• Nie Znaleziono Wyników

of Experimental Parameters when testing the Hardness of Rubber

N/A
N/A
Protected

Academic year: 2021

Share "of Experimental Parameters when testing the Hardness of Rubber"

Copied!
8
0
0

Pełen tekst

(1)

R. Hodson*, S. Lackovic*

The Importance

of Experimental Parameters when testing the Hardness of Rubber

Hardness is one of the most widely measured properties used to charac­

terise rubber as it is a practical way of determining the degree of vulcanisation of a sample. Two scales are in general use throughout the world - the IRHD (International Rubber Hardness Degree) scale and the Shore scale. The two test methods use totally different indentor geometries, applied forces, test times and procedures. This paper looks at the instruments in question, studies the diffe­

rences between the tests and their relationship between scales where possible.

It also highlights the merits of each instrument and test type.

Key words: hardness of rubber, IRHD and Shore test methods, experimen­

tal procedure

Znaczenie parametrów badania podczas pomiaru twardości gumy

Twardość jest najczęściej mierzoną właściwością charakteryzującą gumę i stanowi praktyczny sposób oznaczenia stopnia zwulkanizowania próbki.

Ogólnie na świecie używane są dwie skale: IRHD (International Rubber Hard­

ness Degree) i Shore’a. Te dwie metody badania całkowicie różnią się pod względem geometrii czujnika, przyłożonej siły, czasu pomiaru i procedur. W ar­

tykule przedstawiono różnice między tymi badaniami i możliwe relacje między skalami (twardości). Wyjaśniono także zasadę działania każdego z aparatów i typ pomiaru.

Słowa kluczowe: twardość gumy, metody IRHD i Shore’a, procedury ba­

dania

1. Historical perspective

According to Bassi et al. [1] the Shore instruments had historical priority over the IRHD instruments by more than 30 years. Gurney [2] reported both instru­

ments in use by the early 1920’s, together with other spring and dead load (weight) variants. Results from the spring type varied with the user (Gurney [2], The Rub­

ber Age [3]). This led to the adoption of the dead load instrument where the indentation depth was largely user independent. After Scott [4] stressed the need for a standard to give results some common meaning in 1935, the first British Standard (BS) was introduced in 1940. At the same time, Scott and Newton [5] reported on a reliable pocket type hardness gauge that con­

formed to this new standard. After a comparison with the Shore A Durometer, they concluded that the advan­

tage was always with the BS Hardness Meter. Work was then carried out looking at different instrument types

* H.W. W allace & Co. Ltd, 172 St Jam es’s Road, Croydon, CR9 2 HR, England

(Daynes and Scott [6]) and the new standard (Scott [7]).

They both agreed that there was some correlation be­

tween the Shore A and BS hardness scales. The accu­

racy of a range of hardness testers (Newton [8]) was investigated, concluding that the main limitations were associated with the operator. Instruments with a spheri­

cal indentor and foot gave the smallest errors; the lar­

gest errors were associated with the Shore durometer.

The largest source of variation reported by Scott [9]

was the lack of agreement between laboratories.

The Micro Hardness Tester, a (l/6th) scaled down version of the IRHD Dead Load Hardness Tester was introduced in the 1950’s to test thinner and small pro­

duction samples. Scott and Soden [10] reported results comparable between the Micro and Dead Load tests, with only a few degrees difference noted for rubbers of greater than 65° hardness.

Several papers [1,11,12,13] published in the 1960’s, 70’s and 80’s and two books, Rubber and Plas­

tics Testing [14] and Physical Testing of Rubbers [15]

stated that the most widely used instrument was then the Shore A type even though the IRHD method pro­

duced more repeatable results between operators, with higher accuracy, reproducibility and precision. How-

SćcKLfortt&Uf nr 3 maj - czerwiec 2002 r. TOM 6

(2)

fsc (fumy _____________

ever, Shore A has a less critical dependence than IRHD on sample thickness (Bassi et al [1]). Comparative work by Brown and Soekarnein [16] (1991) between the IRHD Dead Load, IRHD Micro and Shore A instru­

ments demonstrated that inter-laboratory repeatability was likely to be best for the IRHD Dead Load and Micro instruments. In 1993, Briscoe and Sebastian [17]

analysed the durometer indentation, providing an ap­

proximate relationship between IRHD and Shore A of (IRH ~ Ha+ 4), although this is very dependent on the sample compound.

