• Nie Znaleziono Wyników

Irregular ventricular tachycardia underdetected by implantable cardioverter defibrillator device

N/A
N/A
Protected

Academic year: 2022

Share "Irregular ventricular tachycardia underdetected by implantable cardioverter defibrillator device"

Copied!
3
0
0

Pełen tekst

(1)

www.cardiologyjournal.org 281 CASE REPORT

Cardiology Journal 2008, Vol. 15, No. 3, pp. 281–283 Copyright © 2008 Via Medica ISSN 1897–5593

Address for correspondence: Konstantinos G. Kouvelas, MD, Hellenic Red Cross General Hospital, Erythrou Stavrou 1

& Athanasaki str., 11526 Athens, Greece, tel: +30 210-6414 705, 210-8104 554, fax: +30 210-641 4587, e-mail: drkokouv@gmail.com

Received: 24.01.2008 Accepted: 16.03.2008

Irregular ventricular tachycardia underdetected by implantable cardioverter defibrillator device

Athanasios G. Manolis, Spyros Kourouklis, Dimitrios Chatzis, Konstantinos Kouvelas and Zenon Kyriakides

2nd Department of Cardiology. Hellenic Red Cross General Hospital, Athens, Greece

Abstract

A case of sustained monomorphic ventricular tachycardia underdetected by a single chamber implantable cardioverter defibrillator because of RR interval irregularity is presented. The programmed stability criterion is responsible for the underdetection. Special attention must be paid when it comes to programming this detection parameter. (Cardiol J 2008; 15: 281–283) Key words: irregular ventricular tachycardia, implantable cardioverter defibrillator, amiodarone

Case report

Ischemic heart failure has been associated with the increased risk of lethal ventricular tachyarrhy- thmias. Implantable cardioverter defibrillator (ICD) is an established treatment, for the primary and/or the secondary prevention of sudden arrhythmic death in these patients. In ICD detection algori- thms, the stability criterion of consecutive RR in- tervals should be programmed very carefully for the prevention of inappropriately delivered therapies by the device, especially in patients with concomitant supraventricular irregular tachyarrhythmias.

We present a patient with ischemic heart fa- ilure, left bundle branch block in the basal surface ECG and known paroxysmal atrial fibrillation un- der amiodarone therapy. A single chamber ICD device (Biotronik, Belos VR) was implanted for secondary prevention after an episode of hemody- namically unstable sustained monomorphic ventri- cular tachycardia (SMVT). The device has been programmed for a three-zone therapy, slow ventri- cular tachycardia (VT) (130 to 171 bpm), fast VT (171 to 214 bpm) with the stability interval set at

± 20 ms and ventricular fibrillation (VF). The co- unter of detection was set at 26 and at 16 RR inte- rvals in the slow and fast VT zone, respectively.

In the slow VT zone, the programmed therapy consisted of 2 antitachycardia pacing schemes (burst type with 6 consecutive pulses) each containing 3 sequences, followed by cardioversion. In the fast VT zone, the programmed therapy consisted of cardiover- sion sequences (the first attempt was set at 7 J).

In the VF zone, the programmed therapy consisted of defibrillation shocks (the first shock was set at 20 J, and the remaining shocks were set at 30 J).

During a 24 hours ambulatory ECG recording, an episode of SMVT (Fig. 1) occurred at a mean initial rate of 150 bpm, with unstable RR intervals (Fig. 2A, B). The ICD device was unable to detect this irregular rhythm in the slow VT zone. After- wards, this ventricular rhythm was accelerated to a faster but regular SMVT (180 bpm) (Fig. 2C), which was detected in the fast VT zone. The pro- grammed cardioversion therapy of 7 J was followed by degeneration of the fast VT to VF and the sinus rhythm was finally restored after two consecutive shocks of 20 J and 30 J, respectively (Fig. 1).

In order to eliminate underdetected irregular VT episode in the future, the stability interval was reprogrammed to ± 40 ms, based on the calculated mean values of the RR intervals differences during the recorded tachycardia.

