Guiding catheter extension facilities high speed rotational atherectomy for distal right coronary artery lesion

N/A
N/A
Protected

Academic year: 2022

Share "Guiding catheter extension facilities high speed rotational atherectomy for distal right coronary artery lesion"

Copied!
2
0
0

Pełen tekst

(1)

Creative Commons licenses: This is an Open Access article distributed under the terms of the Creative Commons 507

Attribution-NonCommercial-ShareAlike 4.0 International (CC BY -NC -SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/).

Image in intervention

Corresponding author:

Tomasz Pawłowski PhD, MD, Department of Invasive Cardiology, Center of Postgraduate Medical Education, Warsaw, Poland, phone: +48 691 020 998, e-mail: pawtom@gmail.com

Received: 21.02.2020, accepted: 11.05.2020.

Guiding catheter extension facilities high speed rotational atherectomy for distal right coronary artery lesion

Tomasz Pawłowski, Zbigniew Śliwiński, Paweł Modzelewski, Robert J. Gil

Department of Invasive Cardiology, Center of Postgraduate Medical Education, Warsaw, Poland

Adv Interv Cardiol 2020; 16, 4 (62): 507–508 DOI: https://doi.org/10.5114/aic.2020.101779

Extensive coronary calcification may affect proce- dural outcomes of coronary angioplasty, resulting in im- paired device deliverability, especially in distal locations of the stenosis [1]. In some cases, balloon predilatation is not enough for successful dilatation of the lesion and high speed rotational atherectomy is needed to perform the procedure. Due to the limited length of effective ro- tablation (advancer moving length – 7 cm), deliverability to the distal plaque location is impossible.

In the present case, we used a  90 cm length 7 Fr guiding catheter with 7 Fr guiding extension (Guideliner, Teleflex Co) to deliver a 1.25 mm Rotablator burr to the postero-lateral branch of the right coronary artery. Be- cause of the risk of burr stack on the first arterial bend, we placed the tip of the Guideliner below this bend and using the standard Dynaglide protocol, the burr was ad- vanced distally. After successful rotablation with the reg- ular protocol, the burr was removed using the Dynaglide protocol. After that, coronary stenting was performed without any obstacles.

The well-known limitation of the high speed rota- tional atherectomy system is the quite short range of

burr movement. Additionally, the artery’s bending pro- vides a risk of burr stack. In the present case, we used the Guideliner to overcome the mentioned issues [2]. Of course, one should remember that the internal diameter of the guiding extension has to be matched with the out- er diameter of the burr. Based on our experience, we can recommend checking the matching outside the patient.

A very important issue is the protocol of system ma- nipulation. We advise using Dynaglide protocol to ad- vance and remove the burr with careful manipulation of the burr at the tip of the guiding extension during remov- al just to prevent damage of the tip. On the other hand, the risk is comparable to the regular situation with the guiding catheter.

We think that the guiding extension for rotablation burr advancement is a valuable option for some distal lo- cations of a calcified lesion treated with coronary angio- plasty. Of course, it can also be used for tortuous arteries with high degree bends [3].

Conflict of interest

The authors declare no conflict of interest.

(2)

Tomasz Pawłowski et al. Guiding extension for ROTA

508 Advances in Interventional Cardiology 2020; 16, 4 (62)

References

1. Kobayashi Y, Okura H, Kume T, et al. Impact of target lesion cor- onary calcification on stent expansion. Circ J 2014; 78: 2209-14.

2. Fath-Ordoubadi MF, Fraser DG. Distal stent delivery with guide- liner catheter: first in man experience. Catheter Cardiovasc Interv 2010; 76: 102-11.

3. Vo M, Minhas K, Kass M, Ravandi A. Novel use of the GuideLiner catheter to deliver rotational atherectomy burrs in tortuous ves- sels. Case Rep Cardiol 2014; 2014: 594396.

Figure 1. A – Coronary angioplasty of the postero-lateral branch of the right coronary artery (arrow). B – Short guiding catheter (Launcher JR 4.0, 90 cm, Medtronic Co) with guide extension (arrow) and parking position of 1.25 mm burr is shown. C – Presentation of used devices – distal tip of Guideliner extension and rotablation burr are marked. D – Final result of coronary stenting

A

C

B

D

Cytaty

Powiązane dokumenty

Changes in coronary artery blood flow A total of 62 patients for whom the cTFC was evaluated were divided into two groups based on the difference between the postprocedural

First a 2.0 × 20 mm balloon catheter was inflated beyond the site of the occlusion in order to ensure protection from distal embolism and facilitate the guide extension catheter

Material and methods: We identified 35 consecutive patients with stable angina, who underwent virtual 3-Fr PCI using either hydrophilic-coated guiding catheters (Works TM ,

operators successfully intubated the right coronary ar- tery with a guiding catheter, morphology of the lesion. and presence of side branches reduced the

In addition, combinations such as a 5-Fr Heartrail TM guid- ing catheter (Terumo, Japan) and a 4-Fr KIWAMI ST01 Heartrail TM guiding catheter (Terumo, Japan), and a 5-Fr Works

We present a case of successful percutaneous retrieval of a damaged rotational atherectomy burr in a 74-year-old man who underwent a procedure on the right coronary artery..

Below we present a case of a patient with a long, high- ly calcified, critical stenosis of the right coronary artery diagnosed in our center which was treated with the use

Primary Stenting of Totally Occluded Native Coronary Arteries II (PRISON II): a randomized comparison of bare metal stent implantation with sirolimus-eluting stent implantation for