• Nie Znaleziono Wyników

Multipoint pacing of the left ventricle to achieve better resynchronisation and clinical response

N/A
N/A
Protected

Academic year: 2022

Share "Multipoint pacing of the left ventricle to achieve better resynchronisation and clinical response"

Copied!
1
0
0

Pełen tekst

(1)

www.kardiologiapolska.pl

Kardiologia Polska 2016; 74, 1: 84; DOI: 10.5603/KP.2016.0011 ISSN 0022–9032

STUDIUM PRZYPADKU / CLINICAL VIGNETTE

Address for correspondence:

Marcin Grabowski, MD, PhD, FESC, 1st Chair and Department of Cardiology, Medical University of Warsaw, ul. Banacha 1a, 02–097 Warszawa, Poland, tel: +48 22 599 19 58, fax: +48 22 599 19 57, e-mail: marcin.grabowski@wum.edu.pl

Conflict of interest: M. Grabowski: honoraria and grants from SJM Kardiologia Polska Copyright © Polskie Towarzystwo Kardiologiczne 2016

Multipoint pacing of the left ventricle to achieve better resynchronisation and clinical response

Wielopunktowa stymulacja lewej komory w celu osiągnięcia pełniejszej resynchronizacji i odpowiedzi na leczenie

Marcin Michalak

1

, Andrzej Cacko

1, 2

, Przemysław Stolarz

1

, Marcin Grabowski

1

11st Department of Cardiology, Medical University of Warsaw, Warsaw, Poland

2Department of Medical Informatics and Telemedicine, Medical University of Warsaw, Warsaw, Poland

A 65-year-old female with non-ischaemic cardiomyopathy, New York Heart Association III heart failure, and left bundle branch block was treated with cardiac resynchronisation therapy (CRT) of a defibrillator (CRT-D) Quadra Assura MP with a Quartet™

quadripolar left ventricular lead (St. Jude Medical, USA) (Fig. 1). The lead was placed into the lateral cardiac vein with its distal point located in the apical and proximal point in the mid-ventricular area. The defibrillator lead was implanted to the apex of the right ventricle. The device allows pacing of not only four different points, but also provides almost synchronous pacing of two points of the left ventricle (LV). We present how different settings of the CRT-D device affected the QRS morphology in 12-lead electrocardiogram (Fig. 2C–H). Synchronous biven-

tricular pacing (BiVp) with LV paced from the distal electrode (Fig. 2C) caused narrowed QRS (188 ms) with RS morphol- ogy in lead V1. Synchronous BiVp with LV paced from the Mid 2 and Mid 3 electrodes (Fig. 2D, E) gave further R wave amplitude increase in lead V1 and QRS shortening (160 ms and 140 ms, respectively). Synchronous BiVp with LV paced from the proximal electrode (Fig. 2F) was associated with the most dominant R wave in lead V1, but the QRS width (140 ms) was quite similar to the one observed during pacing with the Mid  3 electrode. As shown in Figure 2H, multipoint pacing caused no further QRS width reduction compared to pac- ing from the proximal electrode but gave the most dominant R wave in lead V1. It also led to maximal change in QRS am- plitude in lead V6 from +1.7 mV in native QRS to –1.7 mV, i.e. delta change of 3.4 mV. According to the unpublished subanalysis of the MADIT-CRT trial, higher delta change in QRS amplitude in lead V6 from baseline to post-implantation day was the best electrocardiographic predictor of reduction in heart failure or death during CRT pacing.

Figure 2. QRS morphology as a response to different configurations of biventricular pacing; A. Baseline electro- cardiogram; B. Right ventricle apical pacing; C. Synchronous biventricular pacing (BiVp) with left ventricle (LV) paced from distal electrode; D. Synchronous BiVp with LV paced from Mid 2 electrode; E. Synchronous BiVp with LV paced from Mid 3 electrode; F. Synchronous BiVp with LV paced from proximal electrode; G. BiVp with LV paced from proximal electrode with 50 ms pre-excitation; H. BiVp with multipoint LV pacing in configuration: proximal electrode — 10-ms delay — distal electrode — 40-ms delay — right ventricular lead

Figure 1. Distal end of the left ventricular quadripolar pa- cing lead; D — distal electrode; M2 — Mid 2 electrode;

M3 — Mid 3 electrode; P — proximal electrode

Cytaty

Powiązane dokumenty

titative variables (age, left ventricular ejection fraction, QRS before CRT system implantation, QRS after CRT system implantation, distance between the left and right

The MyoPore (Greatbatch Medical, New York, United States) sutureless screw -in epicardial pacing lead was delivered to the target area with the steerable FasTac Flex

Ventriculography showing the snail-like shape of left ventricle (LV) and focal akinesis of its anterior wall in systole (A) and LV in diastole (B). The same regional wall

Epicardial screw-in sensing lead on the left ventricle to treat undersensing of ventricular arrhythmias in a patient with arrhythmogenic.. right

D – Transthoracic echocardiography in apical four-chamber view before vascular surgery: the tumour mass is not clearly visible in the apex of the left ventricle. E – Com-

In 5 patients undifferentiated connective tissue dis- ease was diagnosed, and 14 consecutive patients were diagnosed with other connective tissue diseases (sclero- derma in 2

If sterile neutrinos exist, the mixing between sterile and active neutrino states must be small and the new massive neutrinos must be mostly sterile in order not to spoil the very

The control problem of the fed-batch fermentor for peni- cillin production was solved with the matrix-free inexact Newton method, presented in the article.. At first, the overall