• Nie Znaleziono Wyników

Transvenous extraction of an implantable cardioverter-defibrillator lead looped and damaged in subclavian vein

N/A
N/A
Protected

Academic year: 2022

Share "Transvenous extraction of an implantable cardioverter-defibrillator lead looped and damaged in subclavian vein"

Copied!
1
0
0

Pełen tekst

(1)

www.kardiologiapolska.pl

Kardiologia Polska 2013; 71, 12: 1328; DOI: 10.5603/KP.2013.0340 ISSN 0022–9032

STUDIUM PRZYPADKU / CLINICAL VIGNETTE

Transvenous extraction of an implantable cardioverter-defibrillator lead looped

and damaged in subclavian vein

Przezżylne usunięcie uszkodzonej i spętlonej w lewej żyle podobojczykowej elektrody ICD

Andrzej Ząbek

1

, Barbara Małecka

1

, Jacek Lelakowski

1

, Małgorzata Hardzina

1

, Roman Pfitzner

2

1Department of Electrocardiology, John Paul II Hospital in Krakow, Krakow, Poland

2Department of Cardiosurgery, John Paul II Hospital in Krakow, Krakow, Poland

We present a case of damaged lead transvenous extraction in a 26-year-old patient after surgical correction of Fallot tetralogy at the age of nine, and after implantable cardioverter-defibrillator (ICD) implantation as secondary prevention at the age of 19. A one-coil passive fixation lead (Medtronic Sprint 6932) was implanted by sub- clavian venipuncture to the right ventricular apex. In follow-up, ad- equate interventions were noted and significant sinus bradycardia in Holter ECG was observed as a result of treatment with metoprolol and dronedaron. In 2011, an exchange of one-chamber to two-chamber ICD was performed. Active fixation atrial lead (Biotronik Setrox) was implanted by left subclavian venipuncture to the right auricle.

In X-ray examination performed after implantation, looping in the

left subclavian vein and significant pulling-up in heart cavities of the ventricular lead was diagnosed (Fig. 1A). Most probably the lead was pulled up in the terminal stage of atrial lead implantation during vascular sheath removal made without X-ray control. After 18 months of correct ICD functioning, episodes of short cycles in intracardiac ventricular electrograms were registered. In TTE, significant tricuspid regurgitation, related to pres- sure on the septal cusp of the tricuspid valve by stretched ventricular lead, was observed.

Therefore, TLE was planned. In left subclavian vein venography, vessel patency and the presence of ventricular lead loop outlining vessel lumen with atrial lead inside this loop was confirmed (Fig. 1B). Additionally, during the procedure atrial lead isolation damage in ICD pocket was diagnosed. The atrial lead was extracted by simple traction (Fig. 2A).

Next, extraction of the ventricular lead was performed with the use of a mechanical Cook set. Inserting metal wire inside the lumen was possible only up to half of the lead loop.

An attempt at lead unlooping was unsuccessful due to the strong adherence of the lead to the subclavian vein wall (Fig. 2B). Gradually, with the help of yellow and then white Byrd dilators, the lead was separated from adherences and the loop was straightened (Fig. 2C). Unlooping allowed us to remove and cut off the lead fragment with closed lumen outside the vein (Fig. 3A; fragments 2b and 2c). Then a metal wire was inserted into the lumen in order to stiffen the remaining lead fragment. The lead was separated from adherences along its whole length with the use of yellow and white Byrd dilators and removed (Fig. 2D). Guide wires for new lead implanta-

tion were inserted into the heart through a Byrd dilator internal lumen. New leads with active fixation to the right atrium and ventricle were implanted. An image of the removed lead fragments is presented in Figure 3. In the early postoperative period, the presence of fluid in the left pleural cavity was observed. During drainage, 900 mL of blood liquid was aspirated. In laboratory tests, a decrease in morphology was noted, however it not require blood transfusion. After 24 h the drain was removed. Further hospitalisation was successful.

Address for correspondence:

Andrzej Ząbek, MD, MSc, Department of Electrocardiology, John Paul II Hospital in Krakow, ul. Prądnicka 80, 31–202 Kraków, Poland, e-mail: andrzej_j_z@poczta.onet.pl

Conflict of interest: none declared

Figure 1. A. X-ray PA projection; B. Left subclavian veno- graphy

Figure 2. A–D. X-ray during lead extra- ction

Figure 3. A. All removed frag- ments of atrial and ventricular leads (1a and 1b — atrial proximal and distal; 2a, 2b, 2c — ventricular proximal, looped and distal with coil);

B. Distal ventricular with con- nective tissue; C. Ventri cular loop with tissue adherence to subclavian vein

B

B C

A

A

C

B

D

A

Cytaty

Powiązane dokumenty

The MyoPore (Greatbatch Medical, New York, United States) sutureless screw -in epicardial pacing lead was delivered to the target area with the steerable FasTac Flex

We aimed to evaluate the utility of inflamma- tory markers (WBC count, CRP level) in the di- agnostic workup of infectious complications of permanent pacemakers and in differentiation

Intravascular techniques for lead extraction including locking stylets and telescoping sheaths significantly reduce the risk of electrode rupture, but helix fracture and

Muhammet Dural, Eskisehir Osmangazi University, Cardiology Department, Büyükdere Mahallesi, 26040 Odunpazarı/Eskişehir, 26040 Eskisehir, Turkey, e-mail:

Cardiac magnetic resonance imaging confirmed the presence of hypertrabeculation with a two-layered structure of the endocardium with an increased noncompacted to compacted

Dual-chamber ICD implantation via a persistent left superior vena cava — use of an innominate vein for the placement of a right ventricular.. cardioverter-defibrillator

Chest X-ray showed a lead breakage at the level of the right clavicle and migration of the proximal fragment of the broken lead to the left pulmonary arteries (Fig.. There were

Due to the large size of the vegetation (the length: 71 mm) connected to the lead and its coarse texture (Fig. 1), the patient was qualified for complete system removal during