Procedural and follow-up insights from optical coherence tomography-guided management

of intracoronary stent loss and crushing

Lorenzo Azzalini, Satoru Mitomo, Luciano Candilio, Antonio Colombo

Division of Interventional Cardiology, Cardio-Thoracic-Vascular Department, San Raffaele Scientific Institute, Milan, Italy

A 55-year-old man underwent left anterior descending (LAD) chronic total occlusion interven- tion (Fig. 1A, B).

The occlusion was crossed antegradely and predilated (Fig. 1C). During delivery of a 3.0 ×

× 33-mm stent, the operator encountered signifi- cant resistance, and soon noticed stent dislodg- ment from the delivery system. Since the stent appeared deformed and was located in front of the circumflex and first diagonal (D1) ostia (Fig. 1D), the operator chose to crush it (Fig. 1E). A 3.5 ×

× 33-mm stent (Fig. 1F) was then implanted and post-dilated with a 4.5-mm non-compliant bal- loon. Optical coherence tomography revealed that stent crushing also involved a 3.5 × 9 mm (likely malapposed) stent that was implanted across the LAD-D1 bifurcation 2 months earlier, which might have caused subsequent stent dislodgment in the following intervention. Final result was good (Fig.

1G, H). Optical coherence tomography confirmed optimal stent crushing (Fig. 1a–d). The patient

was discharged on long-term dual antiplatelet therapy (aspirin + ticagrelor).

A 5-month follow-up coronary angiogram was performed, which confirmed stent paten- cy. Optical coherence tomography showed com- plete endothelialization of the crushed stents with embedment into the vessel wall (Fig. 1a’–d’, Supplementary Video 1 — see journal website).

Therefore, the decision was made to discontinue ticagrelor after completing a 12-month course.

Stent loss is a rare but serious percutaneous coronary intervention complication. Management strategies include implantation, retrieval, or crush- ing. Crushing is preferred when the dislodged stent is deformed and/or in an inappropriate location, like in this case. An intravascular imaging-based man- agement is essential, first to demonstrate adequate crushing, and subsequently to confirm full endothe- lialization of the crushed stent on follow-up, thus reassessing the risk of stent thrombosis and possibly avoiding prolonged dual antiplatelet therapy.

Conflict of interest: Dr. Azzalini reports having received research support from ACIST Medical Systems and honoraria from Guerbet. The other authors have no disclosures to report.

INTERVENTIONAL CARDIOLOGY

Cardiology Journal 2017, Vol. 24, No. 6, 702–703

DOI: 10.5603/CJ.2017.0145 Copyright © 2017 Via Medica

ISSN 1897–5593

702 www.cardiologyjournal.org

IMAGE IN CARDIOVASCULAR MEDICINE

(2)

Figure 1. A, B. Mid left anterior descending chronic total occlusion. The occlusion was crossed with antegrade wire escalation and predilated (C). D. A 3.0 × 33-mm drug-eluting stent (DES) was dislodged from the delivery system, appeared severely distorted (arrowheads) and was located in front of the circumflex and first diagonal ostia. E. Balloon crushing of the dislodged stent was performed. F. A 3.5 × 33-mm DES was implanted and post-dilated, G, H. with optimal final result. In the lower panel, angiographic and optical coherence tomography (OCT) images of the final result after the index procedure in (I) cranial and (J) caudal projections are shown in detail. a–c. Optimal stent apposi- tion is observed. Arrowheads in (b, c and d) point at the crushed stents. The asterisk in panel b indicates that just one layer of metal is seen on the opposite side with regard to the crushed segment, thus suggesting that crushing also involved a short (likely malapposed) stent that had been previously implanted across the left anterior descending-first diagonal bifurcation. At 5-month follow-up coronary angiogram (I’ and J’), a good result is observed and confirmed by OCT imaging (a’–d’). Arrowheads in panel b’ and c’ point at the crushed stents, which are now fully endothelialised.

www.cardiologyjournal.org 703

Lorenzo Azzalini et al., OCT-guided management of stent loss

Cytaty

Powiązane dokumenty

Following stent ablation, OCT showed no visible strut area in the site of the previously underexpanded stent (Fig.. Final angiography showed good distal flow without

Angiography demonstrating vertical filling defect (white arrow) in the left main bifurcation lesion (white dotted line: proximal edge of the previously implanted

Incidence and short-term clinical outcomes of small side branch occlusion af- ter implantation of an everolimus-eluting bioresorbable vascular scaffold: an interim report of

The selection of ap- propriate stent size and stent optimization to avoid under-expansion and malapposition due to the stenotic lesion of ectatic RCA led to an assessment

Very late stent thrombosis derived from thin-cap neoatheroma and fibroatheroma with plaque rupture assessed by optical coherence tomography.. Yongcheol Kim, Myung Ho Jeong, Min

angiography and optical coherence tomography (OCT) imaging revealed a long, spiral, spontaneous coronary dissection (SCAD) of the right coronary artery (Fig.. Right coronary

Conclusions: Optical coherence tomography through a catheter extension system is a useful technique able to overcome some of the limitations associated with this imaging

Different patterns of neointimal coverage between acute coronary syndrome and stable angina after various types of drug-eluting stents implantation; 9-month follow-up optical