• Nie Znaleziono Wyników

Three-dimensional transoesophageal echocardiography as the ultimate diagnostic tool in a case of unintentional left ventricular pacing

N/A
N/A
Protected

Academic year: 2022

Share "Three-dimensional transoesophageal echocardiography as the ultimate diagnostic tool in a case of unintentional left ventricular pacing"

Copied!
1
0
0

Pełen tekst

(1)

www.kardiologiapolska.pl

Kardiologia Polska 2015; 73, 8: 664; DOI: 10.5603/KP.2015.0148 ISSN 0022–9032

Studium przypadku / CliniCal Vignette

Address for correspondence:

Roman Steckiewicz, MD, PhD, Department of Cardiology, Medical University of Warsaw, ul. Banacha 1A, 02–097 Warszawa, Poland, tel: +48 22 599 29 58, e-mail: r.steckiewicz@pro.onet.pl

Conflict of interest: none declared

Three-dimensional transoesophageal echo- cardiography as the ultimate diagnostic tool in a case of unintentional left ventricular pacing

Trójwymiarowa echokardiografia przezprzełykowa, czyli „kropka nad i”

implikacji diagnostycznych niezamierzonej stymulacji lewej komory

Piotr Scisło

1

, Roman Steckiewicz

1

, Przemysław Stolarz

1

, Elżbieta Świętoń

2

, Marcin Grabowski

1

1Department of Cardiology, Medical University of Warsaw, Warsaw, Poland

2Department of Cardiology, Central University Hospital in Warsaw, Warsaw, Poland

Unintentional lead placement in the left ventricle (LV) is a rare complication of pacemaker implantation. Despite the fact that these procedures are routinely guided by electrocardiography and fluoroscopic imaging, these monitoring methods may occasionally prove to be insufficient. These are the cases

when echocardiography, especially three-dimensional (3D) echocardiography, may be of help. An 83-year-old male was admitted to our clinic due to suspected ventricular lead dislocation into the LV, detected in a routine follow-up chest X-ray film. Five years earlier, the patient underwent implanta- tion of a Victory DR (DDDR) cardiac pacing system (from St.

Jude Medical) with screw-in leads (atrial Optisense 52 and ventricular Optisense  58) due to complete atrioventricular block. The pacing parameters were good both initially and at the interim follow-up outpatient interrogations. A 12-lead electrocardiogram (ECG) tracing showed normal cardiac pac- ing in the DDDR mode with pacemaker-generated ventricular complexes of right bundle branch block (RBBB) morphology (Fig. 1A). Lateral-view fluoroscopy revealed the electrode forming a backward-facing arch (Fig. 1C, arrow), which is consistent with the ventricular lead passing from the right atrium (RA) through the foramen ovale into the LV. A detailed, three-dimensional (3D) course of both leads was visualised via 3D echocardiography. During lead insertion the ventricular lead was deflected by an abnormally large muscle rim in the RA, which altered its course (Fig. 2A). The lead passed from the RA through the anterior edge of the foramen ovale, and into the left atrium (LA) (Fig. 2B). Finally, the lead passed through the mitral valve at the border of segments P2 and P3 into the LV (Fig. 3A), where its tip eventually rested in the basal part of the lateral apical segment of the LV (Fig. 3B). Right endo- cavitary pacing produces electrocardiographic evidence of left bundle branch block (LBBB). However, RBBB can also be seen in 8.3% of ECG tracings, a proportion that can be reduced to 4.3% by tracing modification with the Klein’s manoeuvre (lowering precordial leads by one intercostal space). Radio- graphic anterior-posterior images do not always show lead placement accurately and may fail to visualise complications of permanent pacing. In the case presented here, the suspected complication was confirmed echocardiographically. The 3D mode allowed an accurate visualisation of lead placement in the LV and helped explain the likely cause of lead misdirection.

Figure 2. A, B. Three-dimensional transoesophageal echo- cardiography — two views from the middle oesophagus, modified. The lead passes through interatrial septum (IAS) in the anterior part of the foramen ovale; LA — left atrium;

LV — left ventricle; RA — right atrium; TV — tricuspid valve

Figure 3. Three-dimensional transoesophageal echocardio- graphy — the view from the middle oesophagus, modified;

A. View from the left atrium (LA) roof towards the mitral valve (MV). Arrows mark the course of the lead toward seg- ment P3 of the MV; B. View from the left ventricle (LV) apex towards the MV. The arrow marks the tip of the lead in the LV wall; IAS — interatrial septum; AV — aortic valve

Figure 1. A. Atrioventricular (dual-chamber) cardiac pacing at 70 bpm; B, C. Fluoroscopy of the chest: anterior-posterior (A-P) view (B) and left lateral oblique view (ventricular lead

— arrow) (C)

A

A

A

B

B

B

C

IAS

IAS

IAS

Cytaty

Powiązane dokumenty

‑anterior view and left anterior oblique 30º view, respectively): both frames obtained after His bundle lead removal and implantation of left bundle branch (LBB) lead (arrows); f

The MyoPore (Greatbatch Medical, New York, United States) sutureless screw -in epicardial pacing lead was delivered to the target area with the steerable FasTac Flex

kinje system pacing has been proposed, namely, deep septal pacing with direct left bundle branch (LBB) capture. 2 This can be achieved

Assessments of right ventricular volume and function using three-dimensional echo- cardiography in older children and adults with congenital heart disease: comparison with

Conclusions: Male gender, QRS duration greater than 140 ms, discordant LBBB, and residual conduction in the left bundle branch seem to be markers of reduced LVEF in patients

A few days after the coronary angiogram, in ECG record, atrio-ventricular (AV) dissociation, sinus rhythm (HR of 55 bpm) with right bundle branch block (RBBB), and slightly

RIGHT VENTRICULAR PACING AND THE OCCURRENCE OF VT/VF EPISODES The possibility of harmful effects of RV pacing in patients with reduced LV systolic function with an ICD was

A high degree right bundle branch block (RBBB) is easily recognised in the precordial leads which have a peculiar feature: the precordial transition is strongly displaced to