Successful coronary lithotripsy after unsuccessful rotational atherectomy – use of complementary tools for treatment of undilatable calcified coronary artery stenosis

N/A
N/A
Protected

Academic year: 2022

Share "Successful coronary lithotripsy after unsuccessful rotational atherectomy – use of complementary tools for treatment of undilatable calcified coronary artery stenosis"

Copied!
2
0
0

Pełen tekst

(1)

228 Creative Commons licenses: This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY -NC -SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/).

Image in intervention

Corresponding author:

Piotr M. Wańczura MD, PhD, Department of Cardiology, The Ministry of Internal Affairs and Administration, Hospital, Rzeszow, Poland, e-mail: p.wanczura@poczta.fm

Received: 29.03.2021, accepted: 5.05.2021.

Successful coronary lithotripsy after unsuccessful

rotational atherectomy – use of complementary tools for treatment of undilatable calcified coronary artery stenosis

Piotr M. Wańczura1, Wojciech Wojakowski2

1Department of Cardiology, The Ministry of Internal Affairs and Administration, Hospital, Rzeszow, Poland

2Department of Cardiology and Structural Heart Diseases, Medical University of Silesia, Katowice, Poland

Adv Interv Cardiol 2021; 17, 2 (64): 228–229 DOI: https://doi.org/10.5114/aic.2021.107507

Despite the progress in percutaneous treatment of coronary artery disease, massively calcified atherosclerot- ic lesions still constitute a challenge. Rotational atherec- tomy, capable of modifying the atherosclerotic plaque, known as debulking, has become the gold standard in the treatment of calcified lesions [1]. Recently, intravas- cular lithotripsy (IVL) became an alternative treatment option for undilatable calcified vessels. This technique uses a balloon-mounted ultrasound source which emits sonic pressure waves propagating through the fluid and delivered to the vessel wall. It results in fragmentation of superficial and deep calcium deposits allowing balloon expansion [2]. One limitation of this method is the rel- atively stiff, high-profile balloon, with lower crossability than regular semi- or non-compliant balloons routinely used during coronary angioplasty, especially in tight cal- cified lesions. Therefore effective IVL requires predila- tation and may not be possible in uncrossable lesions.

On the other hand, rotational atherectomy, even if per- formed with burs of increasing diameter, may be less effective in large-caliber vessels with calcifications lo- cated in the media. In such cases, an effective method combining initial debulking with rotational atherectomy followed by IVL may be an example of a comprehensive approach overcoming the limitations of both methods.

IVL should be considered when rotational atherectomy, despite the use of a sufficiently large burr, is ineffective.

The criteria for such a  comprehensive approach to the massively calcified lesion in right coronary artery (RCA) were met by a  73-year-old patient with arterial hyper- tension, hyperlipidemia, persistent atrial fibrillation, and history of myocardial infarction treated with left anterior descending (LAD) percutaneous coronary intervention (PCI) followed by elective recanalization of the circumflex

chronic total occlusion (CTO). Due to persistent angina (CCS III class), the patient was qualified for PCI of the RCA. The predilatation with the NC balloon (3.5 × 15 mm) was not successful due to the calcified undilatable lesion located in segment 2 of the RCA. We performed rota- tional atherectomy with the 1.5 mm and then 1.75 mm burrs (Figure 1 A). Despite the high pressure (24 atm) inflation of the 3.5 × 20 mm NC balloon, full expansion was not achieved (Figure 1 B). There was dog-boning of the balloon and non-flow limiting dissection. The patient was clinically stable, and a second attempt with IVL was planned after 4 weeks. The procedure was performed using a 3.5 × 12 mm IVL balloon (Shockwave Medical), which, after completing lithotripsy cycles, allowed for full expansion of the balloon at 6 atm (Figure 1 C). Finally, 2 drug-eluting stents (Promus 3.5 × 24 and Promus 3.5

× 20 mm) were implanted with an optimal angiographic effect (Figure 1 D).

IVL is a valuable supplement to the available revas- cularization toolbox, consisting of rotational atherecto- my, ultra high-pressure balloons, scoring, and cutting balloons. In the case of lesions that are too tight for the delivery of the lithotripsy balloon and, at the same time, not susceptible to expansion despite debulking with the available burr sizes, a hybrid approach based on a com- bination of both methods may be the only effective solu- tion. Coronary aortic bypass surgery always remains an alternative, although lesions not involving the left main or proximal LAD segment are relatively rarely qualified for surgical treatment.

Conflict of interest

The authors declare no conflict of interest.

(2)

Piotr M. Wańczura et al. Coronary lithotripsy after failed ROTA

229

Advances in Interventional Cardiology 2021; 17, 2 (64) References

1. Kassimis G, Raina T, Kontogiannis N, et al. How should we treat heavily calcified coronary artery disease in contemporary prac- tice? From atherectomy to intravascular lithotripsy. Cardiovas- cular Revasc Med 2019; 20: 1172-83.

2. Dini CS, Tomberli B, Mattesini A, et al. Intravascular lithotipsy for salcific coronary and peripheral artery stenoses. Eurointerven- tion 2019; 15: 714-21.

Figure 1. A – Rotational atherectomy with 1.75 mm burr, B – unexpanded 3.5 × 20 mm NC baloon, C – full expanded 3.5 × 12 mm IVL balloon, D – final result after DES implantation

A

C

B

D

Cytaty

Powiązane dokumenty

After initial pharmacological stabilization, coronary angiogram was performed and re- vealed a chronic total occlusion of the right coronary artery (Figure 1A) and circumflex, with

Coronary angiography, fluoro- scopic digital stent enhancement (DSE), and intravascular ultrasound revealed recurrence of ISR in the proximal segment of the LAD caused by

onary artery was intubated with a 4.0/6F extra backup guiding catheter (Medtronic), a Fielder guidewire (Asahi Intecc, Aichi, Japan) was placed distally in the Cx..

1 With enhanced stent design and device technology, the ability to deliver a stent has improved at the cost of necessary and adequate lesion preparation for calcified lesions

Successful shock- wave intravascular lithotripsy for a severely calcified and undilatable left anterior descending coronary artery lesion in a patient with recurrent myo-

Coronary angiography showed a patent left main coronary artery and chronic total occlusions (CTOs) of the following vessels: mid segment of the left anterior descending artery

In the current study, the percentage of complex lesions with advanced calcifications was high (78.3% of type C and 21.7% of type B) and comparable with other studies.

Follow- ing intubation with a 7F AL 0.75 guiding cath- eter, and sequential high-pressure predilatation (1.2 mm semi-compliant balloon, and 2.0 mm to 3.0 mm non-compliant