• Nie Znaleziono Wyników

Implantation of additional defibrillation lead into the coronary sinus: an effective method of decreasing defibrillation threshold

N/A
N/A
Protected

Academic year: 2022

Share "Implantation of additional defibrillation lead into the coronary sinus: an effective method of decreasing defibrillation threshold"

Copied!
2
0
0

Pełen tekst

(1)

www.kardiologiapolska.pl

Elektrofizjologia inwazyjna/Invasive electrophysiology Kardiologia Polska 2011; 69, 12: 1308–1309 ISSN 0022–9032

Address for correspondence:

Address for correspondence:

Address for correspondence:

Address for correspondence:

Address for correspondence:

Aleksandra Czepiel, MD, PhD, Department of Cardiology, Postgraduate Medical School, Grochowski Hospital, ul. Grenadierow 51/59, 03–074 Warszawa, Poland, e-mail: aczepiel@kkcmkp.pl

Copyright © Polskie Towarzystwo Kardiologiczne

Implantation of additional defibrillation lead into the coronary sinus: an effective method of decreasing defibrillation threshold

Wszczepienie dodatkowej elektrody defibrylującej do zatoki wieńcowej:

skuteczna metoda obniżenia progu defibrylacji

Rajmund Wilczek

1

, Maciej Świątkowski

2

, Aleksandra Czepiel

2

, Maciej Sterliński

3

, Ewa Makowska

2

, Piotr Kułakowski

2

1Private Health Care Clinic “CardioPace”, Elblag, Poland

2Department of Cardiology, Postgraduate Medical School, Grochowski Hospital, Warsaw, Poland

3Ischaemic Heart Disease 2nd Department, Institute of Cardiology, Warsaw, Poland

A b s t r a c t

We report a case of successful implantation of an additional defibrillation lead into the coronary sinus due to high defibrilla- tion threshold (DFT) in a seriously ill patient with a history of extensive myocardial infarction referred for implantable cardio- verter-defibrillator implantation after an episode of unstable ventricular tachycardia. All previous attempts to reduce DFT, including subcutaneous electrode implantation, had been unsuccessful.

Key words: additional defibrillation lead, high defibrillation threshold

Kardiol Pol 2011; 69, 12: 1308–1309

INTRODUCTION

High defibrillation threshold (DFT) is a potentially life-thre- atening complication in patients with implantable cardio- verter-defibrillators (ICDs). Although less frequent in the era of high-output ICD devices, this problem still occurs in about 5% of ICD recipients [1].

CASE REPORT

A 54 year-old man was referred for ICD implantation after an episode of unstable ventricular tachycardia which had pre- viously been successfully terminated by external cardiover- sion. He had a history of extensive antero-lateral myocardial infarction, ischaemic stroke and renal insufficiency with a glo- merular filtration rate of 33–40 mL/min due to polycystic kid- ney disease. Echocardiography showed enlarged left ventricle with dyskinetic apex and ejection fraction of 15%. Coronary

angiography revealed an occlusion of the left anterior descen- ding artery and no significant changes in the remaining coro- nary vessels. The patient was selected for ICD implantation.

A dual chamber high-voltage ICD (Atlas II + DR, St. Jude Medical, Valley View Court Sylmar, CA, USA) with a single-coil active-fixation defibrillation lead (Durata 7122–65 cm, St. Jude Medical, Valley View Court Sylmar, CA, USA) positioned in the right ventricular (RV) apex and an atrial lead inserted in the right atrial appendage, was implanted. The defibrillation test perfor- med after implantation was unsatisfactory despite changing the biphasic defibrillating current polarity and duration of both pha- ses, and finally re-positioning the RV electrode.

During a second attempt, six days later, the single-coil RV electrode was replaced by a dual-coil lead (Durata 7120–

–65 cm, St. Jude Medical, Valley View Court Sylmar, CA, USA), but ventricular fibrillation could not be terminated even using

(2)

www.kardiologiapolska.pl

1309 Implantation of additional defibrillation lead into the coronary sinus

maximal shock energy. Following this, a subcutaneous elec- trode (SQ 6996–58 cm, Medtronic Inc, Minneapolis, MN, USA) was inserted, but DFT remained higher than device output. Re-positioning of the subcutaneous electrode did not result in an improvement of the DFT.

At this point, to avoid referring the patient for pericardial implantation of defibrillation patches, an attempt to reduce DFT was performed by the implantation of a defibrillation lead into the coronary sinus (CS), a tested but often over- looked procedure [1] which has been described in medical text- books [2]. This was performed via the insertion of a defibrilla- tion ‘floating’ lead into the CS (6937–110 cm, Medtronic Inc, Minneapolis, MN, USA) using ScoutPro 8 F (Biotronik, Berlin, Germany) delivery system (Fig. 1). In consequence, ventricu- lar fibrillation was terminated using 25 J, with a 11 J safety margin. The patient was successfully discharged, and one year follow-up confirmed stable CS lead position and did not re- veal any complications.

This case report underlines the importance of making all possible efforts to reduce DFT before opting for surgery. The implantation of a floating defibrillation lead into the CS is one of the options.

Conflict of interest: Dr Maciej Sterliński has received consul- ting fees from Biotronik, Medtronic, Sorin and St. Jude Medical.

The remaining authors have no conflicts of interest to disclose.

References

1. Swerdlow CD, Russo AM, Degroot PJ. The dilemma of ICD im- plant testing. PACE, 2007; 30: 675–700.

2. Ellenbogen KA, Kay GN, Lau C-P, Wilkoff BL. Clinical cardiac pacing, defibrillation, and resynchronization therapy. 3rd Ed.

Saunders Elsevier, Philadelphia 2007.

Figure 1.

Figure 1.

Figure 1.

Figure 1.

Figure 1. Lateral chest radiograph showing the ICD system after final implantation of additional floating defibrillation lead into the coronary sinus

Cytaty

Powiązane dokumenty

Po miesiącu od zakończenia antybiotykoterapii pacjent nie zgłaszał objawów niewydolności serca, a w kontrolnym echokardiograficznym badaniu przezprzełykowym uwidoczniono tętniak

Anna Posadzy-Małaczyńska, MD, PhD, Department of Hypertensiology, Angiology and Internal Diseases, Poznan University of Medical Sciences, ul. RCA with the fistulous connection to

Po doszczepieniu elektrody defibrylującej do vena azygos dochodzi do zmiany wektora fali defibrylacyjnej, która obejmuje teraz również ścianę tylno-boczną lewej

The chance of SR return was increased 11 times in patients receiving amiodarone, 6 times in patients with a double-coil lead, 2 times in patients with lower NYHA class, and 1.36

The use of high output devices significantly decreases the number of patients who required system modification in order to obtain an adequate defibrillation safety margin.. Key

Odejście lewej tętnicy wieńcowej (LTW) od bliższego odcinka prawej tętnicy wieńcowej (PTW) lub niezależnie, ale również z prawej zatoki Valsalvy, mo- że być przyczyną

Implantacja zapinki Occlutech Figulla ® u chorej z ubytkiem w przegrodzie międzyprzedsionkowej typu wtórnego.. Implantation of the Figulla Occlutech ® device in a patient with an

Opisany przypadek dotyczy pękniętego tętniaka zatoki Valsalvy z przeciekiem do RA u chorego z wrodzo- ną wadą serca obejmującą dwupłatkową zastawkę aor- talną i