Many contemporary hardness testers have im­

proved accuracy due to the automatic nature of the test, requiring minimal operator intervention. Bench mounted instruments (IRHD Dead Load and Micro and Shore A scales) produce the most repeatable and reli­

able results. Pocket meters are much improved, but do rely entirely on the operator’s hand pressure and reli­

able angular application for repeatable results (varia­

tions can be extreme).

In recent years, there has been an increased interest in the Shore M instrument, which has now been in­

cluded in the latest revision of ASTM D2240-00. The results from such an instrument are not comparable to those obtained from a Shore A instrument.

2. D ifferen ces betw een IR H D and S h o re te st methods

S am p le T h ic k n e s s /In d e n to r Profile

There are four IRHD methods in use: the Normal­

hardness test (Dead Load), High-hardness test, Low­

hardness test and the Micro-test. The Normal test is used for samples greater than or equal to 4 mm thick and preferably used for rubbers in the 35 to 85 IRHD range (but with reservation, may be used for the 30 to 95 IRHD range). The High-hardness test is used for testing samples of the same dimensions as the Normal test, but in the 85 to 100 IRHD range. The Low-hard­

ness test is used for testing samples greater than or equal to 6 mm thick and hardnesses in the 10 to 35 IRHD range. The Micro tests samples less than 4 mm thick and is used for rubbers in the 35 to 85 IRHD range (but with reservation, may be used for the 30 to 95 IRHD range). All four methods use a spherically tipped indentor. The diameters of the ball indentor and foot vary between methods. The applied forces are the same for the Normal, High and Low tests, with only the Mi­

cro test requiring the application of smaller forces. It is worth noting that the IRHD scale is non-linear.

The Shore range of hardness testers incorporates eight scale types: A, B, C, D, DO, O, OO and M. These are used for testing a wider range of materials. The A

scale is used for soft rubbers and elastomers and type C for medium hard rubbers and plastics; both types use a truncated cone shaped indentor. Type A is the most commonly used rubber scale. Type B is used to test moderately hard rubbers and type D is used to test hard rubbers and plastics. Both of these use a 30° indentor.

Type DO is used for very dense textile windings, type O is used for soft rubbers and medium density textiles and OO is used for low density textile windings and sponge. These three use a 3/32 inch spherically ended indentor. All types require samples more than 6 mm thick (unless it can be proved that smaller samples give equivalent results). Type M is used for testing thin and irregular rubbers of hardness in the range 20 to 90 and uses a very small round tipped indentor. Thinner sam­

ples may be used, although the support table starts to affect the value as thickness falls as the indentor pene­

trates the sample.

Indentor Force

Indentor forces vary between instruments. Type A, B and O use the same spring force and it is recom­

mended that a force equivalent of 1 kg (9.8 N) is applied to the durometer presser foot to ensure that the spring force is repeatably overcome. Type C, D and DO use the same spring, requiring a force equivalent of 5 kg to overcome the spring. Type OO uses a different spring and requires 400 g. Type M requires a force sufficient to overcome the calibrated spring force. All Shore scales are linear.

The IRHD method is based on the use of dead loads (weights). A foot is used to hold the sample in place with a force of 8.3 N (Dead Load) or 235 mN in the case of the Micro hardness tester. A primary load of 0.3 N (Dead Load) or 8.3 mN (Micro hardness tester) is then applied for 5 seconds, providing a datum position.

A secondary load of 5.4 N (Dead Load) or 145 mN (Micro) is then applied for 30 seconds. The incremental displacement from the datum is measured and con­

verted to an IRHD value (a non-linear scale defined in the standard). The full-range displacement of (Normal) Dead Load is 1.8 mm; the Micro uses 0.3 mm.

In contrast, the Shore instruments use calibrated springs. For example, the Shore A scale spring force varies from 0.5 N to 8.1 N (over the full displacement) and the Shore M scale from 0.3 N to 0.8 N. The presser foot applies a force sufficient to overcome the spring force. Once the presser foot contacts the sample the indentation depth is recorded after a pre-set dwell time;

the standard ASTM dwell times are 1 and 3 seconds.

The DIN standard uses 3 seconds, since the reading is usually still changing appreciably after 1 second. The force increases linearly with indentor displacement (full range is 2.5 mm for the A scale and 1.25 mm for the M scale).

