Intentional fracture of the bioprosthetic valve ring during transcatheter aortic valve implantation

N/A
N/A
Protected

Academic year: 2022

Share "Intentional fracture of the bioprosthetic valve ring during transcatheter aortic valve implantation"

Copied!
2
0
0

Pełen tekst

(1)

KARDIOLOGIA POLSKA 2020; 78 (4) 352

the coronary artery ostia, and guidewires load‑

ed with stents ready for implantation were po‑

sitioned for backup in both coronaries (FIGURE 1A).

Second, a 23‑mm Medtronic CoreValve Evolut R (Medtronic Inc.) was directly implanted using the right femoral access (FIGURE 1B). At this stage, echocardiography revealed a decrease in pres‑

sure gradients down to maximum (mean) of 23 (11) mm Hg, and reduction in peak velocity down to 2.37 m/s. Subsequently, a 20‑mm Atlas Gold PTA (BARD Canada Inc., Oakville, Ontar‑

io, Canada) dilatation catheter was positioned into both valves. Under rapid ventricular pac‑

ing up to 180 bpm lasting 9 seconds, it was di‑

lated with pressure of 20 atm until the ring of surgical valve cracked (FIGURE 1D and 1E). The mo‑

ment of the fracture was noticeable in the fluo‑

roscopy as a visible release of the balloon waist and confirmed by sudden drop in inflation pres‑

sure on the manometer.

After the completion of the procedure, further improvement of echocardiographic parameters was noticed: a maximum (mean) pressure gra‑

dient of 17 (8) mm Hg and a reduction in peak velocity to 2.1 m/s (FIGURE F). No aortic regurgita‑

tion, paravalvular leak, arrhythmias, ischemia, or pericardial effusion were observed. There was no need for stent deployment (FIGURE C). The patient was transferred to the postoperative care unit in good condition and was discharged 5 days after the procedure with notable improvement of her symptoms to NYHA class I.

There has been a limited number of cases of valve ‑in ‑valve transcatheter aortic valve replace‑

ment with BVF carried out in the world. The pro‑

cedure is strictly connected with the type of An 82‑year‑old woman underwent a minimally

invasive surgical implantation of a bioprosthetic aortic valve (19‑mm Medtronic Mosaic biopros‑

thesis, Medtronic Inc., Minneapolis, Minnesota, United States) in our center in 2016. Since 2018, progressive dyspnea to New York Heart Associ‑

ation (NYHA) class III was observed, accompa‑

nied by an increase in mean pressure gradient postoperatively from 23 to 38 mm Hg and peak transvalvular pressure gradient to 75 mm Hg as well as reduction in the indexed effective orifice area to 0.41 cm2/m2.

Because of a significant risk of death dur‑

ing reoperation (logistic regression version of the European System for Cardiac Operative Risk Evaluation [EuroSCORE], 31.16%; Society of Tho‑

racic Surgeons [STS] risk score, 6.2%), the patient was disqualified from a surgical procedure and referred for transcatheter valve ‑in ‑valve replace‑

ment. Due to the very small size of the surgical bioprosthesis, implantation of a new valve might not guarantee the desired outcome because of further decrease in the orifice area. Therefore, a decision was made to perform controlled ring fracture (bioprosthesis valve fracture [BVF]).

The distance between the coronary ostia and the surgically implanted bioprosthesis valve, es‑

timated by computed tomography angiography to be 2 mm, was an additional risk factor. Con‑

sidering that after cracking the ring, this param‑

eter may be further reduced, it was decided to utilize coronary protection technique.

The procedure was performed by a team of cardiac surgeons and interventional cardiol‑

ogists. First, using the left radial and femoral approaches, guiding catheters were placed in

Correspondence to:

Marek Mak, MD, PhD, Department of Cardiac Surgery, 4th Military Hospital, ul. Weigla 5, 50-981 Wrocław, Poland, phone: + 48 26 166 08 26, email: aureliuszm@tlen.pl Received: October 31, 2019.

Revision accepted:

February 14, 2020.

Published online:

February 21, 2020.

Kardiol Pol. 2020; 78 (4): 352-353 doi:10.33963/KP.15208 Copyright by the Author(s), 2020

C L I N I C A L V I G N E T T E

Intentional fracture of the bioprosthetic valve ring during transcatheter aortic valve

implantation

Marek Mak1, Filip Klausa1, Artur Telichowski2, Agnieszka Wysokińska ‑Kordybach2, Jacek Skiba1, Waldemar Banasiak2 1 Department of Cardiac Surgery, 4th Military Hospital, Wrocław, Poland

2 Department of Cardiology, 4th Military Hospital, Wrocław, Poland

(2)

C L I N I C A L V I G N E T T E Intentional BVF during TAVI 353 REFERENCES

1  Allen KB, Chhatriwalla AK, Saxon JT, et al. Bioprosthetic valve fracture: tech- nical insights from a multicenter study. J Thorac Cardiovasc Surg. 2019; 158:

1317-1328.e.1.

