Balloon leakage dilation over arterio-aorta rail support greatly facilitates a complex aortic paravalvular leak closure

N/A
N/A
Protected

Academic year: 2022

Share "Balloon leakage dilation over arterio-aorta rail support greatly facilitates a complex aortic paravalvular leak closure"

Copied!
3
0
0

Pełen tekst

(1)

Creative Commons licenses: This is an Open Access article distributed under the terms of the Creative Commons 223

Attribution-NonCommercial-ShareAlike 4.0 International (CC BY -NC -SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/).

Short communication

Corresponding author:

Eustaquio Maria Onorato MD, Centro Cardiologico Monzino, Istituto di Ricovero e Cura a Carattere Scientifico (IRCCS), University School of Milan, Milan, Italy, phone: +39 348 6939883, e-mail: eustaquio.onorato@gmail.com

Received: 6.01.2021, accepted: 4.02.2021

Balloon leakage dilation over arterio-aorta rail

support greatly facilitates a complex aortic paravalvular leak closure

Eustaquio Maria Onorato1, Massimo Carasi2, Mirko Zadro2, Umberto Cucchini2, Fabio Chirillo2

1Centro Cardiologico Monzino, Istituto di Ricovero e Cura a Carattere Scientifico (IRCCS), University School of Milan, Milan, Italy

2UOC Cardiologia, Ospedale di Bassano del Grappa, Azienda Sanitaria ULSS 7 Pedemontana, Bassano del Grappa, Italy

Adv Interv Cardiol 2021; 17, 2 (64): 223–225 DOI: https://doi.org/10.5114/aic.2021.107505

In patients with an aortic paravalvular leak (PVL), the retrograde femoral arterial approach is most common- ly used [1, 2]. Nevertheless, in some cases an antegrade (via trans-septal puncture) or a  retrograde transapical approach may be alternative options [3]. Particularly for the anatomically hard-to-approach calcified serpiginous tracts where an extra support is needed, an arterio-ar- terial loop can be established and may be key for a suc- cessful procedure [4].

A  75-year-old male with previous ischaemic stroke, paroxysmal atrial fibrillation, and multiple previous percu- taneous coronary interventions was admitted for conges- tive heart failure (NYHA class III). He had a history of sur- gical removal of subaortic membrane and right coronary leaflet plasty, followed 20 years later by mechanical aor- tic valve replacement and woven fabric ascending aortic surgical graft. Transthoracic echocardiography (TTE) and transoesophageal echocardiography (TEE) showed dilata- tion of left ventricle (LV) with an ejection fraction of 45%, and confirmed the presence of a  calcified oval-shaped 6 × 5 mm PVL located underneath the left coronary sinus.

The patient denied a redo surgery and after heart team discussion a transcatheter PVL closure was planned.

An informed consent form was signed by the patient. The procedure was performed under TEE and fluoroscopic guidance using general anaesthesia. Right femoral artery puncture was performed, and a severe paravalvular aor- tic regurgitation was also demonstrated by aortography (Figure 1 A). With the support of a 4-Fr Amplatz left cathe- ter (AL2) we easily passed a hydrophilic Terumo 0.35 inch guide wire retrograde from the aorta to the LV through the calcific serpiginous PVL; however, none of the catheters used (AL2- 4Fr, Judkins right 5-Fr, Multipurpose 5-Fr, Glide-

cath 4-Fr) could be advanced through the leak over the hy- drophilic wire. It was decided to re-advance the hydrophilic guidewire, establishing a  loop in the left ventricle, and then re-direct it across the mechanical prosthetic leaflets and place it finally into the descending aorta (Figure 2 A).

A Glidecath 4-Fr was subsequently advanced over the hy- drophilic wire that was replaced by 0.035-inch wire, 260 cm in lenght super stiff guidewire (Boston Scientific), thus cre- ating an arterio-aorta rail, without the need to externalize the wire through an additional radial access. Over the ex- change wire a 4.0 mm PTA balloon dilatation catheter was placed across the leakage and dilated twice, not beyond its rated burst pressure, taking care to keep in place the arterio-aorta rail (Figure 2 B). The stronger support finally allowed progression of the retrograde 7-Fr delivery sheath into the LV. A specifically designed, 7 mm square twist PLD device (PLD, Occlutech, Helsingborg, Sweden) was suc- cessfully deployed (Figures 2 C, D) and the final echocardi- ographic and angiographic control confirmed the effective PVL closure (Figure 1 B). The postoperative course was uneventful, and the patient was discharged on the third post-operative day. TTE confirmed stable position of the device with trivial residual leak.

