Proximal optimization technique, proximal rewiring, side-branch dilation and paired balloon inflation to minimize metal carina

in inadequate jailing

Fumiaki Nakao

Department of Cardiology, Yamaguchi Grand Medical Center, Yamaguchi, Japan

Figure 1. Clinical case with SYNERGY™ and schema of PPSP procedure. A. Baseline coronary angiogram; B. Longi- tudinal 3D-OCT cut-away view of inadequate jailing and proximal rewiring; C. Final coronary angiogram; D. Final 3D- -OCT image; E. Schema of PPSP procedure with X-ray fluorogram and 3D-OCT image of case with two-link BMX-J™

stent (Biosensors International Group Ltd.); * and #, same links divide distal cell at rewiring; arrowhead, re-crossed wire.

If distal rewiring is selected for inadequate jailing (distal cell on a side-branch ostium divided by a stent-link) during single two-link stent deploy- ment at the coronary bifurcation with side-branch dilation, kissing balloon inflation (KBI) might result in residual deformed struts in the side-branch ostium, and a sparse strut-covered distal main vessel. Although adequate rewiring during inad- equate jailing might be proximal rewiring [Nakao F.

The impact of proximal cell rewiring compared with distal cell rewiring with single-stent deployment and subsequent kissing balloon inflation in the inadequate jailing of a Kanam stent on the side branch ostium.

Cardiol J. 2017 Jun 27; ahead of print], simple KBI might result in residual metal carina. The sequence of procedures should also be considered.

A SYNERGY™ two-link stent (Boston Scientific Co.) was deployed in the main vessel of a bifurca-

279 www.cardiologyjournal.org

INTERVENTIONAL CARDIOLOGY

Cardiology Journal 2018, Vol. 25, No. 2, 279–280

DOI: 10.5603/CJ.2018.0036 Copyright © 2018 Via Medica

ISSN 1897–5593

IMAGE IN CARDIOVASCULAR MEDICINE

(2)

tion lesion (Fig. 1A) and the proximal optimization technique (POT) was applied. Three-dimensional optical coherence tomography (3D-OCT) con- firmed inadequate jailing and proximal rewiring (Fig. 1B). Side-branch dilation was followed by KBI (paired balloon inflation, Fig. 1C). The final 3D-OCT showed a shift to the distal main vessel and minimization of the metal carina (Fig. 1D).

Figure 1E shows a schema of the POT, proxi- mal rewiring, side-branch dilation, and paired balloon inflation (PPSP) procedure with X-ray

fluorogram and 3D-OCT images of a patient with a different two-link stent. Such stents are weak longitudinally, and their struts can be moved lon- gitudinally if properly handled. POT tilts the metal carina to the side-branch side, proximal rewiring and side-branch dilation lift the metal carina to the distal main vessel, and KBI apposes struts to the vessel wall. The PPSP procedure might help to minimize the metal carina when jailing by a two- link stent is inadequate. Further clinical studies are required.

Conflict of interest: None declared

280 www.cardiologyjournal.org

Cardiology Journal 2018, Vol. 25, No. 2

Cytaty

Powiązane dokumenty

Modified parasternal long axis view showing an antegrade (red) diastolic color Doppler flow in the anterior interventricular sulcus corre- sponding to the distal portion of the LAD;

these previous works do not satisfy the polynomial equation [6]. Uniquely, the authors have circum- navigated these pitfalls in detail, the basics of the assumptions were: 1)

Pre-dilatation was then performed using a 2.75/15 mm non-compliant balloon, and a 2.75/20 mm SYNERGY stent (Boston Scientific, Natick, MA, USA) was deployed at nomi- nal

The baseline coronary angiogram (C), the longitu- dinal cut-away view of the stent-enhanced 3D optical coherence tomography (D) and the stent-oriented 3D-IVUS (E) of another

As there- wrapped 2.0 × 12 mm noncompliant balloon used in pre-stenting KBI could not be advanced to the DG branch, presumably because of the jailed strut, using 2.25 mm

CC-Prox; the metal carina index is defined as the number of frames with malapposed struts continuously present from the distal end of the side branch ostium (SBO) to the proximal

proximal and distal vessel lesions could be the jailed position, and the additional support from the jailed buddy wire could be shared and used in the whole

Methods: Twenty-two 7-month-old male farm pigs underwent implantation of two steel stents, one short (8 mm length) and one long (16 mm length), in the right coronary artery. The