(2)

282

Cardiology Journal 2008, Vol. 15, No. 3

www.cardiologyjournal.org

Figure 1. A 24 hours ambulatory ECG recording (dual channel) in an ICD recipient, showing an undetected episode of irregular ventricular tachycardia, lasting about 1.5 min. This rhythm was then accelerated to a faster but regular ventricular tachycardia followed by cardioversion, which led to the degeneration of ventricular tachycardia to ventricular fibrillation. Finally, the sinus rhythm was restored after two consecutive delivered shocks of 20 J and 30 J, respectively.

Figure 2. A. Initiation of a slow ventricular tachycardia; B. Ongoing slow ventricular tachycardia with irregular cycle length intervals; C. Acceleration of the slow and irregular ventricular tachycardia to a faster but regular cycle length interval form.

A

B

C

22:07:30 22:07:45

22:08:00 22:08:15 22:08:30 22:08:45 22:09:00

22:09:15 22:09:30 22:09:45

22:10:00

(3)

283 Athanasios G. Manolis et al., Irregular ventricular tachycardia, ICD underdetection

www.cardiologyjournal.org

Discussion

The cycle length of SMVT is usually stable.

Stabilization of the cycle length occurs after 10 to 20 beats from the initiation of the arrhythmia [1]. In addition, irregularity in the cycle length during on- going VT has been observed in 20% of episodes [2].

In ICD devices, a programmed stability inte- rval of 50–60 ms combined with a number of 12 to 16 RR intervals has the ability of a proper detec- tion over 90% of irregular VT episodes [3]. The probability of RR intervals irregularity increases in slower VT as well as in patients under antiarrhyth- mic therapy [4–6]. The underlying mechanism of cycle length irregularity is not well understood.

Rate dependent changes in the electrophysiologic properties of the arrhythmogenic substrate (con- duction velocity, refractory period) may contribute to the prescribed irregularity.

Given that the aforementioned underdetection was due to the programmed stability criterion, spe- cial attention must be paid when it comes to pro- gramming this detection parameter.

Acknowledgements

The authors do not report any conflict of inte- rest regarding this work.

References

1. Volosin KJ, Beauregard LA, Fabiszweski R et al. Spontaneous changes in ventricular tachycardia cycle length. J Am Coll Cardiol, 1991; 17: 409–414.

2. Wellens HJJ, Bar FW, Lie KI et al. The value of the electrocardio- gram in the differential diagnosis of a widened QRS complex.

Am J Cardiol, 1978; 64: 27–33.

3. Alberola A, Ili-Mayry S, Block M et al. RR interval variability in irregular monomorphic ventricular tachycardia and atrial fibril- lation. Circulation, 1996; 93: 295–300.

4. Swerdlow CD, Ahern T, Chen PS et al. Undertedection of ven- tricular tachycardia by algorithms to enhance specificity in tiered-therapy cardioverter-defibrillator. J Am Coll Cardiol, 1994; 24: 416–424.

5. Swerdlow CD, Friedman PA. Advanced ICD trouble- shooting: Part I. Pacing Clin Electrophysiol, 2005; 28:

1322–1346.

6. Swerdlow CD, Friedman PA. Advanced ICD troubleshooting:

Part II. Pacing Clin Electrophysiol, 2006; 29: 70–96.

Cytaty

Powiązane dokumenty

AIMS The aim of this study was to adapt and test the reliability of the Polish version of the Implantable Cardioverter ‑Defibrillator Concerns (ICDC) questionnaire, an

or left ventricular wall; and the epicardial pacing lead was attached in the right ventricular apical region via left minithoracotomy..

A 65-year-old woman with highly symptomatic (New York Heart Association [NYHA] class III/IV), non-ischaemic cardiomyopathy with severe systolic left ventricular dysfunction

Therefore, a temporary external implantable cardioverter-defibrillator (ICD) with dual-coil lead and a conventional ICD was implanted on the opposite side of the planned

Third Department of Cardiology, School of Medicine with the Division of Dentistry in Zabrze, Medical University of Silesia, Katowice, Poland.. A 40-year-old obese male without

For the prevention of sudden cardiac death teh patient was equipped with wearable car- dioverter defibrillator (WCD; LifeVest ® by Zoll) as a bridge therapy until the

Hybrid ablation of haemodynamically unstable ventricular tachycardia using a transabdominal minimally-invasive approach and percutaneous left ventricular assist device..

Cardiac magnetic resonance imaging confirmed the presence of hypertrabeculation with a two-layered structure of the endocardium with an increased noncompacted to compacted