• Nie Znaleziono Wyników

Transvenous retained lead fragment removal after incomplete extraction assisted by three-dimensional transoesophageal echocardiography

N/A
N/A
Protected

Academic year: 2022

Share "Transvenous retained lead fragment removal after incomplete extraction assisted by three-dimensional transoesophageal echocardiography"

Copied!
1
0
0

Pełen tekst

(1)

www.kardiologiapolska.pl

Kardiologia Polska 2016; 74, 2: 195; DOI: 10.5603/KP.2016.0024 ISSN 0022–9032

STUDIUM PRZYPADKU / CLINICAL VIGNETTE

Address for correspondence:

Marcin Grabowski, MD, PhD, FESC, 1st Chair and Department of Cardiology, Medical University of Warsaw, ul. Banacha 1a, 02–097 Warszawa, Poland, tel: +48 22 599 19 58, fax: +48 22 599 19 57, e-mail: marcin.grabowski@wum.edu.pl

Conflict of interest: none declared

Kardiologia Polska Copyright © Polskie Towarzystwo Kardiologiczne 2016

Transvenous retained lead fragment removal after incomplete extraction assisted by three- -dimensional transoesophageal echocardiography

Przezżylne usunięcie pozostawionego fragmentu elektrody wspomagane obrazowaniem trójwymiarowej echokardiografii przezprzełykowej

Marcin Grabowski

1

, Andrzej Kutarski

2

, Agnieszka Kołodzińska

1

, Radosław Piątkowski

1

, Franciszek Majstrak

3

11st Department of Cardiology, Medical University of Warsaw, Warsaw, Poland

2Department of Cardiology, Medical University of Lublin, Lublin, Poland

3Department of Cardiosurgery, Medical University of Warsaw, Warsaw, Poland

A 79-year-old woman with diagnosed pacemaker lead endocarditis was admitted for percutaneous lead extraction.

Pacemaker and leads removal using Byrd polypropylene sheets was performed with incomplete radiographic success with retained 4 cm long atrial pacemaker lead fragment (Fig. 1A). Due to the patient’s poor clinical state a second at- tempt of the fragment extraction was performed. The procedure was assisted by transoesophageal echocardiography with three-dimensional visualisation showing the localisation of the lead fragment (Fig. 2A, B). Several unsuccessful attempts of traction of the lead fragment were performed using basket, pig-tail, and loop catheters and Needle’s-eye snare, but after each successful catch it slid and was lost by the catheter (Fig. 1B). Finally, success was achieved when the lead fragment was fixed with the pig-tail catheter and grasped with the basket catheter through the two transvenous sheets (Fig. 1C, D). A cutting sheet was used to dilatate and free the lead fragment (Fig. 1E–G). The patient was found to reach final radiological and clinical success regarding indication for leads removal. A film of the procedure is available at:

http://youtu.be/OiCqiJp7VAQ.

Figure 2. Real-time three-dimensional (3D) transoesophageal echocardio- graphy (TEE) showing the fragment of atrial lead placed in the right atrial appendage (A) and the temporary ventricular lead running crosswise (B) (Philips® iE33 with live 3D software;

X7-2t TEE probe) Figure 1. The 4 cm long atrial

lead fragment in right the atrial appendage and the tem- porary ventricular pacing lead with active fixation (A). Byrd Working Station was utilised.

Attempt to catch, reposition, and extract the lead fragment by lasso catheter (B). Stabili- sation of the free part of the lead fragment with the pig-tail catheter (C) allowed us to ca- tch it distinctly with the basket catheter (D). Dissection of con- necting tissue bridges using rotation of truncated internal sheet of Byrd Working Station (13 French) (E) allowed us to free the lead fragment (F).

Basket catheter, both sheets of the Byrd Femoral Work Station™ (Cook Vascular Inc.), and removed fragment of the lead with tip. Cutting catheter allowed dissection of connec- ting tissue from the distal part of the lead and its successful extraction (G)

C

G F

E

D A

A B

B

Cytaty

Powiązane dokumenty

During im- plantation, there is potential for the ventricular lead to be inadvertently inserted into LV, either through a patent forman ovale, ventricular septal defect, or

I would take issue with this definition as it is highly dependent on how the device is programmed: too short an AV delay that usurps control from the nor- mal AV conduction system

The purpose of the present study was to ex- amine pacemaker dependency (PD) in paced pa- tients during long-term follow-up and to evaluate the difference between patients with

Figure 2 Monitoring of the process of pulling on the cardiac walls during lead dissection: A ‒ fluoroscopy showing  the extraction of a high ‑voltage lead, adhesion of

Laboratory tests also demonstrated prolonged thrombin time of 196 s (normal range, 18–22 s), prolonged activated partial thromboplastin time of 66.4 s (normal range, 27–34

nous lead extraction of the abandoned lead with surgical treatment of the fistula in the axillary fossa, leaving the pacing system on the right side of the chest.. First,

We aimed to evaluate the utility of inflamma- tory markers (WBC count, CRP level) in the di- agnostic workup of infectious complications of permanent pacemakers and in differentiation

In FIGURE 1 on page 1228, “HR” should have read “OR” and, consequently, the abbreviation list below the figure should contain “OR – odds ratio” instead of “HR –