IMAGES IN CARDIOLOGY

Cardiology Journal 2011, Vol. 18, No. 5, pp. 570–572 10.5603/CJ.2011.0017 Copyright © 2011 Via Medica ISSN 1897–5593

570 www.cardiologyjournal.org

Address for correspondence: S. Serge Barold, MD, Florida Heart Rhythm Institute, Tampa, Florida, USA, e-mail: ssbarold@aol.com Received: 28.04.2011 Accepted: 10.05.2011

Committed function for the first delivered shock of an uncommitted implantable cardioverter-defibrillator

Roland X. Stroobandt1, S. Serge Barold2

1Heart Center, Department of Electrophysiology, University Hospital, Ghent, Belgium

2Florida Heart Rhythm Institute, Tampa, Florida, USA

A 75 year-old man with a Biotronik Lumax VR-T540 uncommitted implantable cardioverter- -defibrillator (ICD) received a shock during general anesthesia for the resection of a basal cell carcino- ma on the nose. The surgeon who was unaware of the ICD, used unipolar electrocautery during the procedure. No ICD parameters were reprogrammed preoperatively and the device remained pro- grammed as follows: Low rate = 40 ppm, ventricu- lar tachycardia — 1, rate = 167–222 bpm, antitachy- cardia pacing (ATP) followed by 40 J shocks, ven- tricular fibrillation (VF) zone 12/16 and > 222 bpm, ATP, followed by 40 J shocks. ATP in this device is only activated by tachycardia with relatively stable cycle lengths which was not the case in our patient

because the interference from electrocautery pro- duced sensed signals with marked irregularity of rate or detected intervals. Figure 1 shows how in- terference from electrocautery was detected as VF by the device which then initiated capacitor charg- ing. ATP was not delivered. The capacitor charge was then aborted when the interference abated based on the uncommitted function of the ICD (Fig. 2). The subsequent recurrence of another bout of interfer- ence failed to produce enough “sinus” intervals between the aborted shock and VF redetection (in- terference) to fulfill detection of a normal termina- tion (Fig. 2). Instead the ICD redetected or recon- firmed the presence of a continuing tachyarrhyth- mia. The ICD again began to charge its capacitor

Figure 1. The ICD senses interference from electrocautery. The markers on top show that the ICD interprets the signals as ventricular fibrillation (VF) according to 12 out of 16 VF intervals. The cycles labeled VT in the marker channel (M) are not counted for VF diagnosis. The interference subsides as the capacitor is charging. The underlying rhythm is atrial fibrillation; V — near-field ventricular electrogram; FF — far-field ventricular electrogram; VT — ventricular tachycardia; VS — ventricular sensed event outside the tachyarrhythmia zones; (see text for details).

Onset VT1 Detection VF

Capacitor charge progress bar

(2)

571 Roland X. Stroobandt, S. Serge Barold, Committed function for the first delivered shock of an uncommitted ICD

www.cardiologyjournal.org

(Fig. 2). The interference terminated again during capacitor charging (Fig. 3) but this time the shock was not aborted and the ICD eventually fired (com- mitted ICD function) during a relatively stable su- praventricular rhythm (Fig. 3). Subsequent ICD evaluation revealed normal shock and pacing imped- ances and no evidence of lead malfunction.

As far as the shocks are concerned, most ICDs function in an uncommitted mode for the first shock but in the committed mode for all subsequent shocks if the reconfirmation phase detects a con-

tinuing tachyarrhythmia. Yet, in our case, it was the first delivered shock that was committed.

If an uncommitted device aborts tachyarrhyth- mia therapy, it reverts immediately to the pro- grammed parameters. It then resumes monitoring for ventricular tachyarrhythmias or their equivalent usually starting at the first paced or sensed ventri- cular event after the capacitor has stopped charging.

