88 Advances in Interventional Cardiology 2017; 13, 1 (47)

Image in intervention

Corresponding author:

Jakub Baran, Department of Interventional Cardiology, Medical College Jagiellonian University and the John Paul II Hospital, 80 Pradnicka St, 31-202 Krakow, Poland, phone: +48 791 330 302, e-mail: jakub_baran@yahoo.pl

Received: 26.10.2016, accepted: 19.02.2017.

Intravascular ultrasonography guided 60-mm long

drug-eluting tapered stent implantation in a long calcified lesion in a patient with stable coronary artery disease

Jakub Podolec, Jakub Baran, Łukasz Niewiara, Piotr Odrowąż-Pieniążek, Krzysztof Żmudka

Department of Interventional Cardiology, Medical College Jagiellonian University and the John Paul II Hospital, Krakow, Poland

Adv Interv Cardiol 2017; 13, 1 (47): 88–89 DOI: https://doi.org/10.5114/aic.2017.66196

Patients with stable coronary artery disease and dif- fuse long lesions are becoming more often a challenge for interventional cardiologists. Complete revascularization often requires implantation of many stents into a single artery. Overlapping of stents, required for drug-eluting stents (DES), leads to higher metal density, which is also potentially more favorable for healing problems and car- diac events [1–3].

A  62-year-old patient with hypertension, diabetes and dyslipidemia presented to the Department of Inter- ventional Cardiology, Jagiellonian University College of Medicine in the John Paul II Hospital with moderate to severe angina (class II/III according to the Canadian Car- diovascular Society Classification), diagnosed 2 months ago with coronary artery disease. According to Heart Team qualification, the patient was admitted to a  re- ferral invasive cardiology center to perform left anterior descending (LAD) angioplasty (Figure 1). The fractional flow reserve (FFR) result of 0.6 confirmed severity of the stenosis. Coronary angiography was performed and re- vealed a  long calcified lesion in the proximal and mid segment of the LAD. Both the LAD and the diagonal (Dg) branch were secured with guide wires. To estimate the vessel diameter intravascular ultrasound (IVUS) was performed, but due to extensive arterial calcification in the proximal segment introduction of the Eagle eye IVUS catheter was impossible (Figure 1). Predilatation with a non-compliant balloon 3.0/15 mm up to 24 atm in the proximal and mid part was performed. Vessel di- ameter estimation after IVUS measurements was per- formed and revealed dissection in the mid part of the LAD. A  60  mm Sirolimus drug-eluting tapered Biomime morph stent with the proximal diameter of 3.0 and dis- tal diameter of 2.5 mm was implanted without compli- cations. The guidewire from the Dg branch was again

introduced through the struts of the stent. The IVUS im- aging showed a  need of postdilatation in the mid part of the stent. Non-compliant 3.75/15  mm was used for optimization of the result. Kissing balloon technique was used to optimize the bifurcation result. The IVUS of the LAD also revealed non-critical distal dissection distally to the stent with slower blood flow to the distal part of the artery. A short 2.5/18 mm DES was implanted to cover the whole dissection distally. Final IVUS measurements revealed minimal lumen diameter (MLA) in the proximal part of the stent of 8.0 mm² and in the distal part 3.8 mm².

Proximally the plaque edge burden was 45.5%, and dis- tally 52.8%. The IVUS measurements confirmed the good angiographic result.

The stent introduction after optimal lesion preparation with a non-compliant balloon was easy and predictable.

Treatment of long lesions with single stent implantation is an interesting option for most patients and reduces the need for implantation of many overlapping stents.

Conflict of interest

The authors declare no conflict of interest.

References

1. Räber L, Jüni P, Löffel L, et al. Impact of stent overlap on angio- graphic and long-term clinical outcome in patients undergoing drug-eluting stent implantation. J Am Coll Cardiol 2010; 55:

1178-88.

2. O’Sullivan CJ, Stefanini GG, Räber L, et al. Impact of stent overlap on long-term clinical outcomes in patients treated with new- er-generation drug-eluting stents. Eurointervention 2014; 9:

1076-84.

3. Colombo A, Al-lamee R. Interventional cardiology: does stent overlap make a difference to clinical outcome? Nature Rev Car- diol 2010; 7: 362-4.

(2)

Jakub Podolec et al. IVUS-guided implantation of a long, tapered DES

89

Advances in Interventional Cardiology 2017; 13, 1 (47)

Figure 1. Long stenosis in the proximal and mid segment of the LAD prior to stent im- plantation (A, A’); stent introduction and po- sitioning (B) stent implantation (B’); IVUS re- sult prior to stent implantation (C) and after dilatation (C’); Final angiographic result (D)

A

B

C

D

A’

B’

C’

Cytaty

Powiązane dokumenty

Coronary angiography, fluoro- scopic digital stent enhancement (DSE), and intravascular ultrasound revealed recurrence of ISR in the proximal segment of the LAD caused by

ing catheter with side holes, the right coronary artery intravascular ultrasound confirmed an underexpanded stent with heavy circumfer‑.. ential calcification and IVL treatment

We provided evidence that DES implantation was suc- cessfully performed for the treatment of SCA reocclusion after BMS implantation, and the vessel patency was maintained with

After intracoronary vascular ultrasound (IVUS) for artery size and plaque morphology estimation (Fig. 2A) a tapered sirolimus drug-eluting stent Meril, Biomime Morph, 60 mm

According to our findings and previous studies, we suggest that adequate lesion preparation, higher pressure or longer bal- loon inflation time, and adjuvant postdilatation may

Study flow chart; COACT — CathOlic University of Korea: percutAneous Coronary inTervention; DES — drug eluting stents; GFR — estimated glomerular filtration rate; MDRD

Randomized comparison of everolimus-eluting and paclitaxel-eluting stents: two-year clinical follow-up from the Clinical Evaluation of the Xience V Everolimus Eluting

Methods: Twenty-two 7-month-old male farm pigs underwent implantation of two steel stents, one short (8 mm length) and one long (16 mm length), in the right coronary artery. The