• Nie Znaleziono Wyników

Complicated Holter tracing with an incidence of inappropriate mode switch due to sensing abnormalities in a patient with dual-chamber pacemaker

N/A
N/A
Protected

Academic year: 2022

Share "Complicated Holter tracing with an incidence of inappropriate mode switch due to sensing abnormalities in a patient with dual-chamber pacemaker"

Copied!
3
0
0

Pełen tekst

(1)

INTERESTING ELECTROCARDIOGRAMS

Cardiology Journal 2011, Vol. 18, No. 1, pp. 94–96 Copyright © 2011 Via Medica ISSN 1897–5593

94 www.cardiologyjournal.org

Address for correspondence: Małgorzata Sobieszczańska, MD, PhD, Department of Pathophysiology, Wroclaw Medical University, Marcinkowskiego 1, 50–368 Wroclaw, Poland, tel/fax: +48 71 784 12 47, e-mail: sogood@poczta.onet.pl Received: 13.09.2010 Accepted: 29.09.2010

Complicated Holter tracing with an incidence of inappropriate mode switch due to sensing

abnormalities in a patient with dual-chamber pacemaker

Małgorzata Poręba

1

, Stefan Karczmarewicz

2

, Bartłomiej Szwarc

3

, Małgorzata Sobieszczańska

1

, Jerzy Krzysztof Wranicz

4

1Department of Pathophysiology, Wroclaw Medical University, Wroclaw, Poland

2Warsaw Education Center, CRDM Medtronic Poland, Warsaw, Poland

3CRDM Medtronic Poland, Warsaw, Poland

4Department of Electrocardiology, Medical University of Lodz, Lodz, Poland

Abstract

24-hour Holter monitoring of a 59 year-old man with DDDR pacemaker (programmed mode:

DDD) implanted for sick sinus syndrome with paroxysmal atrial fibrillation was performed one month after implantation, due to palpitations. Several episodes of rapid pacing of decreasing rate were detected. Intracardiac recording stored in the pacemaker memory had shown epi- sodes of atrial lead oversensing, which led to the mode switch, resulting in DDIR mode. Signals which caused oversensing were not seen in Holter tracing. (Cardiol J 2011; 18, 1: 94–96) Key words: DDD pacemaker, Holter monitoring, mode switch, atrial oversensing

Introduction

A 59 year-old male patient who had been im- planted with the Medtronic Kappa KDR 901 DDDR pacemaker for symptomatic sick sinus syndrome, underwent 24-hour Holter monitoring (24HM) due to palpitations, which occurred one month after implantation, especially when he moved his left arm. Programmed pacemaker (PM) parameters during 24HM are shown in Table 1. 24HM revealed:

numerous episodes of atrial fibrillation (AF) lasting from seconds to minutes, with the ventricular rate from about 70 beats/min to 150 beats/min, numer- ous premature supraventricle complexes and dual chamber pacing — DDD: Ap-Vp and As-Vp.

Several ECG stripes recorded in 24HM moni- toring could not be explained with PM programmed parameters. The examples are shown in Figures

1–3. Episodes started from rapid ventricular pac- ing, which could not be explained by the preceding ventricular rate, or by apparent atrial activity, which potentially might have triggered ventricular pacing.

Ventricular pacing was of continuously decreasing rate, and then some beats with sequential pacing started and continued with decreasing rate as well.

Episodes ended with the supraventricular (sinus?) beat and with markedly lower heart rate thereaf- ter. In a few episodes, some artifacts difficult to precise analysis could be seen.

The PM was interrogated with a programmer and 12 episodes of the high atrial rate with subse- quent mode switch interventions were revealed.

Analysis of the stored EGMs showed that detected high atrial rate episodes were actually due to atrial oversensing (Fig. 4). The patient’s description of events suggested that oversensing episodes and

(2)

95 Małgorzata Poręba et al., Complicated Holter tracing in pacemaker patient

www.cardiologyjournal.org

subsequent mode switch interventions were caused by the sensing of the skeletal muscles activity due to the unipolar-sensing programmed device. The decreasing rate of pacing was due to DDIR mode, which is routinely activated in Medtronic DDD pacemakers, as the result of the mode switch in- tervention, regardless of whether the sensor was activated before the intervention, or not. Since oversensing was caused by the patient’s physical activity, the sensor was activated just after the mode switch, and then (due to the halt of activity) deceleration occurred, resulting in a decrease of the pacing rate.

The problem was solved by reprogramming both atrial and ventricular sensitivity to bipolar.

Thereafter, it never recurred.

