

The authors declare no conflict of interest.

References

1. Kiviniemi MS, Pirnes MA, Eranen HJK, et al. Complications relat- ed to permanent pacemaker therapy. Pacing Clin Electrophysiol 1999; 22: 711-20.

2. Morris DC, Scott IR, Jamieson WR. Pacemaker electrode reposi- tioning using the loop-snare technique. Pacing Clin Electrophysi- ol 1989; 12: 996-9.

3. Sharif AY, Davoodi G, Saeed AK, et al. New technique: reposi- tioning of dislodged atrial pacing lead with a specially designed urological basket. Europace 2007; 9: 105-7.

4. Favale S, Nacci F. Percutaneous transcatheter repositioning of displaced permanent pacemaker lead. Pacing Clin Electrophysiol 1999; 22: 1817-9.

Cytaty

Powiązane dokumenty

Paradoxical atrial undersensing was also re- ported in a single case of paroxysmal atrial flutter where device interrogation revealed atrial under- sensing of 5 mV flutter waves at

During im- plantation, there is potential for the ventricular lead to be inadvertently inserted into LV, either through a patent forman ovale, ventricular septal defect, or

Basal characteristics in patients with chronic atrial fibrillation (AF) and chronic AF with atrioventricular node ablation (AVNA). There were no significant changes of medical

We then reprogrammed the generator to a VVI stimulation mode for the last set of RF pulses in the lateral hepatic lesions and at a 50 bpm rate to limit the R on T phenomenon (Fig..

Hazard ratio estimates stratified according to cause of death between atrial fibrillation (AF) pa- tients who underwent atrioventricular junction ablation (AVJ-abl) and patients

The lead was extracted via the left femoral vein in two stages: untying a loop on the lead using a pig-tail catheter and Dotter basket followed by traction and dissection of

6‑8 However, about two ‑thirds of the analyzed procedures were performed for paroxysmal AF, while two ‑thirds of patients had normal or Catheter ablation (CA) is

nous lead extraction of the abandoned lead with surgical treatment of the fistula in the axillary fossa, leaving the pacing system on the right side of the chest.. First,