In conclusion, the balloon leakage dilation over an arterio-aorta rail strong support allowed us to get across the anatomically hard-to-approach calcified serpiginous leak with the 7-FR delivery sheath and successfully ac- complish the procedure.

Conflict of interest

Eustaquio Maria Onorato is a consultant for Occlu- tech, manufacturer of the device. The remaining authors declare no confict of interest.

(2)

Eustaquio Maria Onorato et al. Complex aortic paravalvular leak closure

224 Advances in Interventional Cardiology 2021; 17, 2 (64)

Figure 1. A – Baseline aortic angiogram showing the regurgitant jet through the calcific serpiginous aortic paravalvular leak (black arrow); B – post-procedure aortography with the device in situ (white arrow) showing mild regurgitant jet

Figure 2. Procedural steps showing retrograde aortic paravalvular leak crossing and closure thanks to a modi- fied technique. A – The hydrophilic guidewire from the aorta crossed retrogradely the serpiginous leak and after forming a big loop in the left ventricle (white arrows) was advanced across the mechanical prosthetic leaflets into the aorta (black arrows), establishing an arterio-aorta rail; B – the hydrophilic wire has been replaced by a super stiff exchange guidewire, and over it a 4.0 mm PTA balloon dilatation catheter was placed across the leakage and dilated (orange arrow); C, D – after 7-Fr delivery sheath progression into the LV, a 7 mm square twist PLD (white arrow) was correctly deployed

B A

B

D A

C

(3)

Eustaquio Maria Onorato et al. Complex aortic paravalvular leak closure

225

Advances in Interventional Cardiology 2021; 17, 2 (64) References

1. Calvert PA, Northridge DB, Malik IS, et al. Percutaneous device closure of paravalvular leak. Combined experience from the United Kingdom and Ireland. Circulation 2016; 134: 934-44.

2. Cruz-Gonzalez I, Rama-Merchan JC, Rodriguez-Collado J, et al.

Transcatheter closure of paravalvular leaks: state of the art.

Neth Heart J 2017; 25: 116-24.

3. Damluj AA, Kaynak HE, Heldman AW. Combined retrograde/an- tegrade approach to transcatheter closure of an aortic paraval- vular leak. Tex Heart Inst J 2015; 42: 443-7.

4. Jiménez-Brítez G, Panaro A, Maldonado G, et al. Use of radial access to create an arterio-arterial loop to facilitate the percu- taneous closure of paravalvular aortic leak. Is it feasible? REC Interv Cardiol 2020; 2: 228-31.

Cytaty

Powiązane dokumenty

non–ST -segment elevation myocardial infarc- tion treated with percutaneous coronary inter- vention (PCI) with a drug -eluting stent (Supra- flex Cruz 2.5 × 44 mm; SMT Polonia, Gdańsk,

ence of >48° angle between the aortic annulus Transcatheter aortic valve implantation (TAVI)4. is a well ‑established modern treatment of severe symptomatic

Single -photon emission computed tomogra- phy (SPECT) showed exercise -induced revers- ible perfusion defect in the anterolateral wall en- compassing 10% of the left ventricular

Clinical manifestations of heart failure abate with transcatheter aortic paravalvular leak closure using Amplatzer Vascular plug II and III devices. Paravalvular leak from

Ascending aortic pseudoaneurysms are very rare entities generally occurring after heart surgery associated with cannulation for cardiopulmonary pump or aorta-coronary artery

Korelacje liniowe między wiekiem pacjenta w czasie operacji, czasem, który upłynął od operacji, wartościami ciśnienia tętniczego oraz GrAoD a wybranymi parametra-

Transcatheter aortic valve implantation (TAVI) is a less invasive treatment option for elderly, high-risk patients with symptomatic severe aortic stenosis (AS) than surgical

Sagittal phase contrast cine magnetic resonanse imaging of the thoracic aorta (velocity encoding gradient 150 cm/s).. Flow in the superior direction (towards the head) is shown