Adaptation of Nanoarrays for the Study of

Synuclein Aggregation

-Preliminary Results

Heidelinde R. C. Dietrich

a

, Richard L. van den Doel

a

, Wolfgang Hoyer

b

, Wim van Oel

a

, Guus Liqui

Lung

a

, Yuval Garini

a

, Thomas Jovin

b

, and Ian T. Young

a

a

Department of Imaging Science and Technology, Faculty of Applied Sciences, Delft University of

Technology, Lorentzweg 1, 2628 CJ Delft, The Netherlands

b

Department of Molecular Biology, Max-Planck-Institute for Biophysical Chemistry, Am Fassberg

11, 37077 Goettingen, Germany

ABSTRACT

In previous publications we have shown that we can perform enzymatic reactions in nanoarrays by means of a microarray-reader based on a conventional microscope. In this publication we report on a modification of this system in order to monitor the aggregation kinetics of the natively unfolded protein-synuclein. We describe the motivation for this development, the problems associated with the miniaturization of the aggregation assay, and the validation of our modifications.

Keywords: Nanoarrays, High Throughput Screening, Evaporation Control, -Synuclein, Thioflavin T, Quantitative

Fluorescence Microscopy

1. INTRODUCTION

A large number1,2 of proteins show either partially disordered regions or lack an ordered secondary structure under physiological condition. These natively unstructured proteins are classified into two groups: 1) proteins that are flexible but still possess a secondary structure, e.g. molten globules; and 2) proteins that behave essentially as random coils. The term “natively unfolded” has been employed for proteins of the second class. The primary structure of a protein determines its secondary structure. More than 15,000 proteins share a pronounced degree of structural disorder and these are mostly rich in certain amino acids such as glutamic acid, lysine, and glycine. Natively unfolded proteins characteristically exhibit a high net-charge and low hydrophobicity, as well as the tendency to form depositions associated with several neurodegenerative disorders. Examples are Parkinson’s disease (PD), Alzheimer’s disease, dementia with Lewy body, and multiple system atrophy (MSA). The lack of an ordered secondary structure also potentiates low affinity interactions with other molecules. Thus, the regulation of many cellular processes may be controlled by the transition from a disordered to an ordered state of a participating protein.

-Synuclein1,3

is a member of the class of natively unfolded proteins and constitutes the major component of Lewy bodies, fibrillar cytoplasmic inclusions found in cells of the substantia nigra in the brain of individuals suffering from PD. -Synuclein is a 14 kDa (140 amino acids) pre-synaptic protein abundant in various regions in the brain but its natural functions are still a matter of speculation. -Synuclein lacks a well-defined secondary structure under physiological conditions but exhibits well demarcated domains, some of which adopt alternative conformations under different conditions. An interesting recent finding4,5is the existence of distinct intramolecular interactions that appear to establish a native state of autoinhibition with respect to aggregation, a process triggered by mutation, low pH, high temperature and the binding of numerous ligands. The-sheet conformation is characteristic of the resulting amyloid structures, including the fibrils pathognomonic of the dopaminergic neurons affected in PD.

Thioflavin T6,7,8. (ThioT) is a classical stain for amyloid depositions in brains from Alzheimer’s disease infected individuals. It is also in widespread use in kinetic studies of fibril formation in vitro. Unbound ThioT has an absorption peak at 350 nm and an emission peak at 430 nm. When bound to aggregated -synuclein, particularly fibrils, the absorption peak shifts to 450 nm and the emission peak to 485 nm; the intensity also increases dramatically. Whereas

Imaging, Manipulation, and Analysis of Biomolecules and Cells: Fundamentals and Applications III, D. V. Nicolau, J. Enderlein, R. C. Leif, D. L. Farkas, R. Raghavachari, eds., Proc. of SPIE Vol. 5699 (SPIE, Bellingham, WA, 2005) · 1605-7422/05/$15 · doi: 10.1117/12.587493

395

(2)

ThioT binds rapidly and specifically to the anti-parallel-sheet fibrils, it does not interact with the monomeric form or oligomeric intermediates.

