KARDIOLOGIA POLSKA 2019; 77 (11) 1084

the Heart Team decided to refer the patient for transcatheter aortic valve implantation.

The procedure was done under general anes- thesia with puncture of femoral arteries and preclosure technique. The 23-mm Lotus Edge valve (Boston Scientific, Marlborough, Massa- chusetts, United States) was successfully im- planted over Safari guidewire (Boston Scien- tific) under transesophageal echocardiography and angio control to avoid interaction between an implanted transcatheter aortic valve and a mitral valve prosthesis. After implantation, the peak pressure gradient decreased to 5 mm Hg, and diastolic aortic pressure was 65 mm Hg.

No paravalvular leak was detected (Figure 1C–1F).

The Lotus valve has been examined in several studies (REPRISE I, II -IIE, RESPOND) with re- markably low rate of paravalvular leaks (PVLs) (about 0.6% of patients displaying moderate to severe PVL); however, there was a relative- ly high need for permanent pacemaker im- plantation (24%–38% of pacemaker -naive pa- tients).1 More recently, the REPRISE III study has shown that Lotus valve was noninferior to self -expanding CoreValve / Evolut R (Medtron- ic, Minneapolis, Minnesota, United States).2 The previous Lotus valve system was recalled from the market due to problems with the device’s locking system.

The new Lotus Edge valve system consists, similarly to the previous version, of a braided single -wire nitinol frame with 3 bovine pericar- dial leaflets and features an adaptive polymer membrane seal at the lower half to reduce PVL.

It is available in 3 sizes (23, 25, and 27 mm) and will also be obtainable in 21 and 29 mm in the near future.3 The Lotus Edge delivery catheter We present data on the first implantation

of the Lotus Edge transcatheter aortic valve (Figure 1A and 1B) in Poland and other Central and Eastern European countries performed in our hospital on June 26, 2019.

A 78-year -old woman with a history of se- vere aortic stenosis after previous cardiac sur- gery in 2014 (biological mitral prosthesis and tricuspid valve ring implantation) complicated by third -degree atrioventricular block requiring implantation of VVI type pacemaker, with arte- rial hypertension, mild renal insufficiency, per- manent atrial fibrillation (CHA2DS2-VASc score, 4 points; HAS -BLED score, 2 points) presented with dyspnea (New York Heart Association class III; N-terminal fragment of the prohormone brain natriuretic peptide, 2632 pg/ml) and chest pain. Coronary angiography was normal. Echo- cardiography revealed good ejection fraction with severe aortic stenosis and moderate aortic insufficiency (maximal aortic gradient, 82 mm Hg; mean aortic gradient, 53 mm Hg; aortic valve area, 0.6 cm2), left ventricular hypertro- phy, proper function of mitral valve prosthesis, and mild tricuspid insufficiency. Multislice com- puted tomography showed a moderately calci- fied tricuspid aortic valve with calcium extend- ing into left ventricular outflow tract, aortic an- nulus perimeter of 65.1 mm and area of 329.1 mm2, distance to coronary arteries was low, 8.9 mm on the left side, and quite good femoral ac- cess, although some calcifications were pres- ent at the puncture site. After considering the patient’s past medical history and risk scores (EUROSCORE II [ European System for Cardiac Operative Risk Evaluation II], 5.17%; STS [the Society of Thoracic Surgeons score], 5.324%),

Correspondence to:

Assoc. Prof. Marek grygier,  MD, PhD, 1st Department  of Cardiology, Poznan  university of Medical Sciences,  ul. Długa 1/2, 61-848 Poznań,  Poland, phone: +48 61 854 92 23,  email: marek.grygier@skpp.edu.pl Received: July 23, 2019.

Revision accepted:

August 30, 2019.

Published online:

August 30, 2019.

