Atrial septal defect occluder dislocation engaged through the tricuspid valve: surgical removal via right thoracotomy

N/A
N/A
Protected

Academic year: 2022

Share "Atrial septal defect occluder dislocation engaged through the tricuspid valve: surgical removal via right thoracotomy"

Copied!
1
0
0

Pełen tekst

(1)

www.kardiologiapolska.pl

Kardiologia Polska 2017; 75, 3: 279; DOI: 10.5603/KP.2017.0052 ISSN 0022–9032

Studium przypadku / CliniCal Vignette

Address for correspondence:

Marzia Cottini, MD, Department of Heart and Vessels, Cardiac Surgery Unit, “S. Camillo-Forlanini” Hospital, Circonvallazione Gianicolense 87, 00149, Rome, Italy, tel: +39-347-3245331, fax +39-06-58704511, e-mail: marzia.cottini@hotmail.it

Conflict of interest: none declared

Kardiologia Polska Copyright © Polskie Towarzystwo Kardiologiczne 2017

Atrial septal defect occluder dislocation engaged through the tricuspid valve:

surgical removal via right thoracotomy

Przemieszczenie zapinki Amplatza zamykającej ubytek przegrody

międzyprzedsionkowej poprzez zastawkę trójdzielną: chirurgiczne usunięcie za pomocą prawostronnej torakotomii

Marzia Cottini, Amedeo Pergolini, Francesco Musumeci

Department of Heart and Vessels, Cardiac Surgery Unit and Heart Transplantation Centre, “S. Camillo-Forlanini” Hospital, Rome, Italy The increasingly widespread use of atrial septal defect (ASD) transcatheter occlusion devices is associated with the identification of new and uncommon complications. We present the case of a 29-year-old female admitted for per- cutaneous closure of ASD. During device implantation, the transoesophageal echocardiogram revealed dislocation of the Amplatzer into the right atrium and straddling the tricuspid valve (Fig. 1A, B, Supplementary Movie Clip 1 — see journal website). The ASD occluder device engaged through the tricuspid valve, producing severe tricuspid regurgitation (Fig. 1C, D). The patient was referred for surgery to remove the device and to repair tricuspid regurgitation (Fig. 2). The percutaneous inter-atrial communication occluder is an internationally certificated device that is being increasingly used.

The most common periprocedural complications are cerebro-ischaemic event (transitory ischaemic attack, stroke, etc.), residual shunting, and device dislocation. We reported a case of original device dislocation: it had been lacerating the ASD borders, then migrating through the right atrium to the tricuspid valve, engaging into the tricuspid valve orifice, and finally causing severe tricuspid regurgitation. In scientific literature, reports of ASD occluder dislocation are presented, but device dislocation into the tricuspid valve is extremely rare and uncommon.

Figure 1. A. Transoesophageal echocardiogram (TEE) medi- oesophageal four-chamber view: atrial septal defect occluder dislocated into the right atrium and right ventricle through the tricuspid valve; B. TEE medioesophageal 50-degree view:

inter-atrial defect and the dislocated Amplatzer; C. TEE view of the engaging of Amplatzer occluder through the tricuspid valve; D. TEE view of tricuspid valve regurgitation due to iatrogenic damage of posterior leaflet

Figure 2. Transoesophageal echocardiogram medioeso- phageal four-chamber view: successful removal of dislocated device, closure of atrial septal defect with a pericardial patch and repair of posterior tricuspid leaflet with right thoraco- tomy approach; LA — left atrium; LV — left ventricle;

RA — right atrium; RV — right ventricle; TV — tricuspid valve

A B

C D

Cytaty

Powiązane dokumenty

According to our assumptions, early closure of ASD II in small children should result, together with a child’s growth, in a decrease of the implant diameter and septum length

There was a small defect of the membranous inter- ventricular septum with left-to-right shunt and maximum left ventricle to right ventricle gradient of 110 mm Hg1. In the atrial

Transthoracic four-chamber view (A) showing interatrial septal aneurysm (arrow) extending to the right ventricular inflow and left-to-right shunt (dotted arrow) with color Doppler

Subsequent transesophageal echocardiography, after failure of closure with two sepa- rate closure devices, showed another defect and an ongoing left to right shunt.. During

In 29 children in whom 24-hour Holter ECG monitoring was performed prior to the procedure, sinus rhythm was dominant in 28 children with in- termittent junctional rhythm in 2

The transesophageal echocardiogram showed a 17 × 10 mm ostium secundum ASD with a left-to-right shunt, an atrial septal aneurysm protruding into the right atrium, a deficiency of

D – the FASD, a self ‑expanding nitinol wire mesh device with fenestration (fenestration diameter, 6 mm; arrow); e, f – final transesophageal echocardiography visualizing

Right heart catheterization indicated precapillary pul- monary hypertension (PH), with a mean pulmo- nary artery pressure (mPAP) of 32 mm Hg, pulmo- nary artery wedge pressure