The development of rational therapeutic strategies requires an understanding of the underlying mechanisms and factors mediating oligomerization and fibrillization. However, it is difficult to explore this multidimensional reaction scheme because of the characteristically long lag times and short transition times for these reactions4. The aggregation kinetics of-synuclein are usually determined in milliliter cuvettes or in 96-well microtiter plates. We are interested in scaling down the assay volume to eight nanoliter “cuvettes” that we term nanowells, implying a500,000-fold reduction in the overall assay volume. The miniaturization leads to numerous improvements:

• a dramatic reduction in the amounts of -synuclein and all other compounds necessary for the assay; • a significant reduction in the physical space required for the measurement instrumentation;

• the ability to perform a large number of tests in parallel through the array of nanowells, and;

• a reduction in the time required for the biochemical reactions in which the rate-limiting process is diffusion. Our read-out platform is based on a conventional microscope system. In this paper we discuss the issues involved in miniaturizing the aggregation assay.

2. BACKGROUND

In previous publications, we reported on a miniaturized system designed to monitor enzymatic reactions9-13. This system is based on nanoarrays with volumes of 6.3 and 8 nanoliters depending on the shape of the wells; see Table 1.

Table 1: Well sizes used for enzyme reactions studies and-synuclein experiments

Depth [µm] Square well 400 x 400µm2 Round well 400µµm2(diameter) 50 8.0 nL 6.3 nL

The wells are etched into silicon and have a top layer of silicon nitride. The nanoarray chip itself has a physical size of 2  1 cm2

and two arrays of 5 5 wells have been implemented per chip14.

The read-out system is based on a conventional, epi-illumination Zeiss Axioskop microscope equipped with a scientific grade CCD camera15,16. The system uses a relatively high numerical aperture objective: 20 / 0.75 FLUAR Zeiss. With the 20 magnification, an entire 400 m-wide well occupies one image, while the 0.75 NA permits the capture of a significant amount of the light emitted by the fluorescent ThioT. The trinocular tube with the eyepieces and the relay optics have been replaced with an inverted 5 / 0.25 FLUAR Zeiss objective (Figure 1). This inverted lens concentrates the measured light onto the smallest possible area of the CCD sensor, producing a higher signal-to-noise ratio (SNR) in the final digital image than if the light had been distributed over a larger number of CCD pixels14. The 100 W mercury-arc light source (USH-102D, Ushio Inc., Tokyo, Japan) and the associated optics are configured for Köhler illumination. The CCD camera is a Princeton Versarray 512B Back-illuminated CCD camera with 512  512 pixels (Princeton Instruments, Trenton, New Jersey, USA). Each pixel has a size of 24µm  24 µm.

(3)

Figure 1: The microscope based read-out system is shown: a Zeiss Axioskop using a 20 x / 0.75 objective, an

inverted 5 x / 0.25 objective and a Princeton Versarray CCD camera.

3. EVAPORATION ISSUES

Aggregation assays17,18 with-synuclein can require substantial amounts of time; an experiment taking three or more days is not unusual. Nanoliter quantities of fluid, however, can evaporate in a few seconds. In order to prevent evaporation of the liquid in the wells, we house the nanoarray chips in a transparent, plastic chamber that provides one opening for the 20 microscope objective, two openings for forced air transport (input and output), and electrical ports for temperature and humidity measurements and controls. Within this chamber we provide a feedback-controlled temperature and humidity environment for the nanoarrays used in-synuclein experiments as well as other experiments. Temperature control of the chip is straightforward and accomplished with a PE120 heating/cooling stage and a PE94 temperature controller (Linkam, Tadworth, UK). The control of the humidity, however, is a more complicated issue. Distilled water is held in a small plastic reservoir. The humidity inside the experimental chamber is measured and the data controls a nebulizer, a device that turns liquids into a fine mist. The mist is then transported by forced air to the chamber to achieve the desired humidity. Figure 2 shows the front view of the chamber plus connection tubes.

Figure 2: Front view of the self-made humidifying chamber build for an upright Zeiss Axioskop microscope.

Figure 3 depicts the plastic container holding the smaller water reservoir. The mist is produced using a miniature nebulizer DK-24 (Conrad Electronic GmbH, Hirschau, Germany). Distilled water is atomized by means of an ultrasonic resonator at a rate of 250 mL per hour. Because of the vibrations associated with the nebulizer, it cannot be placed upon the same table as the microscope. A rigid plastic tube with a large section removed (see Figure 3) is mounted

Proc. of SPIE Vol. 5699 397

(4)

horizontally above the water tank and goes through the larger container to the outside. There it is connected to two 100 mm diameter flexible tubes. The large hole in the rigid tube allows the mist to be directed to the experimental chamber by a 100 mm inline fan. The whole system is closed.

