Where is the lead? An uncommon twist in a defibrillator Twiddler syndrome

Gdzie jest elektroda? Nietypowy przypadek skręcenia elektrody w zespole Twiddlera Zainab Raissuni*, Marie Pierre Douchet*, Michel Chauvin*

University Hospitals of Strasbourg, Strasbourg, France

*These authors contributed equally to this work.

A 70-year-old woman had a single-chamber implantable cardioverter-defibrillator (ICD) inserted four years previously for ventricular sustained tachycardia caused by ischaemic heart disease. She was admitted for assessment of grade II dys- pnoea and muteness of the ICD. Chest X-ray showed that the generator had been rotated clockwise through almost 180°, and the lead displaced and fully wrapped around the defibrillator (Fig. 1A, B). During device extraction, the generator was mobile within the pocket, and the lead found to be twisted around itself multiple times between the generator and suture sleeve (Fig. 1C). We found also fibrosis connecting the loops of the lead (Fig. 1C, D) and several traces of inap- propriate shocks on the pulse generator (Fig. 1E) not felt by the patient. The device was safely removed and a new one was placed in a pocket and sutured tightly to the underlying muscle for stability. The patient denied having manipulated the generator box; the post implantation period was without event during 12 months of follow-up. Twiddler’s syndrome is well described in patients with pacemakers, but is unusual in patients with a defibrillator, because this device is larger and usually more difficult to move. The risk of Twiddler’s syndrome in patients with ICD is greater in obese, elderly patients who have abundant subcutaneous fat or loose connective tissue. Chest radiographs may allow the clinician to recognise this syndrome before a major fatal event.

Address for correspondence:

Zainab Raissuni, MD, Pôle d’Activité Médico-Chirurgicale Cardiovasculaire, Service de Cardiologie, Nouvel Hôpital Civil, 1, Place de l’Hôpital, 67000 Strasbourg, France, tel: + 33 3 69 55 08 21, fax: + 33 3 69 55 08 69, e-mail: zainab.raissouni@hotmail.com

Conflict of interest: none declared

Figure 1. A–E. Twiddler defibillator syndrome: from the diagnosis to the device extraction

A C E

B D

Cytaty

Powiązane dokumenty

Conclusions: Thyroid dysfunction, particularly sick euthyroid syndrome, was found to be related to in-hospital and long term mortality in patients with STEMI undergoing primary

The studied parameters included mHR and the coefficient of irregularity (CI), based on HM and the percentage of fast ventricular rates (tachy AF episodes defined as a heart rate of

In 1930 Mauriac described a syndrome characterized by the presence of growth impairment, hepatomegaly, delayed puberty and cushingoid features in patients with poorly controlled

or left ventricular wall; and the epicardial pacing lead was attached in the right ventricular apical region via left minithoracotomy..

In patients presenting with metastases of adenocarcinoma in axillary lymph nodes, and no evidence of primary cancer on physical and radiological examination, the most probable source

The former pacemaker unit was removed, but the two leads were left with distal tips localised in the right atrium and right ventricle and proximally cut off and secured in the area

The most im- portant factor associated with the high preterm birth rate in pregnant women with OSAS is the high prevalence of perinatal complications; moreover, there is a

Is there an association between the development of metabolic syndrome in PCOS patients and the C677T MTHFR gene polymorphism?. Czy istnieje związek między rozwojem