The IRHD scale was set in 1948 to correspond to the Shore scale, in that a high number indicates a hard rubber and a low number indicates a softer rubber. The original Micro hardness test was designed to be a scaled

TOM 6 maj - czerwiec 2002 r. £CaAta*ptenxfi nr 3

(3)

down version of the Normal Dead Load test (displace­

ments in the ratio 6 to 1). The forces applied were in the ratio 36 to 1. Therefore if the limited thickness sample tested in the case of a Micro instrument is 1 /6th of the thickness of the Dead Load piece, l/6th of the result is obtained. Scaling is set so that the same result should be obtained from both instruments. Similarly, results from the Normal, High and Low Dead Loads show correla­

tion.

The Shore M test was not designed as a scaled down version of the Shore A test, but merely as an instrument that was capable of testing smaller samples.

It uses an unrelated indentor and spring so there is no easy relationship between the two instruments.

3. E xperim ental Proce­

dure

Only bench mounted instruments were used to ob­

tain the experimental results; however, the conclusions drawn will also be relevant to hand held instruments.

All instruments were calibrated before starting and the calibration was rechecked at the end. A standard tem­

perature of 23±2°C was used. The Shore instruments were set to both 1 and 3 second dwell time (since the results from these times differ). Test times are defined by the standard for the IRHD instruments (5 and 30 se­

conds). Each flat sample was tested in 5 different places and curved samples were tested as specified below.

Standard Wallace test blocks (varying compounds of natural rubber, supplied by Malaysian Rubber Pro­

ducers Research Association) for both dead load and micro instruments were used to provide comparative results for each instrument.

The ISO standard (ISO 48:1994) for the IRHD mi­

cro instrument allows 1mm thick samples to be used but the preferred thickness is 2 ± 0.5 mm. In the case of the Shore M instrument, a 1.25 mm sample is specified although thinner samples may be tested if it is known that results equivalent to the 1.25 mm values are ob­

tained. Tests were performed on a range of thinner ma­

terials. The Shore standard (ASTM D2240 - 00) sug­

gests that samples be plied to increase their effective thickness; this was done to determine the effect of vary­

ing sample thickness. This was extended to similar work on the IRHD dead load and Shore A instruments.

The IRHD standard thickness is 8-10 mm whilst the Shore A is 6 mm. A selection of thinner samples were tested and plied to determine the effect of varying sam­

ple thickness.

Tests were carried out on the Dead Load and Shore A instruments at a raised temperature to determine any effect.

Curved samples, such as ‘O’ rings, are often tested and the effect of testing these on different instruments was investigated. ‘O’ rings (of varying outer and core diameters) were placed on a specially designed table so that they could be accurately displaced laterally to de­

termine the effect of testing away from the top dead centre.

4. Results

Standard Test Blocks

The standard test blocks gave repeatable results us­

ing Micro and Normal Dead Load instruments. The 1 and 3 second dwell times (Shore A and M) also pro­

duced repeatable, results. The Dead Load readings were consistently a few units higher than the Shore A read­

ings (see figure 1) over the range tested (40 - 90 IRHD).

However, there was an increasing tendency for the Shore M results to diverge from the Micro IRHD result with increasing hardness values, as shown in figure 2.

Figure 1. D e a d L o a d & S h o r e A r e s u lts

Figure 2. I R H D M i c r o & S h o r e M r e s u lts

Effects of Thickness

The IRHD Dead Load and Shore A instruments were used to test the standard Wallace Micro samples (2 mm thick). As expected, the results differed from those obtained using the specified instrument for the sample thickness, i.e. the Micro IRHD and the Shore M types. The softer rubbers, and also the IRHD instru­

ments, exhibited greater differences between the micro and macro instrument results. For the hardest rubber

nr 3 maj - czerwiec 2002 r. TOM 6

(4)

(76-79 IRHD), the IRHD Dead Load instrument gave a very close value to that of the IRHD micro (see figure 3). The Shore A instrument read a few units lower as expected from the tests carried out on the standard blocks (above) but the readings were closer between the Shore A and Shore M instruments. The Shore A value of the hardest rubber differed by only 1 unit to the Shore M value (figure 3).

Figure 3. Testing 2 mm thick samples on micro and macro instruments

Once the 2 mm thick samples were plied to 8 mm thick (the standard thickness required for the IRHD Dead Load tester) the results came within the specified tolerances of the test pieces. Increasing the thickness further made little difference to the result. These results may be seen in figure 4. Using the Shore A instrument, the results after plying pieces to provide a 6 mm thick sample (the standard thickness for Shore A) was not equivalent to the value when tested with a Shore M instrument.

