• Nie Znaleziono Wyników

Successful removal of side branch-jailed stent struts by reused balloon

N/A
N/A
Protected

Academic year: 2022

Share "Successful removal of side branch-jailed stent struts by reused balloon"

Copied!
2
0
0

Pełen tekst

(1)

Address for correspondence: Kenichi Tsujita, MD, PhD, FJCC, FJCS, FESC, FACC, Department of Cardiovascular Medicine, Graduate School of Medical Sciences, Kumamoto University, 1-1-1, Honjo, Chuo-ku, Kumamoto, 860-8556, Japan, tel: +81-96-373-5175, fax: +81-96-362-3256, e-mail: tsujita@kumamoto-u.ac.jp

Received: 16.07.2018 Accepted: 17.08.2018

INTERVENTIONAL CARDIOLOGY

Cardiology Journal 2018, Vol. 25, No. 5, 644–645

DOI: 10.5603/CJ.2018.0130 Copyright © 2018 Via Medica

ISSN 1897–5593

644 www.cardiologyjournal.org

IMAGE IN CARDIOVASCULAR MEDICINE

Successful removal of side branch-jailed stent struts by reused balloon

Kazumasa Kurogi

1

, Kenji Sakamoto

2

, Kyohei Marume

1

, Hiroaki Kusaka

1

, Nobuyasu Yamamoto

1

, Kenichi Tsujita

2

1Department ofCardiovascular Center, Miyazaki Prefectural Nobeoka Hospital, Nobeoka, Japan

2Department of Cardiovascular Medicine, Graduate School of Medical Sciences, Kumamoto University, Kumamoto, Japan

Three-dimensional (3D) optical coherence to- mography (OCT) (St. Jude Medical, St. Paul, MN)- -guided coronary stenting was performed for steno- sis at the bifurcation of the left anterior descending artery (LAD) and diagonal (DG) branch. A coronary angiography revealed true bifurcation lesion with Medina classification (1,1,1) (Fig. 1A). After lesion modification with rotational atherectomy and kissing balloon inflation (KBI) technique, crossover stent- ing with a Resolute Onyx zotarolimus-eluting stent (2.25 × 30 mm, Medtronic Cardiovascular, Santa Rosa, CA) was done in the LAD lesion (Fig. 1B).

A guidewire was recrossed to the DG branch fol- lowing the proximal optimization technique with a 2.25 × 8 mm noncompliant balloon. Post-rewired 3D-OCT imaging visualized quite small distal cell and recrossed guidewire position at the proximal larger cell (Fig. 1C, D). After confirming proximal rewiring by 3D-OCT imaging, rewiring the distal cell was attempted. This option was not selected as the distal cell was quite small and it was decided to perform KBI in the proximal cell. As there- wrapped 2.0 × 12 mm noncompliant balloon used in pre-stenting KBI could not be advanced to the DG branch, presumably because of the jailed strut, using 2.25 mm noncompliant balloon located in the distal main vessel as an anchor balloon, the balloon was finally pushed into the DG branch (Fig. 1E).

Final angiography after post-stenting KBI (2.25 ×

× 8 mm to LAD and 2.0 × 12 mm to DG branch) showed favorable findings with optimal dilatation of both LAD/DG branch lesions (Fig. 1F). A final 3D-OCT image pullback from the LAD revealed

excellent stent expansion and disappearance of the jailed strut at the DG orifice (Fig. 1G), and 3D-OCT pullback from the DG branch clearly im- aged good apposition of the strut folded toward the DG branch at a sharp angle covering the LAD/DG carina (Fig. 1H).

In bifurcation stenting, 3D-OCT imaging ena- bles visualization of the link of stent struts and also to control the recrossing point of the side-branch guidewire. It has been demonstrated that distal rewiring resulted in favorable stent positioning against the side branch ostium leading to a reduction of incomplete stent apposition and potentially bet- ter clinical outcomes in the free carina type. As the jailed strut may remain at the carina even after KBI when recrossing the “proximal” cell, “balloon push- fold” method with proximal cell guidewire crossing has been proposed to invert the strut toward the side branch in this situation. In the current case, it was speculated that pushing the rewrapped bal- loon with larger crossing profile combined with the

“Anchor balloon technique” folded the jailing strut, and final KBI appose it to the side-branch-sided ca- rina. Furthermore, the attachment of struts on the distal main vessel by anchor balloon may prevent stent deformation and relate to folding of the jailing strut to the side branch. 3D-OCT-based guidewire recrossing strategy can help the development of optimal bifurcation stenting strategy.

Conflict of interest: Dr. Kenichi Tsujita has re- ceived grants from Goodman Co., Ltd. All the other authors have nothing to disclose.

(2)

Figure 1. Coronary angiography and three-dimensional (3D) optical coherence tomography (OCT) findings. Initial coronary angiography (CAG) (A) showed significant stenoses at the bifurcation of the left anterior descending artery (LAD) and the diagonal (DG) branch with a Medina classification (1,1,1). After LAD main branch stenting (B), post- stenting 3D-OCT pullback from LAD (C. Frontal view; D. Lateral view) visualized existence of jailed stent strut (*). Fol- lowing rewrapped balloon delivery to DG branch with the anchor balloon technique (E), final CAG confirmed optimal angiographic findings (F). Final 3D-OCT pullback from LAD (G) and DG branch (H) revealed favorable folding of jailed strut toward DG branch.

www.cardiologyjournal.org 645

Kazumasa Kurogi et al., 3D OCT of a push-fold method with a reused balloon

Cytaty

Powiązane dokumenty

The building work for the extension will begin after the end of the school term, but there will probably be a certain amount of disruption when the students return

Wynika z niego, Ŝe transport masy dla struktur krótkokanałowych jest znacznie intensywniejszy, niŜ dla klasycznych monolitów o długich kanałach: liczby Sherwooda

these previous works do not satisfy the polynomial equation [6]. Uniquely, the authors have circum- navigated these pitfalls in detail, the basics of the assumptions were: 1)

— Group 2: Elliptical ostial shape at SB ostium assumed after MV stenting — calculated with SB reference diameter, taken into account for long axis ellipse calculation according to

The baseline coronary angiogram (C), the longitu- dinal cut-away view of the stent-enhanced 3D optical coherence tomography (D) and the stent-oriented 3D-IVUS (E) of another

CC-Prox; the metal carina index is defined as the number of frames with malapposed struts continuously present from the distal end of the side branch ostium (SBO) to the proximal

The successful retrieval of a broken guide wire from the diagonal branch of the left anterior descending coronary artery complicated by partial stent rolling.. Rafał

We present direct results and clinical short- and mid-term follow-up of 40 consecutive patients treated with Tryton stent implantation for bifurcation coronary artery lesions..