Transvenous extraction of a broken atrial lead embolised into the pulmonary artery using a pigtail catheter

Przezskórne usunięcie złamanej elektrody przedsionkowej migrującej do krążenia płucnego za pomocą cewnika pigtail

Marcin Michalak

1

, Andrzej Kutarski

2

, Maria Zawadzka-Byśko

1

, Grzegorz Opolski

1

, Marcin Grabowski

1

11st Chair and Department of Cardiology, Medical University of Warsaw, Warsaw, Poland

2Department of Cardiology, Medical University of Lublin, Lublin, Poland

A 57-year-old man with an AAI pacemaker was referred to the hospital due to suspicion of an atrial lead failure. The pace- maker was implanted 9 years earlier because of sick sinus syndrome. The atrial lead was inserted by puncture of the right subclavian vein with the pocket located at the same site. Currently the patient presented with mild asymptomatic brady- cardia of 50 bpm. For 3 months he had noticed pectoral muscle contractions that depended on body position. Pacemaker control revealed atrial lead impedance exceeding 2000 Ohms and lack of either proper sensing or stimulation. Chest X-ray showed a lead breakage at the level of the right clavicle and migration of the proximal fragment of the broken lead to the left pulmonary arteries (Fig. 1). There were no significant abnormalities in echocardiogram and the patient was referred for elective transvenous lead extraction. The procedure of transvenous extraction of the

broken atrial lead was conducted under general anaesthesia and consisted of the fol- lowing steps. First, after percutaneous puncture of the left subclavian vein, a coronary sinus (CS) cannulation catheter (Medtronic Attain Command®, USA) was inserted to the level of the right ventricle. Then, the pigtail catheter (Abbot Laboratories, USA) was advanced and positioned in parallel to the atrial lead in the pulmonary trunk.

The lead was caught with the pigtail catheter by rotating it (Fig. 2A) and pulling it down (Fig. 2B) to the vena cava superior. The lead was grabbed with the lasso wire (Fig. 2C) and pulled into the CS catheter. A Green inner (10.0/12.1-French) Byrd dila- tor sheath (Cook Medical, Germany) was glided down over the CS catheter (Fig. 2D), and the lead was extracted completely. The procedure, as well as the postprocedural period, was uncomplicated. The patient was discharged home in good condition on the third day after the procedure. The pulse generator and remaining proximal part of the atrial lead removal was planned as a next step procedure. Complete breakdown and migration of the lead is a rare complication of electrotherapy. The presence of a broken lead in the pulmonary arteries, although asymptomatic at the beginning, may end with perforation and, in the case of invective endocarditis when left in place, is a life-threatening and incurable

condition. In situations like this, early transvenous lead extraction before any complications appear is the management of choice. Procedures with extraction of embolised leads are very difficult and should be performed in experienced centres with cardiac surgery backup. The situation is more complicated because there are no tools dedicated to that kind procedure. We present how typical instruments available in haemodynamic and electrophysiology laboratories, such as pigtail or CS catheters, may be used to achieve a successful and safe result in

such a complex lead extraction. Figure 2. A–D. Following steps of percutaneous atrial lead extraction

Figure 1. Chest X-ray showing bro- ken atrial lead that migrated into the pulmonary bed

A B C D

Cytaty

Powiązane dokumenty

The extraction of three implanted (18-, 16-, and ten year-old) damaged nonfunctional leads was complicated by a lead breaking and losing its external silicone tube.. The missing part

In conclusion, our case calls attention to the prospect of right to left shunt from isolated P-AVF as a potential source of paradoxical embolism caus- ing stroke.. Most

In the last stage the flattened Dotter’s basket was introduced to the heart inside the Byrd dilator and then fastened to the described lead as the external ‘splint’.. This solution

Given the above, we sought to characterize the pop- ulation and examine the outcomes, including risk factors for in-hospital complications and 12-month mortality and

Figure 2 Monitoring of the process of pulling on the cardiac walls during lead dissection: A ‒ fluoroscopy showing  the extraction of a high ‑voltage lead, adhesion of

The MyoPore (Greatbatch Medical, New York, United States) sutureless screw -in epicardial pacing lead was delivered to the target area with the steerable FasTac Flex

Intravascular techniques for lead extraction including locking stylets and telescoping sheaths significantly reduce the risk of electrode rupture, but helix fracture and

Com- puted tomography angiography depicted the right aortic arch (RAA) with isolated left subclavian artery (ILSA) originating from the left-sided patent ductus arteriosus