• Nie Znaleziono Wyników

New imaging device for the detection of lipid core-containing plaques

N/A
N/A
Protected

Academic year: 2022

Share "New imaging device for the detection of lipid core-containing plaques"

Copied!
2
0
0

Pełen tekst

(1)

www.cardiologyjournal.org 386

NEW TECHNOLOGIES IN CARDIOLOGY

Cardiology Journal 2008, Vol. 15, No. 4, pp. 386–387 Copyright © 2008 Via Medica ISSN 1897–5593

New imaging device for the detection of lipid core-containing plaques

S. Suave Lobodzinski

California State University Long Beach, CA, USA

Address for correspondence: S. Suave Lobodzinski, PhD, Department of Electrical and Biomedical Engineering, California State University Long Beach, 1250 Bellflower Blvd, Long Beach, CA 90840, USA, tel: (562) 985 5521, fax: (562) 985 5899, e-mail: slobo@csulb.edu

A new device for the detection of lipid core- containing plaques called a Catheter-based Lipi- Scan™ Coronary Imaging System manufactured by InfraReDx has received 510(k) clearance from the U.S. Food and Drug Administration on 29 April 2008.

The LipiScan device uses near-infrared spec- troscopy to identify lipid core-containing plaques of interest in the coronary arteries in patients already undergoing cardiac catheterization. Such plaques, which cannot be detected by commonly-used tests such as a treadmill examination or even coronary angiography, are suspected to be the cause of most sudden cardiac deaths and non-fatal heart attacks.

The InfraReDx LipiScan Imaging System uses infrared imaging to detect lipid core-containing pla- ques of interest and assess a patient’s coronary artery lipid core burden index. The LipiScan Coro- nary Imaging System is a catheter-based device that uses laser light to detect how much fat and other substances are contained in a plaque.

Near infrared (NIR) diffuse spectroscopy is a technique based on the absorption of light in the NIR spectrum, in a specific manner, by organic molecu- les. NIR spectroscopy has demonstrated the ability to identify plaques with lipid pools through blood [1].

The device works by placing a catheter equip- ped with a fiber-optic laser into the artery. The de- vice shines the near infrared light through the blo- od to the artery wall and measures the light reflec- ted back from it, a technique called spectroscopy.

The reflected wavelengths vary depending on how much fat and other substances are in the plaque in the illuminated portion of the wall.

The identification of the chemicals present is based on the differential absorption of light in the NIR spectrum by different molecules. An important feature of near-infrared light is that it can penetra- te tissue and can therefore identify a tissue despite the presence of blood between the detector and the target. This is an important advantage for imaging within the human coronary artery.

The system consists of a laser light source, an automated pullback and rotation device and a small fiber-optic catheter. While the catheter is similar in size and ease of use to an intravascular ultraso- und catheter, the information it provides is quite dif- ferent since it is based on an optical rather that an ultrasonic signal (Fig. 1).

The NIR system obtains signals from patients that are analyzed with algorithms validated by com- parison to tissue histological findings in ex-vivo coronary specimens. It is therefore possible to per- form a pullback in a patient’s artery and provide an image of the NIR signals which indicate the presen- ce of lipids and other chemicals of interest. It is expected that these images, which are called Intra- vascular Chemograms™, will provide information for interventional cardiologists to help in the care of patients already undergoing cardiac catheterization for coronary events. The identification of the che- mical composition of coronary plaques is of value to cardiologists in the selection of medical, stenting or surgical therapy for coronary lesions. The device is also expected to be of value to the pharmaceutical industry as a means to assess the effect of novel anti- atherosclerotic agents on lipid core plaque burden.

(2)

387 S. Suave Lobodzinski, New imaging device for the detection of lipid core-containing plaques

www.cardiologyjournal.org

The initial use of the InfraReDx device will be for diagnosis, which will help in prevention of a se- cond coronary event in the approximately 2 million individuals world-wide who undergo a coronary in- tervention each year.

Figure 1. Comparison of intravascular to near-infrared plaque imaging.

References

1. Moreno PR, Lodder RA, Purushothaman KR et al. Detection of lipid pool, thin fibrous cap, and inflammatory cells in human aortic atherosclerotic plaques by near-infrared spectroscopy.

Circulation, 2002; 105: 923–927.

Cytaty

Powiązane dokumenty

The major findings of the present study are as follows: 1) there was no correlation between LDL-C, TC and TG plasma levels and the number of li- pids in coronary plaque assessed

The aim of this study was to evaluate the prognostic value of daily monitored device-based diagnostic parameters as predictors of malignant ventricular arrhythmias in

The perioperative use of either dual antiplate- let therapy or heparin ‘bridging’ significantly in- creases the risk of procedure-related bleeding com- plications following

Leczenie farmakologiczne za pomocą statyn wdraża się u wszystkich pacjentów z cukrzycą i cho- robami układu sercowo-naczyniowego, niezależnie od początkowego stężenia

Zdecydowana wiêkszoœæ blaszek pêkniêtych obrazo- wanych z zastosowaniem IVUS by³a ekscentryczna, a pêkniêcie lokalizowa³o siê na granicy fragmentu zdro- wego obwodu œciany

Pulse wave transit time measurements of car- diac output in patients undergoing partial hepatectomy: a comparison of the esCCO system with thermodilution. Yamada T, Tsutsui M, Sugo

The first part discusses the psychometric proper- ties of the adapted tool, while the second one describes the meaning of core beliefs in the process of post-trau- matic

The incidence of mixed, hard, calcified and concentric plaques increased, whereas the percent- age of soft, non-calcified and eccentric plaques decreased gradually with the increase