Intracardiac damage of silicone insulation of a single lead and infective endocarditis

Wewnątrzsercowe uszkodzenie silikonowej izolacji pojedynczej elektrody i infekcyjne zapalenie wsierdzia

Barbara Małecka

1

, Anna Rydlewska

1

, Andrzej Ząbek

1

, Jerzy Sadowski

2

, Jacek Lelakowski

1

1Department of Electrocardiology, John Paul II Hospital in Krakow, Poland

2Department of Cardiosurgery, John Paul II Hospital in Krakow, Poland

A 59-year-old woman after artificial mitral valve and pacemaker implantation (VVI with ventricular lead Biotronik Synox 60BP) underwent cardiac surgery due to infective endocarditis. Transoesophageal echocardiography performed to diagnose recurrent septic fevers revealed vegetations connected to the pacemaker lead and artificial valve, and com- puted tomo graphy confirmed multiple pulmonary embolisms. No pacing or sensing abnormalities of the pacemaker were present. Endocardial pacing lead, together with artificial valve, was removed using extracorporeal circulation.

Simultaneously, a new mitral valve and an epicardial pacing system were implanted. The lead unseal with metal wire exposure was discovered (Fig. 1B). In this location, vegetation infected with Staphylococcus was attached to the lead (Fig. 1A). The 9-year-old lead was damaged at its passing through the tricuspid valve. The use of an optical microscope confirmed silicone layer reduction up to a complete absence of the outer lead insulation (Fig. 1C). In spite of antibiotic and cardiosurgical therapy, the patient died due to septic shock. Lead-dependent infective endocarditis (LDIE) associa- ted with abrasion of silicone insulation of two endocardial leads as a result of their friction in the right atrium has been previously described. We have also described abrasion of silicone insulation in intracardiac parts of pacemaker leads as a new mechanism of LDIE. In this case, abrasion with disruption of silicone insulation occurred in a single lead system implanted 9 years previously. A possible cause of damage was lead abrasion due to its contact with the tricuspid valve and silicone wear caused by lead bending. LDIE followed by pulmonary embolism and artificial mitral valve infection was associated with mechanical damage of the lead and resulted in the patient’s death. Damage of lead insulation was electrically silent. Damage of the endocardial pacemaker lead with disruption of outer silicone insulation and metal wire exposure may occur in the presence of a single lead. LDIE with severe clinical presentation and lack of heart stimulation malfunction may be a symptom of such damage.

Address for correspondence:

Anna Rydlewska, MD, PhD, Department of Electrocardiology, John Paul II Hospital in Krakow, ul. Prądnicka 80, 31–202 Kraków, Poland, tel: +48 12 614 22 77, fax: +48 12 633 23 99, e-mail: annarydlewska@op.pl

Conflict of interest: none declared

Figure 1. A. Ope - rating table:

surgically removed lead with attached infected vegetation (arrow); B. Site of isolation disruption with exposure of the lead inside filled with haemolysed blood (arrow);

C. Macrophoto- graphy — silicone insulation re- duction, up to complete lack of isolation and metal wire exposure

A C

B

Cytaty

Powiązane dokumenty

The MyoPore (Greatbatch Medical, New York, United States) sutureless screw -in epicardial pacing lead was delivered to the target area with the steerable FasTac Flex

nous lead extraction of the abandoned lead with surgical treatment of the fistula in the axillary fossa, leaving the pacing system on the right side of the chest.. First,

Any Riata and Riata ST lead advisory concerning lead extraction or replacement should be considered carefully, especially bearing in mind the results of analysis of the Accufix

Due to the large size of the vegetation (the length: 71 mm) connected to the lead and its coarse texture (Fig. 1), the patient was qualified for complete system removal during

Tissue fragments encapsulating endocardial lead body except the vegetation were mainly connective tissue with partial hyalinisation, with features of vasculogenesis (Fig.. Fragments

In the next stage of tribological destruction of silicone insulation there is abrasive wear caused by detachment of particles from the surface of polymer as the effect of fatigue

It has been demonstrated that the presence of fQRS in patients with coronary artery disease (CAD) has been related to regional myocardial damage, increased adverse cardiac

Based on the thrombus described in the right atrium (TEE before operation showed a 5 × 8 mm structure in contact with the lead), anamnesis of pulmonary embolism and inflamma-