www.kardiologiapolska.pl

Kardiologia Polska 2017; 75, 1: 81; DOI: 10.5603/KP.2017.0006 ISSN 0022–9032

STUDIUM PRZYPADKU / CLINICAL VIGNETTE

Address for correspondence:

Radosław Kozaryn, MD, Medical University of Gdansk, ul. Dębinki 7, 80–291 Gdańsk, Poland, e-mail: radekkozaryn@gmail.com Conflict of interest: Dariusz Jagielak is a proctor for Edwards Lifesciences company.

Kardiologia Polska Copyright © Polskie Towarzystwo Kardiologiczne 2017

Balloon expandable transcatheter heart valve eccentricity in the non-calcified aortic

annulus — should the aortic annulus stiffness come on stage?

Ekscentryczny kształt zastawki aortalnej rozprężalnej na balonie w niezwapniałym pierścieniu aortalnym — czy ma znaczenie kliniczne?

Dariusz Jagielak, Radosław Kozaryn, Maksymilian Mielczarek, Dariusz Ciećwierz

Medical University of Gdansk, Gdansk, Poland

An 85-year-old male with symptomatic severe aortic stenosis was admitted for transcatheter aortic valve replacement (TAVR). Multi-detector computed tomography (MDCT) revealed elliptical configuration of the aortic annulus with a long axis diameter of 31 mm and short axis diameter of 24 mm; the annulus perimeter and area were 87 mm and 584 mm2, respectively (Fig. 1A). No calcifications within the aortic annulus and symmetrically distributed calcifications of the leaflets were detected (Fig. 1B). The decision to implant a balloon expandable transcatheter heart valve (THV) 29 mm Sapien XT (Edwards Lifesciences) was made. The calculated annular area oversizing was 13%. The prosthesis implantation was preceded with a 20 mm diameter balloon predilation. Transoesophageal echocardiography confirmed good result of the procedure, with no paravalvular aortic regurgitation. During the postoperative period the patient developed severe pulmonary infection and died. In a post-mortem examination the absence of calcifications of the aortic annulus and sym- metric distribution of the leaflet calcifications were confirmed (Fig. 1C, D). Surprisingly, the aortic annulus was found to have the same configuration as shown in MDCT prior to the procedure. Due to the non-circular shape of the prosthesis frame the valve leaflets and the line of coaptation were significantly rotated (Fig. 1E). The updated VARC-2 document implements a novel parameter of TAVR success, time-related valve safety, which combines valve dysfunction, endocar- ditis, and thrombotic complications of the prosthesis [Kappetein AP et al., J Thorac Cardiovasc Surg, 2013; 145: 6–23].

All of these seem to be extremely dependent on proper leaflet alignment, which provides laminar blood flow through the THV. Post-implant THV non-circularity may result in distortion of the leaflet coaptation line. Therefore, the indices of correct post-implant THV frame geometry should be assessed in patients with long life expectancy referred for TAVR.

The case presented herein points out the need to complement pre-TAVR assessment with a (frequently omitted) analysis of the aortic annulus deformation over the cardiac cycle (annulus stiffness), potentially the second (after its calcifications) indicator of post-implant THV eccentricity [Schuhbaeck A et al., Eur J Radiol, 2015; 84: 1242–1248; Marsi et al., J Thorac Cardiovasc Surg, 2014; 147: 1847–1854].

Figure 1. A. Multi-detector computed tomograp- hy (MDCT), pre-operative estimation of the aortic annulus diameters (31 mm, 24 mm); B. MDCT, no calcifications of the aortic annulus and symmetric calcification of aortic leaflets; C, D. Post-mortem examination pictures; E. Significant twist of the leaflets and the coaptation line

E

C D

A B

Cytaty

Powiązane dokumenty

Transesophageal echocardiography (TEE) showed bicuspid aortic valve (BAV) with severe eccentric regurgitation jet along anterior mitral leaflet (Fig. 1A), aneurysm of the root

It has been shown that new-generation valves reduced the rates of paravalvular leak (PVL), annular rupture and second valve implantation compared with the early-generation ones

Early regression of left ventricular mass associated with diastolic improvement after transcatheter aortic valve implantation.. Levy D, Garrison RJ, Savage DD, Kannel WB,

Transcatheter aortic valve implantation for the treatment of severe symptomatic aor- tic stenosis in patients at very high or prohibitive surgical risk:.. Acute and late outcomes of

An ECG showed a sinus rhythm with very frequ- ent premature ventricular contractions (PVC) and runs of non-sustained repetitive monomorphic ven- tricular tachycardia (RMVT) with

In the short- and medium-term, the treatment of pa- tients with severe aortic stenosis and the bicuspid aortic valve is as effective as that with the tricuspid aortic valve

Valve concept and implantation process: A – valve design, B – projections (green and red arrows) where the bottoms of three sinuses and commissures are visible in a single line and

1 Department of Cardiology, Congenital Heart Diseases and Electrotherapy, Medical University of Silesia in Katowice, Silesian Center for Heart Diseases in Zabrze, Poland.. 2 3 rd