Image integration of cine-angiography with 3D electroanatomical mapping. Atrial flutter ablation with CARTO UniVuTM module support

N/A
N/A
Protected

Academic year: 2022

Share "Image integration of cine-angiography with 3D electroanatomical mapping. Atrial flutter ablation with CARTO UniVuTM module support"

Copied!
1
0
0

Pełen tekst

(1)

www.kardiologiapolska.pl

Kardiologia Polska 2015; 73, 10: 941; DOI: 10.5603/KP.2015.0186 ISSN 0022–9032

STUDIUM PRZYPADKU / CLINICAL VIGNETTE

Address for correspondence:

Krzysztof Myrda, MD, 3rd Department of Cardiology, Silesian Centre for Heart Diseases, ul. M. Skłodowskiej-Curie 9, 41–800 Zabrze, Poland, e-mail: k_myrda@interia.pl

Conflict of interest: none declared

Image integration of cine-angiography with 3D electroanatomical mapping. Atrial flutter ablation with CARTO UniVu TM module support

Zintegrowanie systemu elektroanatomicznego 3D z obrazem fluoroskopowym w czasie rzeczywistym. Wstępne doświadczenia z modułem CARTO UniVu

TM

Piotr Buchta, Krzysztof Myrda, Mariusz Gąsior, Lech Poloński

3rd Department of Cardiology, SMDZ in Zabrze, Medical Uniwersity of Silesia, Katowice, Poland

Radiation exposure during cardiac electrophysiological procedures is still a major concern. Until now reduction of fluoro­

scopy time and dose during ablation procedures was achieved mostly through the use of three­dimensional (3D) electro­

anatomical systems. For more effective radiation exposure control and improvement of ablations results Biosense Webster created new 3D non­fluoroscopic navigation system software. The CARTO UniVuTM module of CARTO® 3 system allows real­time electroanatomical localisation of ablation catheters in pre­recorded X­ray images. Other technical details of this module were described previously. Here we report the first Polish experiences with mapping and ablation of typical atrial flutter performed with support of a CARTO UniVuTM module. Under fluoroscopy control, two diagnostic catheters were placed — one in the right ventricle apex position (quadripolar, nonsteerable, CRD, SJM) and the second, a 10­polar, steerable catheter (Inquier, SJM), in the coronary sinus. For electroanatomical mapping and ablation the Thermocool® SmartTouchTM catheter (Biosense Webster, Inc., Diamond Bar, CA, USA), which also allows measurements of catheter contact force, was chosen; 41 s and 12 mGy fluoroscopy for diagnostic catheter positioning was used. Afterwards, with the CARTO UniVuTM module, the reference fluoro frame at the standard right (RAO 30o) and left (LAO 60o) anterior oblique projections was captured (Fig. 1). We terminated the arrhythmia during the ablation of the cavotricuspid isthmus (CTI), and we created a bidirectional block after a total procedure time of 50 min. The total fluoroscopy time was 50 s, with a total fluoroscopy dose of 15 mGy. Ablation of CTI is recommended in the treatment of typical atrial flutter. During the procedure, the ablation catheter is typically guided by conventional fluoroscopy. Nevertheless, a non­fluoroscopic or low­fluoroscopic approach is preferable. Here we demonstrated the feasibility and safety of the CARTO® 3 system with a UniVu module.

Using this software combined with a contact force ablation catheter, a reduction in fluoroscopy ablation time to 10 s was possible. Decreasing the fluoroscopy time by CTI ablation was also shown with other systems. Using 3D navigation tech­

niques in pre­recorded 2D fluoros­

copy — MediGuide technology — allowed a decrease in total fluoro­

scopy time of up to 2.5 ± 2 min in 10 consecutive patients. Ablation with remote magnetic navigation decreased the fluoroscopy time by up to 7.2 (3.2–12.2) min in series data. There are published experi­

ences with a CARTO 3 UniVu mod­

ule in the ablation of atrial flutter. In 58 patients a significant reduction in radiation exposure time was observed, from 8.6 ± 0.8 min, with a CARTO® 3 to 2.9 ± 0.3 min with a CARTO UniVuTM. The authors em­

phasise that fluoroscopy reduction was achieved without prolongation of the procedure time.

Figure 1. Standard RAO 30o and LAO 60o projections with placed diagnostic catheters captured during cavotricuspid isthmus (CTI) ablations with support of a CARTO® 3 with a CARTO UniVuTM module and Thermocool® SmartTouchTM ablation catheter

Cytaty

Powiązane dokumenty

Methods and Results: One hundred and ten patients were treated with pulmonary vein isolation with 4 different technologies: open-irrigated tip radiofrequency (RF) catheter

Pul- monary vein isolation (PVI)-focused ablation, which eliminates triggers, and partially elimi- nates substrate located near LA-PVs junction, results in high success rate in

There are no randomized prospective trials comparing the safety and efficacy of catheter abla- tion for paroxysmal or persistent AF in the elderly to best medical therapy or

The complication was observed in a female patient who had been using oral contraceptives for many years and consisted in sud- den dyspnoea which developed on the first night af-

Annotation of conduction block based on the vector of propagation (opposed direction, automatic annotation), low voltage, and double potentials; B – potential map of the RA in the

The aim of this study was to compare safety, efficacy, and cost ‑effectiveness of 2‑electrode (2C) versus 3‑electrode (3C) approach and to identify clinical predictors of

The growing number of interventions was accompanied by relatively low in ‑hospital and 1‑year mortality rates of patients with atrial fibrillation / atrial flutter referred

Efficacy of ganglion- ated plexi ablation in addition to pulmonary vein isolation for paroxysmal versus persistent atrial fibrillation: a meta -analysis of randomized