Kardiologia Polska 2017; 75, 1: 79; DOI: 10.5603/KP.2017.0004 ISSN 0022–9032

STUDIUM PRZYPADKU / CLINICAL VIGNETTE

Address for correspondence:

Krzysztof Boczar, MD, Department of Electrocardiology, John Paul II Hospital in Krakow, ul. Prądnicka 80, 31–202 Kraków, Poland, e-mail: krzysiek.boczar@gmail.com Conflict of interest: none declared

Kardiologia Polska Copyright © Polskie Towarzystwo Kardiologiczne 2017

Spontaneous dislocation of the endocardial lead into the left ventricle through

the intraventricular septum

Samoistne przemieszczenie elektrody endokawitarnej do światła lewej komory poprzez przegrodę międzykomorową

Krzysztof Boczar

1

, Monika Komar

2, 3

, Andrzej Ząbek

1

, Jacek Lelakowski

1, 3

, Barbara Małecka

1, 3

1Department of Electrocardiology, John Paul II Hospital in Krakow, Krakow, Poland

2Department of Cardiac and Vascular Diseases, John Paul II Hospital in Krakow, Krakow, Poland

3Institute of Cardiology, Jagiellonian University Medical College, Krakow, Poland

We report a case of a 66-year-old woman with a history of DDD pacemaker implantation in 2005 due to vasovagal syncope and generator change in 2014. The patient was implanted with an atrial lead Biotronik SX 53-JBP positioned in the right atrial appendage and a ventricular lead Biotronik SX 60-BP positioned in the right ventricular apex. A few months after generator change the patient underwent a diagnostic process due to heart palpitations and stabbing pain in the chest. The chest pain occurred during body movements and deep breathing. In electrocardiographic Holter moni- toring the sensing disturbances in the ventricular channel were detected. During pacemaker interrogation the electrical parameters of both leads were normal except for an increased number of short V-V intervals. The patient was qualified to transvenous ventricular lead extraction and implantation of a new ventricular lead. In the meantime the patient un- derwent a computed tomography scan in search of coronary heart disease, and it was found that the ventricular lead was perforating the intraventricular septum on the level of the apical septal segment. The length of the lead in the left ventricle amounted to 7–9 mm (Fig. 1A). The displacement of the lead was suspected also in a chest X-ray of the current location of the lead (Fig. 1B). In transthoracic and transoesophageal echocardiography the thrombus on the lead tip and left-to-right shunt through the septum along the lead were ruled out (Fig. 1C). The decision to extract the ventricular lead transvenously with mechanical systems with surgical backup and transoesophageal echocardiography monitoring was upheld. The procedure was a complete success. There was no residual left-to-right shunt after the extraction of the lead. Subsequently the new ventricular lead was implanted (Fig. 1D). The presented case is an example of a rare and dangerous complication of permanent pacing. The existence of a short portion of ventricular lead inside the left ventricle posed a risk of thrombus formation and peripheral embolisation, of which the most dangerous is stroke. The late perforation of intraventricular septum by a passive fixation lead has not been reported in English-language literature yet. The probable cause of the distant (in time) perforation was the application of excessive force on the tip of the lead during implantation, which was then aggravated by the ingrown lead into the tricuspid valve region. Transvenous extrac- tion of the lead displaced into the left ventricle requires surgical intervention according to the current Heart Rhythm Society (HRS) expert consensus. In this case a less invasive approach was successfully used. However, close attention was directed toward identification and prevention of potential perioperative complications such as left-to-right shunt after lead extraction and arterial embolisation. Furthermore, the procedure was conducted with surgical backup and continuous intraoperative transoesophageal monitoring.

Figure 1. A. Computed tomography scan, arrow indicates the tip of ventricular lead in the left ventricle; B. Chest X-ray before transvenous lead extraction procedure, arrow indicates displaced tip of the ventricular lead; C. Transoesophageal echocardiography with colour Doppler, arrow indicates right ventricular lead in left ventricle (LV); RV — right ventricle;

IVS — intraventricular septum; D. Chest X-ray after implantation of new DDD pacemaker

A B C D

Cytaty

Powiązane dokumenty

In addition, numerous real ‑life registries have reported favorable outcomes with Watchman, ACP, and Amulet devices in patients at higher bleeding risk and / or

The MyoPore (Greatbatch Medical, New York, United States) sutureless screw -in epicardial pacing lead was delivered to the target area with the steerable FasTac Flex

The coronary computed tomographic angiography showed an abnormal, wide branch of 4 mm calibre, extending from the mid segment of the left anterior descending artery (LAD) (Fig.

Due to the large size of the vegetation (the length: 71 mm) connected to the lead and its coarse texture (Fig. 1), the patient was qualified for complete system removal during

Percutaneous closure of the left atrial appendage versus warfarin therapy for prevention of stroke in patients with atrial fibrillation: a randomised non-inferiority trial.

A 46-year-old Caucasian male patient with brady-tachycardia syndrome and a presentation of sinus bradycardia with oc- casional, atypical atrial flutter episodes had received the DDD

1 Division of Clinical Electrophysiology, Department of Cardiology, Grochowski Hospital, Postgraduate Medical School, Warsaw, Poland.. 2 Department of Cardiology, Grochowski

In the current case, an increase in the rate during ventricular fibrillation/flutter resulted in an augmentation of the depolarisation problem over the right ventricle manifested