If the device redetects the tachyarrhythmia before the ICD is able to declare termination of the epi- sode (criteria according to design), it will then de- Shock abort RD VF (12 in 16)

1 2 3 4 5 x x x6 7 8 9 10 11 x12

Capacitor charge progress bar

Figure 3. Although the interference has stopped, the ICD charges its capacitor and the full 40 J charge is delivered in a committed fashion during a supraventricular rhythm (atrial fibrillation). The finding of 3/4 “sinus” intervals during capacitor charging did not abort the shock. The shock impedance was 46 Ohms. The termination detection was fulfilled 24 s after the initial detection. Same arrangement as in Figures 1 and 2 (see text for details).

Figure 2. The shock is aborted indicated by the capacitor charge progress bar on the top left side (vertical arrow).

This occurred because 3 of 4 intervals were classified as “sinus” intervals during charging. Repeated electrocautery occurs soon after the capacitor has aborted its charge. Tachyarrhythmia termination requires 12 out of 16 “sinus”

intervals. As there are only four “sinus” intervals, there are not enough “sinus” cycles to fulfill a termination detection. The ICD re-detects VF (RD VF 12/16) and again charges its capacitor. The interference stops during capacitor charging. Same arrangement as in Figure 1 (see text for details).

Capacitor charge progress bar

40 J 46 Ohm M

M

(3)

572

Cardiology Journal 2011, Vol. 18, No. 5

www.cardiologyjournal.org

liver the programmed shock that was previously aborted [1]. This time, the ICD functioned as a com- mitted device (for the first delivered shock to the patient) so that spontaneous termination of the de- tected tachyarrhythmia (or its equivalent) as in our case no longer aborted the shock. The rationale for this behavior is based on the assumption that the first shock might have been aborted because of in- termittent undersensing of VF. In fact, the ICD sees two episodes of VF in this situation. Consequently in virtually all ICDs (except the recently released Medtronic Protecta ICD + cardiac resynchroniza- tion) [2] capacitor discharge cannot be aborted twice in succession when reconfirmation (after the capa- citor has stopped charging) suggests an ongoing tachyarrhythmia and the device fails to detect the return to “sinus rhythm” according to criteria based on its design. The committed function for the first

delivered shock after an aborted one therefore pro- vides a therapeutic safeguard.

Disclosures

The authors do not report any conflict of inte- rest regarding this work.

References

1. Mansour F, Guash E, Berruezo A, Brugada J, Tolosana JM, Mont L.

Inappropriate shocks or inappropriate programming? A review of Guidant’s reconfirmation algorithm. Europace, 2009; 11:

1120–1122.

2. Technical Manual, Medtronic Inc., Protecta CRT-D D364TRG Digital Implantable Cardiovertetr-Defibrillator with Cardiac Re- synchronization therapy (DDE-DDDR). Medtronic Inc., Minnea- polis MN 55432, USA, 2010.

Cytaty

Powiązane dokumenty

This alternative to transvenous implantable cardioverter-defibrillator (TV-ICD) is superior for patients with difficult vascular access, high risk of infection, and expected

AIMS The aim of this study was to adapt and test the reliability of the Polish version of the Implantable Cardioverter ‑Defibrillator Concerns (ICDC) questionnaire, an

1,2 The procedure was performed under regional anesthesia of the hemithorax, with combined ultrasound ‑guided serratus anterior plane block (SAPB), 3 pectoral nerve block

A 65-year-old woman with highly symptomatic (New York Heart Association [NYHA] class III/IV), non-ischaemic cardiomyopathy with severe systolic left ventricular dysfunction

Therefore, a temporary external implantable cardioverter-defibrillator (ICD) with dual-coil lead and a conventional ICD was implanted on the opposite side of the planned

A 69-year-old patient with moderate ischaemic left ventricular systolic dysfunction (ejection fraction of 45%) and a his- tory of recurrent ventricular tachycardia (VT)

The broken lead fragment was successfully removed using a bioptome (Jawz™ Endomyocardial Biopsy For- ceps, Argon Medical Devices, USA) which was inserted through the transseptal

The overall beneficial effects of primary ICD implantation in the older age groups as shown in this meta-analysis and review of literature, however, should be interpreted with