Discussion

Only a few papers have assessed the function of the single and double chamber pacemakers in the early post-implantation period [1, 2]. In the presented Table 1. Pacemaker parameters during Holter

monitoring.

Parameter Value

Mode DDD

Lower rate 60 bpm

Upper tracking rate 130 bpm

Paced AV delay 150 ms

Sensed AV delay 120 ms

Post ventricular 180 ms

atrial blanking

PVARP AUTO

Minimum PVARP 410 ms

Pacemaker mediated Off

tachycardia intervention

Mode switch On, 140 bpm

Atrial lead amplitude/ 3.5 V (0.4 ms)/0.5 mV;

sensitivity unipolar/unipolar

Ventricular lead 3.5 V (0.4 ms)/2.8 mV;

amplitude/sensitivity unipolar/unipolar

PVARP — post ventricular atrial refractory period

Figure 1. The beginning of the episode, with rapid ventricular pacing in spite of sufficient cardiac automaticity and conduction, then decreasing pacing rate; some inconclusive artifacts preceding the episode and occurring during the first part of the episode can be seen.

Figure 2. The beginning of the episode.

(3)

96

Cardiology Journal 2011, Vol. 18, No. 1

www.cardiologyjournal.org

case, 24HM was a helpful tool, allowing more ade- quate programming.

Normally, mode switch is interpreted as a sur- rogate marker of the total AF duration and their re- currence. However, in our patient it was used in an inappropriate manner due to oversensing [3, 4]. It seems reasonable that in modern PM, sensing in atrial lead should be bipolar. It has been reported that oversensing of ventricular far-field signals (tail end of the QRS complex) represents one the com- monest causes of false positive mode switching, whereas this is rarely caused by miopotentials, as in our case [4].

The additional goal of PM implantation, beyond treating arrhythmias, is to improve the quality of life, and in our patient this goal was not achieved after implantation [5]. Only 24HM and intracardiac stored electrograms helped to solve the problem and helped the patient.

Figure 4. Intracardiac recording from the pacemaker memory — the oversensing episode with the subsequent mode switch (marked as an ”MS”) can be seen.

Figure 3. The end of the episode.

Acknowledgements

The authors do not report any conflict of inte- rest regarding this work.

References

1. Wranicz JK, Chudzik M, Cygankiewicz I et al. The usefulness of 24 hour Holter monitoring in asymptomatic pacemaker patients in early post-implantation period. Folia Cardiol, 2006; 13: 390–395.

2. Wranicz JK, Chudzik M, Cygankiewicz I et al. Pacing and sens- ing disturbances in patients with DDD pacemakers in the early period after implantation. Acta Cardiol, 2006; 61: 289–294.

3. de Voogt WG, van Hemel NM, van de Bos AA et al. Verification of pacemaker automatic mode switching for the detection of atrial fibrillation and atrial tachycardia with Holter recording.

Europace, 2006; 8: 950–961.

4. Stabile G, De Simone A, Romano E. Automatic mode switching in atrial fibrillation. Indian Pacing Electrophysiol J, 2005; 5: 186–196.

5. Młynarski R, Włodyka A, Kargul W. Changes in the mental and physical components of the quality of life for patients six months after pacemaker implantation. Cardiol J, 2009; 16: 250–253.

Cytaty

Powiązane dokumenty

On April 28, the implantation of a Micra TM VR Transcatheter Pacing System (Medtronic, Dublin, Ireland) was carried out in the Electrophysiological Laboratory under

nous lead extraction of the abandoned lead with surgical treatment of the fistula in the axillary fossa, leaving the pacing system on the right side of the chest.. First,

Conflict of interest: Beata Średniawa — con- sultant: Medtronic, Zoll, Bayer, lectures fee for: Boehringer-Ingelheim, Bayer, Pfizer; Ewa Jędrzejczyk-Patej

The transthoracic echocardiography (TTE) revealed a mass in the left atrium (1.6 × 2.2 cm) with connective tissue peduncle attached in the area of fossa ovalis in the

Autopsy imaging revealed severe pericardial effusion, with the temporary pacemaker lead visualized beyond the right ventricular wall (red arrow); LV — left ventricle... A B

Paradoxical atrial undersensing was also re- ported in a single case of paroxysmal atrial flutter where device interrogation revealed atrial under- sensing of 5 mV flutter waves at

Methods: This was a retrospective analysis of the factors predictive of significant TR in 60 patients with systemic right ventricles following an atrial switch procedure for

We then reprogrammed the generator to a VVI stimulation mode for the last set of RF pulses in the lateral hepatic lesions and at a 50 bpm rate to limit the R on T phenomenon (Fig..