Kardiologia Polska 2018; 76, 2: 472; DOI: 10.5603/KP.2018.0043 ISSN 0022–9032

STUDIUM PRZYPADKU / CLINICAL VIGNETTE

Address for correspondence:

Dr. Muhammet Dural, Eskisehir Osmangazi University, Cardiology Department, Büyükdere Mahallesi, 26040 Odunpazarı/Eskişehir, 26040 Eskisehir, Turkey, e-mail: muhammetdural@yahoo.com

Conflict of interest: none declared

Kardiologia Polska Copyright © Polskie Towarzystwo Kardiologiczne 2018

Fracture of the distal part of an implantable cardioverter-defibrillator lead two months after implantation

Złamanie dalszej części elektrody wszczepionego kardiowertera-defibrylatora po dwóch miesiącach od implantacji

Muhammet Dural, Kadir Ugur Mert, Selda Eraslan, Kemal Iskenderov, Yuksel Cavusoglu

Cardiology Department, Eskisehir Osmangazi University, Büyükdere Mahallesi, Eskisehir, Turkey

Most lead fractures occur near the subclavian venous entry site due to compression of the lead between the clavicle and the first rib. We present a patient with a distal complete ventricular lead fracture that occurred two months after implantation. A 51-year-old man with ischaemic dilated cardiomyopathy, whose ejection fraction was 21% underwent dual-chamber implantable cardioverter-defibrillator (ICD) implantation (St. Jude Medical, DR 2377-36C ICD) in our centre. The ventricular lead and atrial lead were placed passively by using subclavian vein puncture (Fig. 1A, B). Atrial and ventricular lead impedances were 530 Ω and 740 Ω, respectively. After two months, in routine pacemaker control, the ventricular lead impedance was found to be dramatically increased (> 3000 Ω). Fluoroscopy showed a complete ventricular lead fracture at the distal part of the electrode (Fig. 2A). Then extraction of the fractured lead and insertion a new ventricular lead was planned. Two days later, just before the operation, fluoroscopy showed that the fractured distal part of the lead was embolised to the distal right pulmonary artery (Fig. 2B). The fractured lead was extracted (Fig. 3) and a new ventricular lead was implanted with no complications (Fig. 4). In our patient the fracture occurred a short time after implantation. The patient stated that he used a chainsaw to cut wood about one month after pacemaker implanta- tion. It was thought that this severe physical activity, which could generate intense vibration, could result in lead fracture.

Our case documented that fracture may occur in any segment of the lead, so routine pacemaker control should be made carefully. In the presence of any problem that indicates lead fracture, leads must be investigated with fluoroscopy.

Figure 4. Implanted new ventricular lead

A B

A B

Figure 1. A. Successfully implanted atrial lead in the first procedure;

B. Successfully implanted ventricular lead in the first procedure

Figure 2. A. Complete ventricular lead fracture at the distal part of the electrode; B. Embolised distal part of the lead to the right pulmonary artery

Figure 3. Macroscopic view of the fractured lead

Cytaty

Powiązane dokumenty

The study had no statistical power to prove that LITR is associated with worsening of the prognosis in patients with ARVC and high risk of sudden cardiac death.. We assume

AIMS The aim of this study was to adapt and test the reliability of the Polish version of the Implantable Cardioverter ‑Defibrillator Concerns (ICDC) questionnaire, an

1,2 The procedure was performed under regional anesthesia of the hemithorax, with combined ultrasound ‑guided serratus anterior plane block (SAPB), 3 pectoral nerve block

Other independent predictors for mortality in patients with heart failure after implantable cardioverter ­defibrillator placement are: age older than 65 years, impaired

Intravascular techniques for lead extraction including locking stylets and telescoping sheaths significantly reduce the risk of electrode rupture, but helix fracture and

In the last stage the flattened Dotter’s basket was introduced to the heart inside the Byrd dilator and then fastened to the described lead as the external ‘splint’.. This solution

As shown in (B), when the distal (–) and proximal (+) pins are physically reversed in the implantable cardioverter defibrillator header ports, a broad unipolar sensing

Reproducibility of microvolt T-wave alternans (MTWA) in vector leads; VM — vector magnitude; X, Y, Z — Frank leads, HR (BPM) — continuous heart rate trend, %Bad — the percentage