An ordinary hair dryer connected to the larger container is used to heat the circulating air. This allows a higher humidity uptake. The inline fan, the hair dryer and the nebulizer are controlled by PAC2 phase-angle power controllers (United Automation, Southport, UK) that use analog control voltages to regulate the devices.

Figure 3: Water reservoir with mist generator and air control.

The humidifying chamber is equipped with a PIC16F873 microcontroller (Microchip Technology Inc., Chandler, Arizona, USA) that utilizes a humidity and temperature sensor SHT-71 (Sensirion, Zürich, Switzerland). This sensor (Figure 4) consists of a single chip that includes a calibrated digital output. The chip can be used for temperatures between –40°C and 120°C and relative humidities between 0% and 100%. The SHT-71 is factory calibrated with calibration coefficients programmed into the OTP (one-time programmable) memory of the sensor. As shown in Figure 5, the desired values for temperature and relative humidity are set from 00 to 99 by means of BCD switches. A dual DAC is used to generate the control voltages. The PIC16F873 chip reads all settings, calculates the temperature, humidity and the dew point in the humidifying chamber and controls the nebulizer, the fan and the hair dryer.

Figure 4: SHT71 humidity and temperature sensor Figure 5: Control box for the required temperature and humidity

Water tank

Nebulizer

Hair dryer output

(5)

4. VALIDATION

Evaporation:

The humidifying chamber reduces the evaporation rate of the liquid inside the sealed wells of the nanoarray. The wells are filled with the solution by means of a sliding motion of a Pyrex coverslip on top of the 55 array, which pushes the liquid into the wells. Subsequently, the edges of the coverslip are sealed by means of a colorless nail polish. Figure 6 depicts the progress curve of evaporation in a controlled and uncontrolled environment over a period of time for a 25 well array.

In an environment with no control of humidity, 50 percent of the wells were empty after 30 minutes. This time period was extended by a factor of eight, i.e. to four hours, by using feedback control to maintain 90% humidity. Preliminary experiments have been performed with bovine serum albumin (data not shown) as a coating for the nanoarray and the coverslip in order to reduce evaporation even further. This protein alters the surface properties of both materials in such a way that hydrophilicity is increased. The filling of the wells improves accordingly, preventing the formation of air bubbles and subsequent evaporation of the liquid.

Temperature Control:

For biological measurements it is important to work at controlled, constant temperatures. Most aggregation studies of-synuclein are performed at 37° C. Due to the poor conductance of the silicon nanoarrays it is necessary to know the actual temperature inside a well. We used a Sanwa DMM digital multimeter RD700 and K-AD sensor coupler (Sanwa Electric Instrument Ltd, Tokyo, Japan) in order to measure the temperature inside the wells of the nanoarray. Figure 7 shows the calibration curve of the reference temperature as given by the PE94 controller and the actual well temperature as measured by the multimeter. This curve indicates clearly that a reference temperature of 46° C corresponds to an actual temperature of 37° C inside the wells.

Figure 6: The number of evaporated wells as a function of time in an

uncontrolled and controlled humidity environment. 0 5 10 15 20 25 0 200 400 600 800 1000 1200 Tim e [m in] Wells

evaporated uncontrolled room humidity

controlled humidity 90%

Proc. of SPIE Vol. 5699 399

(6)

y = 0.8349x - 0.6596 R2= 0.9864 25 27 29 31 33 35 37 39 41 32 34 36 38 40 42 44 46 48 Reference Temperature [C] Actual Temperature [C]

Figure 7: Actual measured temperature inside a well of a nanoarray and its reference

temperature.

Fluorescence:

ThioT was measured in the free (unbound) state and bound to-synuclein fibrils in two independent experiments to examine the intensity as a function of 1) concentration and 2) integration time.

For the first experiment different concentrations of the native dye were used. The camera integration time was 80 ms and the humidity within the humidifying chamber was 90%. A Zeiss 01 filter set was used. All solutions were made with MilliQ distilled water, which also served as the control. Figure 8 shows the calibration curve for unbound thioflavin T. The minimum detectable signal is drawn as a horizontal dashed line and represents a signal level that is 3 above the background signal level of the blank (control) solution. The lower detection limit with the Köhler illumination configuration was 100µM thioflavin T.

y = 1286.4x + 14512 R2= 0.9975 10000 14000 18000 22000 26000 30000 0 1 2 3 4 5 6 7 8 9 10 Concentration [mM] Intensity [AU] 3 detection limit

Figure 8: The fluorescence intensity of free ThioT as a function of its concentration.