2  Olasińska -Wiśniewska A, Grygier M, Araszkiewicz A, et al. Transcatheter aortic valve implantation in degenerated aortic bioprosthesis complicated by a “frozen”

leaflet. Kardiol Pol. 2019; 77: 1089-1091.

3  Chhatriwalla AK, Allen KB, Saxon JT, et al. Bioprosthesis valve fracture im- proves the hemodynamic results of valve -in -valve transcatheter aortic valve re- placement. Circ Cardiovasc Interv. 2017; 10.

4  Allen KB, Chhatriwalla AK, Cohen JD, et al. Bioprosthesis valve fracture to fa- cilitate transcatheter valve -in -valve implantation. Ann Thorac Surg. 2017; 104:

1501-1508.

the previously used bioprosthesis.1,2 Especial‑

ly bioprosthetic valves of small size result in high residual transvalvular gradients that have been associated with increased mortality.3 Pro‑

cedures with BVF that are described in the liter‑

ature are associated with 2.6% 30‑day mortali‑

ty. Postprocedural mean (SD) transvalvular gra‑

dient was 9.2 (6.3) mm Hg (P <0.001).1 A multi‑

center study of 70 procedures revealed a reduc‑

tion in mean (SD) transvalvular gradients after the placement of transcatheter implanted valve in surgical implanted valve of 19.0 (8.8) mm Hg (P <0.001) and a further mean (SD) gradient re‑

duction after BVF of 8.1 (4.8) mm Hg (P <0.001) as well as an increase in mean (SD) effective orifice surface of 1.4 (0.8) cm2 (P <0.001) and a further mean (SD) increase after BVF of 2.1 (0.8) cm2 (P <0.001).4

Intentional fracture of bioprosthetic valve ring during valve ‑in ‑valve TAVI is a procedure with low mortality risk which may be considered in a patient with a very small surgically impant‑

ed bioprosthesis to achieve the lowest transval‑

vular gradient.

ARTICLE INFORMATION

CONFLICT OF INTEREST  None declared.

OPEN ACCESS  This is an Open Access article distributed under the terms of the Creative Common s Attribution -NonCommercial -NoDerivatives 4.0 In- ternational License (CC BY -NC -ND 4.0), allowing third parties to download ar- ticles and share them with others, provided the original work is properly cited, not changed in any way, distributed under the same license, and used for non- commercial purposes only. For commercial use, please contact the journal office at kardiologiapolska@ptkardio.pl.

HOW TO CITE  Mak M, Klausa F, Telichowski A, et al. Intentional fracture of the bioprosthetic valve ring during transcatheter aortic valve implantation. Kardi- ol Pol. 2020; 78: 352-353. doi:10.33963/KP.15208

FIGURE 1 A – the anatomy of the coronary arteries with guidewires. Bioprosthesis with new Evolut valve inside before (B) and after (C) ring fracture. The moment before (D) and after (E) fracture of the valve ring. F – ultrasonography results directly after the procedure

A

D E F

B C

Cytaty

Powiązane dokumenty

This echocardiographic evaluation performed in a real-world setting in European patients with severe AS who received a transcatheter SAPIEN 3 demonstrated good valve performance,

For patients with degenerated bioprosthetic aortic valves, transcatheter aortic valve implantation (TAVI) represents a less invasive option.. Herein reported are valve-in-valve

1) Sustained hemodynamic changes are observed as peak velocity and transvalvular gradient dif- ferences between discharge and follow-up within 1 year after the procedure; 2) BAV

figure 1 A – multislice contrast computed tomography of implanted bioprosthesis and the ascending aorta, the short axis view; B – multislice contrast computed tomography of

ence of &gt;48° angle between the aortic annulus Transcatheter aortic valve implantation (TAVI)4. is a well ‑established modern treatment of severe symptomatic

However, a higher risk of THVT may be expected in patients with chronic obstructive pulmonary disease, a smaller aortic valve area, a higher mean postprocedural

short ‑axis view; C – echocardiography imaging of implanted valvular prostheses after severe paravalvular regurgitation: long ‑axis view; D – removed infected valvular

Figure 1 A – Lotus Edge valve; B – Lotus Edge valve attached to the delivery system; C – angiogram from the procedure showing the valve just before the final locking; D –