Dislodgement of a magnesium bioresorbable scaffold from the delivery system in the left main: A true “nightmaris”

N/A
N/A
Protected

Academic year: 2022

Share "Dislodgement of a magnesium bioresorbable scaffold from the delivery system in the left main: A true “nightmaris”"

Copied!
2
0
0

Pełen tekst

(1)

Address for correspondence: Prof. Juan Luis Gutiérrez-Chico, MD, PhD, FESC, FACC, Head of Interventional Cardiology, Klinikum Frankfurt (Oder), Müllroser Chaussée 7, 15236 – Frankfurt (Oder), Germany, tel: +49 (0) 176 30585019, +34 615 319370, e-mail: juanluis.gutierrezchico@ictra.es

Received: 10.05.2017 Accepted: 13.07.2017

*Equally contributed

Dislodgement of a magnesium bioresorbable scaffold from the delivery system in the left main:

A true “nightmaris”

Carlos Cortés

1, 2*

, Milosz Jaguszewski

1, 3, 4*

, Michele Schincariol

3

, María-Alexandra Durán-Cortés

1

, Ula Limon

1

, Juan Luis Gutiérrez-Chico

1, 3

1Interventional Cardiology, Klinikum Frankfurt (Oder), Germany

2Interventional Cardiology Department, Hospital Clínico Universitario de Valladolid, ICICOR, Valladolid, Spain

3Institute of Cardiovascular Translational Research of Atlantic (ICTRA), Berlin, Germany

4First Department of Cardiology, Medical University of Gdansk, Poland

This paper was guest edited by Prof. Sang-Wook Kim

A 69-year-old male patient with hypertension and hypercholesterolemia underwent percutane- ous coronary intervention (PCI) of the left anterior descending artery (LAD), from mid to ostial, with implantation of three overlapping everolimus- eluting stents (EES). PCI of a lesion in the distal left circumflex (LCx) (Supplementary Video 1

— see journal website) was scheduled 6 months later. Then an optical coherence tomography (OCT) pullback pre-PCI showed a predominantly fibrotic plaque of 19 mm length in the distal LCx. Predilation 1:1 with a scoring-balloon (AngiosculptTM 3.5 × 15 mm, Biotronik, Berlin, Germany) was performed.

A magnesium bioresorbable scaffold (MagmarisTM 3.5 × 20 mm, Biotronik, Berlin, Germany) was then advanced towards the lesion, but it got unexpectedly stuck at the distal left main (LM), unable to move forward. The attempts to retrieve the device back into the guiding catheter resulted in dislodgement from the delivery system. OCT showed an angio- graphically inadvertent protrusion of the previously implanted EES into the LM (Fig. 1, red arrowheads), in which’s mess the Magmaris was entangled (Fig. 1, white arrowheads; Supplementary Video 2 — see journal website). Some struts were visible in the guiding-catheter (Fig. 1F). Three-dimensional OCT image fusion depicted clearly the pattern of the crimped Magmaris (Fig. 2.A, B), even though the automatic detection software failed to recognise the struts. The scaffold retrieval by means of a snare

(Amplatz Gooseneck microsnareTM, 4 mm, Covidien, Plymouth, MN) was performed under stationary OCT guidance, due to its complete radiotranspar- ency. The device was firmly snared, but it broke twice in small fragments when the operator tried to pull it back (Fig. 2C, arrowheads). In a third attempt, the scaffold was successfully snared in a slightly more distal position. In this occasion the device did not break while pulling back, so it could be finally retrieved (Fig. 2C).

This case highlights the instrumental role of in- vasive imaging in critical problem solving, becoming even more prominent in radiotransparent scaffolds [1]. It also points out the importance of ensuring a clear way from the guiding catheter up to the lesion:

neither the previous OCT, focused on assessing the lesion, nor predilation manoeuvres noticed the pro- trusion into the LM of a previously implanted EES.

A careful assessment of the whole navigation pathway with invasive imaging might be advisable if PCIs have been previously performed in adjacent segments.

Conflict of interest: None declared Reference

1. Jaguszewski MJ, Cortés C, Gutiérrez-Chico JL. Implantation of magnesium-bioresorbable scaffolds in a bifurcation under opti- cal coherence tomography guidance. Eur Heart J. 2017; 38(25):

2017–2018, doi: 10.1093/eurheartj/ehw539, indexed in Pub- med: 27941012.

INTERVENTIONAL CARDIOLOGY

Cardiology Journal 2017, Vol. 24, No. 6, 700–701

DOI: 10.5603/CJ.2017.0144 Copyright © 2017 Via Medica

ISSN 1897–5593

700 www.cardiologyjournal.org

IMAGE IN CARDIOVASCULAR MEDICINE

(2)

Figure 2. Optical coherence tomography image fusion, rendering a three-dimensional depiction of the left main (A, B).

The platform pattern of the crimped Magmaris bioresorbable scaffold (MgBRS) can be clearly recognised, with its proximal struts overhanging the guiding-catheter (A, B). The MgBRS could be removed from the left main by means of a snare (C). The snared scaffold broke twice in small pieces (C, arrowheads) while pulled back, before being suc- cessfully retrieved.

Figure 1A–F. Optical coherence tomography (OCT) after dislodgement of the magnesium bioresorbable scaffold (MgBRS) from its delivery system in the left main (LM). The crimped MgBRS (white arrowheads) is clearly visible, with a guidewire inside the scaffold lumen. A second guidewire outside the scaffold lumen supports the OCT catheter.

A stent previously implanted in the left anterior descending artery is amply protruding into the LM (red arrowheads), so the MgBRS got entangled in its mesh (A–C). Struts of the MgBRS can be seen in the guiding catheter (F).

www.cardiologyjournal.org 701

Carlos Cortés et al., Dislodgement of a magnesium bioresorbable scaffold in LM

Cytaty

Powiązane dokumenty

tomography (OCT) at 6-month ( Figure 1A ; Supple- mentary material, Figure S3) and 2-year follow- -up ( Figure 1B ; Supplementary material, Figures S4 and S5) demonstrated

The presented single-centre study is an analysis of 91 patients with acute myocardial infarction (MI) treated with primary percutaneous coronary intervention (PCI) with Absorb BVS

Incidence and short-term clinical outcomes of small side branch occlusion af- ter implantation of an everolimus-eluting bioresorbable vascular scaffold: an interim report of

Absorb bioresorbable vascu- lar scaffold versus everolimus-eluting metallic stent in ST-seg- ment elevation myocardial infarction: 1-year results of a propen- sity score

Clinical utility of optical coherence tomography (OCT) in the optimisation of Absorb bioresorbable vascular scaffold deployment during percutaneous coronary intervention.. Prati

The aim of this study was an optical coherence tomography (OCT) assess- ment of acute procedural result of the everolimus-eluting BVS implantation in patients with ST segment

Repeat coronary angiography and intravas- cular ultrasound confirmed a patent left main coro- nary artery.. Successful percutaneous coronary in- tervention with drug-eluting

The decline in unemployment in subsequent years was influenced primarily by two factors: a strong economic recovery (in 2004-2007 the average annual GDP growth was 5.5%), and