Kardiologia Polska 2017; 75, 12: 1355; DOI: 10.5603/KP.2017.0235 ISSN 0022–9032

STUDIUM PRZYPADKU / CLINICAL VIGNETTE

Address for correspondence:

Monika Gawałko, 1st Chair and Department of Cardiology, Medical University of Warsaw, ul. Banacha 1a, 02–097 Warszawa, Poland, e-mail: mongawalko@gmail.com Conflict of interest: M. Grabowski, M. Michalak, A. Cacko, G. Opolski — investigator’s fees from Zoll

Kardiologia Polska Copyright © Polskie Towarzystwo Kardiologiczne 2017

Ventricular tachycardia successfully treated with wearable cardioverter-defibrillator

Częstoskurcz komorowy skutecznie przerwany przez zewnętrzny kardiowerter-defibrylator noszony przez pacjenta

Marcin Grabowski

1

, Monika Gawałko

1

, Marcin Michalak

1

, Andrzej Cacko

1, 2

, Grzegorz Opolski

1

11st Chair and Department of Cardiology, Medical University of Warsaw, Warsaw, Poland

2Department of Medical Informatics and Telemedicine, Medical University of Warsaw, Warsaw, Poland

A 68-year-old woman with diabetes, paroxysmal atrial fibrillation (PAF), and chronic kidney disease stage 3 was admitted to hospital because of acute lower limb ischaemia, requiring emergency surgical embolectomy of common iliac arteries with a Fogarty balloon catheter. Unfortunately, the procedure was complicated by ST-elevated myocardial infarction of the left ventricle posterolateral wall, subsequently treated with marginal branch angioplasty with drug-eluting stent implantation. In addition, the patient’s general condition deteriorated due to cardiogenic pulmonary oedema, which required intermittent mechanical ventilation with positive end-expiratory pres- sure support. Post-procedural echocardiography revealed decreased left ventricular ejection fraction (LVEF 35%) and severe mitral regurgitation. For the prevention of sudden cardiac death teh patient was equipped with wearable car- dioverter defibrillator (WCD; LifeVest® by Zoll) as a bridge therapy until the final decision regarding the need for an implantable cardioverter-defibrillator (ICD). Ventricular tachycardia (VT) and ventricular fibrillation zones were set as presented in Figure 1. The patient was educated and trained in WCD and discharged from the hospital. The patient had been wearing a LifeVest® for 16 days (23 h and 26 min per day on average) when, during a PAF with average heart rate of 175 bpm, the WCD recognised VT and alarmed the pa- tient (Fig. 1). She stopped therapy by simultaneously pressing the abort bottoms (Fig. 1, blue arrow). About 30 s later VT (about 195 bpm) was initiated, probably as a result of the patient’s decompensation because of fast heart rate (Fig. 1, green arrow). She aborted therapy five times until a shock (energy of 150 J) was delivered after 92 s of VT (Fig. 1, red arrow). The therapy was successful, but about 25 s later another VT was initiated. Then she aborted therapy three times, but because of severe symptoms allowed another shock within about 50 s. The therapy was successful. WCD is recommended as a bridge therapy for patients with high risk of sudden cardiac death due to potentially reversible causes and waiting for qualification to ICD. According to registries less than 2% of patients equipped with a WCD receive therapy. LifeVest® allows patients to abort therapy until losing consciousness or other severe symptoms. Patients should be trained not to take the WCD off after the shock because arrhythmia may recur. LifeVest® may deliver up to twelve therapies when its battery is fully charged.

Figure 1. Electrocardiogram recording of channel ‘SS’ in wearable cardioverter defibrillator during the shock episode.

The shock sequence was canceled by the patient for the first time 7 s after ventricular tachycardia (VT) detection (blue arrow). After 30 s later VT was initiated (green arrow). Patient aborted therapy five times until a shock (energy of 150 J) was delivered after 92 s of VT (red arrow)

Cytaty

Powiązane dokumenty

During the mean follow-up of 14 ± 13 months we observed the appropriate function of the systems and a ventricular fibrillation episode successfully terminated by the device in one

Cardiac magnetic resonance imaging confirmed the presence of hypertrabeculation with a two-layered structure of the endocardium with an increased noncompacted to compacted

Układy podskórnych kardiowerterów-defibrylatorów są istotną pod względem medycznym opcją terapeutyczną dla chorych obarczonych ryzykiem wystąpienia nagłego zgonu

Electrical signal recorded by automated external defibrillator during resuscitation of pre-hospital

Implantable cardioverter-defibrillator placement via a single persistent left superior vena cava in secondary prevention of sudden cardiac death in a patient with Turner

a report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee

The aim of this study was to determine whether MTWA testing could be useful in establishing the order of ICD im- plantation procedures in a group of patients with LV systolic

RIGHT VENTRICULAR PACING AND THE OCCURRENCE OF VT/VF EPISODES The possibility of harmful effects of RV pacing in patients with reduced LV systolic function with an ICD was