In a second experiment a solution of 10µm ThioT and 45 µm -synuclein fibrils (both final concentrations) was used in order to determine the relation of the intensity of the complex of ThioT and -synuclein fibrils as a function of the integration time of the CCD camera. The array was filled and sealed as explained above. The humidity was 95%. An Omega filter set XF114-2 (OmegaOptical Inc., Brattleboro, USA) was used in order to detect the shifted emission peak of 485 nm. The read out rate of the CCD camera was 1 MHz and the electronic gain was set to medium. The camera was cooled to40° C.

(7)

y = 6946.6x + 2079.7 R2= 0.9814 0 2000 4000 6000 8000 10000 12000 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1

Integration time [sec]

Intensity

[AU]

Figure 9: The fluorescence intensity of ThioT and-synuclein fibrils as a function of the

integration time The scale is not the same as that of Figure 8.

5. SUMMARY AND CONCLUSIONS

Our nanoarray system based on a conventional microscope is capable of measuring the weak fluorescence signals that we expect from the fluorophore ThioT. The minimum detection limit for free ThioT is 100 µM. For the detection of thioflavin T with-synuclein fibrils a minimum integration time of 80 ms should be applied. Evaporation of the solution in the wells is significantly reduced through the use of a self-made humidifying chamber. Due to this innovation approximately 50 % of the wells are still filled after a time period of four hours. In experiments that we have performed with nanoarrays coated with bovine serum albumin this time is increased even further. The temperature inside the wells of the nanoarray can be controlled within an accuracy of ±1° C. We are continuing to improve this nanoarray instrumentation so that the study of aggregation kinetics of-synuclein can be implemented in the miniaturized assay.

6. ACKNOWLEDGEMENTS

This work was partially supported by the Delft Inter-Faculty Research Centers LifeTech and Life Science & Technology and by the Bsik research program Cyttron.

7. REFERENCES

1. A. K. Dunker and Z. Obradovic, “The protein trinity – linking function and disorder”, Nat. Biotech., 19, pp. 805-806, 2001.

2. V. N. Uversky, “Why are “Natively Unfolded” Proteins Unstructured Under Physiologic Conditions?”, Proteins:

Structure, Function, and Genetics, 41, pp. 415-427, 2000.

3. A. Recchia, P. Debetto, A. Negro, D. Guidolin, S. D. Skaper, and P. Giusti, “-Synuclein and Parkinson’s disease”,

FASEB, 18, pp. 617-626, 2004.

4. W. Hoyer, D. I. Cherny, V. Subramaniam, and T. M. Jovin, "Impact of the acidic C-terminal region comprising amino acids 109-140 on a-synuclein in vitro." Biochemistry, 43, pp. 16233-16242, 2004.

5. C. W. Bertoncini, Y.-S. Jung, C. O. Fernández, W. Hoyer, C. Griesinger, T. M. Jovin, and M. Zweckstetter, "Release of long-range tertiary interactions potentiates aggregation of natively unstructured a-synuclein." Proc. Nat. Acad. Sci. USA, in press (2005).

6. H. LeVine III, “Stopped-Flow Kinetics Reveal Multiple Phases of Thioflavin T Binding to Alzheimer (1-40) Amyloid Fibrils”, Arch. Biochem. Biophys., 342:2, pp. 306-316, 1997.

7. H. LeVine III, “Quantification of-Sheet Amyloid Fibril Structures with Thioflavin T”, Methods in Enzymology, 309, pp. 274-284, 1999.

Proc. of SPIE Vol. 5699 401

(8)

8. H. LeVine III, "Thioflavine T interaction with synthetic Alzheimer’s disease -amyloid peptides: Detection of amyloid aggregation in solution", Protein Sci., 2, pp. 404-410, 1993.

9. I. T. Young, R. Moerman, L. R. van den Doel, V. Iordanov, A. Kroon, H. R. C. Dietrich, G. W. K. van Dedem, A. Bossche, B. L. Gray, L. Sarro, P. W. Verbeek, and L. J. van Vliet, “Monitoring enzymatic reactions in nanolitre wells”, J. Microsc-Oxfort, 212, pp. 254-263, 2003.