Kardiol Pol. 2019; 

77 (11): 1084-1086 doi:10.33963/KP.14957 Copyright by the Author(s), 2019

C L I N I C A L V I G N E T T E

First implantation of the new Lotus Edge transcatheter aortic valve in Poland

Marek Grygier1, Marcin Misterski2, Aleksander Araszkiewicz1, Anna Olasińska ‑Wiśniewska1, Anna Komosa1, Marek Jemielity2, Maciej Lesiak1

1  1st Department of Cardiology, Poznan university of Medical Sciences, Poznań, Poland

2  Department of Cardiac Surgery and Transplantology, Poznan university of Medical Sciences, Poznań, Poland

(2)

C L I N I C A L V I G N E T T E New Lotus Edge valve in Poland 1085 is locked after reaching the final position. An- other new feature is the Depth Guard—modifi- cation to the delivery system with the aim to re- duce the need for new pacemaker implantation (in RESPOND Extension, a significant reduction to 17.8% was observed).3

has increased flexibility with an optimized pre - -shape curve and reduced (by 3 French) proxi- mal catheter profile. Guided by a marker, the device is mechanically expanded in the desired position and allows for full recapture and repo- sition after evaluation of the initial result and

Figure 1 A – Lotus Edge valve; B – Lotus Edge valve attached to the delivery system; C – angiogram from the procedure showing the valve just before the final locking; D – angiogram from the procedure showing the valve after release; E – angiogram at the end of the procedure—no leak around the device; F – final transesophageal echocardiography showing good function of the transcatheter aortic valve and mitral valve bioprosthesis; A and B – reproduced with permission from Boston Scientific (Marlborough, United States)

A

D C

B

E F

(3)

KARDIOLOGIA POLSKA 2019; 77 (11) 1086

ArtiClE inFormAtion

ConFliCt oF intErEst Mg and Boston Scientific concluded a proctoring  agreement. Mg is a Boston Scientific Advisory Board Member and receives hono- raria for lectures. Other authors do not report any conflict of interest.

opEn ACCEss This is an Open Access article distributed under the terms  of  the  Creative  Commons  Attribution  -NonCommercial  -NoDerivatives  4.0  in- ternational License (CC BY  -NC  -ND 4.0), allowing third parties to download ar- ticles and share them with others, provided the original work is properly cited,  not changed in any way, distributed under the same license, and used for non- commercial purposes only. For commercial use, please contact the journal office  at kardiologiapolska@ptkardio.pl.

How to CitE grygier M, Misterski M, Araszkiewicz A, et al. First implantation  of the new Lotus edge transcatheter aortic valve in Poland. Kardiol Pol. 2019; 77: 

1084-1086. doi:10.33963/KP.14957

rEFErEnCEs

1 Falk V, Wohrle J, Hildick  -Smith D, et al. Safety and efficacy of a repositionable  and fully retrievable aortic valve used in routine clinical practice: the reSPOND  Study. eur Heart. J 2017; 38: 3359-3366.

2 Feldman Te, reardon MJ, rajagopal V, et al. effect of mechanically expanded  vs self- expanding transcatheter aortic valve replacement on mortality and major  adverse clinical events in high  -risk patients with aortic stenosis: The rePriSe iii  randomized Clinical Trial. Jama. 2018; 319: 27-37.

3 Solomonica A, Choudhury T, Bagur r. The mechanically expandable LOTuS  valve and LOTuS edge transcatheter aortic valve systems. exp rev Med Dev. 2018; 

15: 763-769.

Cytaty

Powiązane dokumenty

figure 1 A – multislice contrast computed tomography of implanted bioprosthesis and the ascending aorta, the short axis view; B – multislice contrast computed tomography of

figure 1 A–D – baseline computed tomography showing a calcified nodule (arrows); e–L –postprocedural assessment by means of angiography (e; black arrows indicate the stent

However, a higher risk of THVT may be expected in patients with chronic obstructive pulmonary disease, a smaller aortic valve area, a higher mean postprocedural

tion in mean (SD) transvalvular gradients after the placement of transcatheter implanted valve in surgical implanted valve of 19.0 (8.8) mm Hg (P <0.001) and a further mean

short ‑axis view; C – echocardiography imaging of implanted valvular prostheses after severe paravalvular regurgitation: long ‑axis view; D – removed infected valvular

Prosthetic valve endocarditis and acute heart failure in a patient after transcatheter aortic valve implantation procedure.. Tomasz Skowerski, Patryk Grzywocz, Mariusz Bałys,

— higher graded AV-conduction abnormalities following TAVR implantation can be mark- edly reduced by exact valve sizing and a high implantation technique relative to the aortic

Early regression of left ventricular mass associated with diastolic improvement after transcatheter aortic valve implantation.. Levy D, Garrison RJ, Savage DD, Kannel WB,