10. H. R. C. Dietrich, J. Knoll, L. R. van den Doel, G. W. K. van Dedem, P. A. S. Daran-Lapujade, L. J. van Vliet, R. Moerman, J. Pronk, and I. T. Young, “Nanoarrays: A Method for Performing Enzymatic Assays”, Anal. Chem., 76, pp. 4112-4117, 2004.

11. R. Moerman and G. W. K. van Dedem, “A cover slide method for controlled parallel sample introduction into arrays of (sub) nanoliter wells for quantitative analysis”, Anal. Chem., 75, 4132-4138, 2003.

12. R. Moerman, J. Knoll, C. Apetrei, L. R. van den Doel, and G. W. K. van Dedem, “Quantitative Analysis in Nanoliter Wells by Prefilling of Wells Using Electrospray Deposition Followed by Sample Introduction with a Coverslip Method”, Anal. Chem., publication in 2005.

13. I. T. Young, K. T. Hjelt, L. R. van den Doel, M. J. Vellekoop, and L. J. van Vliet, “Measuring liquid volumes in subnanoliter wells”, Biomedical Instrumentation Based on Micro- and Nanotechnology II (ed. By R. P. Mariella & d. V. Nicolau), 4265, pp. 75-80, SPIE, San Jose, CA, 2001.

14. L. R. van den Doel, Quantitative Microscopic Techniques for Monitoring Dynamic Processes in Microarrays, Ph.D. Thesis, Delft University of Technology, ISBN 90-77017-48-8, 2002.

15. L. R. van den Doel, M. E. Vellekoop, P. M. Sarro, S. Picioreanu, R. Moerman, J. Frank, G. van Dedem, K. T. Hjelt, L. J. van Vliet, and I. T. Young, “Fluorescence detection in (sub-) nanoliter microarrays”, Micro- and

Nanofabricated Structures and Devices for Biomedical and Environmental Applications II (ed. by M. Ferrari), 3606,

pp. 28-39, SPIE, San Jose, CA, 1999.

16. L. R. van den Doel, R. Moerman, G. van Dedem, I. T. Young, and L. J. van Vliet, “Monitoring enzyme-catalyzed reactions in micromachined nanoliter wells using a conventional microscope based microarray reader”, Novel

Micro- and Nanotechnologies for Bioengineering Applications (ed. By R. P. Mariella & C. J. Murphy), 4626, pp.

366-377, SPIE, San Jose, CA, 2002.

17. V. N. Uversky, E. M. Cooper, K. S. Bower, J. Li, and A. L. Fink, “Accelerated-synuclein fibrillation in crowded milieu”, FEBS Lett., 515, pp. 99-103, 2001.

18. W. Hoyer, T. Antony, D. Cherny, G. Heim, T. M. Jovin, and V. Subramaniam, “Dependence of -synuclein Aggregate Morphology on Solution Conditions”, J. Mol. Biol., 322, pp. 383-393, 2002.

Cytaty

Powiązane dokumenty

O n Borel sets and immeasurable functions in metric spaces.. In this note we give simple proofs of the theorems proved

Using Lipschitz- type hypotheses on the mth Fr´ echet derivative (m ≥ 2 an integer) instead of the first one, we provide sufficient convergence conditions for the inexact

Definition 4.2. Consider the Γ-semigroup S of Example 2.3. Let S be the set of all integers of the form 4n+1 and Γ be the set of all integers of the form 4n+3 where n is an integer.

We will get infor- mation from all the ideals, by means of Theorem 1.1 and noting that in fact an ideal is, in some sense, a divisor of its elements.. In this way, we give a new

In 1842 Dirichlet proved that for any real number ξ there exist infinitely many rational numbers p/q such that |ξ−p/q| < q −2.. This problem has not been solved except in

In case µ = 0 of the ordinary Hermite transform the product formula (27) of GHP is not correct in case of even degree n of the polynomials, since they are connected with the

In [1] and [2] we introduced a modification of classical Bernstein oper- ators in C([0, 1]) which we used to approximate the solutions of suitable parabolic problems.. In this paper

The model of generalized quons is described in an algebraic way as certain quasiparticle states with statistics determined by a commutation factor on an abelian group.. Quantization