Figure 4. Decreasing hardness with increasing sample thickness

In contrast, the standard Dead Load blocks of 8 mm thick, when tested on the Micro IRHD and the Shore M instruments, tended to give approximately the same re-

Figure 5. Macro&Micro Instruments being used to test Dead Load samples

Figure 6. Increasing the thickness of neoprene sample

Figure 7. Increasing the thickness of a nitrile sample

suits as the Dead Load IRHD and Shore A instruments, as shown in figure 5.

Various thinner samples were used with the micro instruments. Up to 5 pieces of neoprene (0.6 mm thick­

ness) were plied, taking the thickness of the sample into the standard tolerance region (and slightly beyond).

Both the IRHD Micro and Shore Micro showed a con­

tinual decrease in hardness with increasing thickness

TOM 6 maj - czerwiec 2002 r. Sta& tM ienxf, nr 3

(5)

(see figure 6). As before, the IRHD hardness values were consistently higher than the Shore M values. For a nitrile sample, the readings at the initial thickness of 1.5 mm (within the tolerance given in the standard) were similar between instruments. The Shore M results re­

mained constant during the thickness increase but the IRHD Micro instrument showed a decrease in hardness with increasing thickness to 4.5 mm as shown graphi­

cally in figure 7. In the case of a sample of silicone, the Shore M results were consistently lower than the IRHD Micro values, but both instruments exhibited a decrease in hardness of 1 unit, when doubling the thickness of the sample from 0.9 mm to 1.8 mm (within the stan­

dard) as shown in figure 8.

Figure 10. Differences in testing an O Ring of core diameter 2.5 mm on IRHD micro and Shore M

Figure 8. Increasing the thickness of a silicone sample

Effect of Temperature

Raising the temperature by 10°C appeared to make little difference to the results from the IRHD Dead Load on the standard test blocks (natural rubber compound).

However, slightly lower values were observed on the harder samples tested on the Shore A instrument. These results may be seen in figure 9.

Figure 9. IRHD Dead Load&Shore A Hardness Tests at room temperature and 32°C

Effect of Curved Surfaces

The smaller diameter ‘O’ rings were laterally dis­

placed in increments of 0.25 mm. The larger curved sur-

Effects of repeated testing in one location

The same spot was repeatedly measured. There was a decrease in hardness with increasing number of tests shown on both instruments. However, the Micro IRHD instrument gave more stable results, with only a total decrease of 1.5 IRHD over 28 tests on a rubber of ap­

proximately 65 IRHD. In contrast, the Shore-M instru­

ment results decreased much more rapidly before level­

ling out. The total decrease was approximately 5 units over 28 tests on the same sample. It was also noted that results produced from 1 second dwell time were consis­

tently lower than those produced with a 3 second dwell

Figure 11. Repeated testing in one location

S ta & tw i& ity nr 3 maj - czerwiec 2002 r. TOM 6

(6)

śćjum y _____________

time on the same sample. These results are represented graphically in figure 11.

5. Discussion

From the results it is clear that there is a correlation between the Dead Load and the Micro IRHD instru­

ments. This is apparent when the IRHD Dead Load result of plied micro samples corresponds with the standard result on an IRHD Micro instrument. In contrast, the same cannot be said for the Shore A and M scales.

The results indicate that the thickness of the sample used on the IRHD Dead Load affects the result more than on the Shore A, in agreement with Bassi et al [1].

In general, when the nitrile, neoprene and silicone samples were plied, a trend of decreasing hardness with increasing thickness was observed. Some differences were noted and it appears that different rubber types influence the results in slightly different ways.

Generally, the micro instruments can be used for testing both micro and macro samples, whilst the macro instruments (IRHD Dead Load and Shore A) are better for macro samples. Indeed, many people now use the Micro IRHD instead of the Dead Load instrument.

Flatter curves are produced with the IRHD Dead Load, Micro and the Shore A instruments when testing curved samples, implying that there is less critical de­

pendence on accurate sample positioning with these.

Since the graphs produced when using the Shore M instrument are more peaked, it is important to accu­

rately place the sample (to within ~ 0.1 mm). However, this is controlled when using an instrument attachment to centralize ‘O’ rings, but remains important when testing curved shapes that cannot be held accurately in such an attachment.

Increasing the temperature by approximately 10°C appears to make a greater difference to harder natural rubber samples only on the Shore A and little difference using the IRHD Dead Load instrument.

It is clear that repeated testing at the same location makes an appreciable difference to the results. This is more apparent when using the Shore M instrument. It is important to ensure that the sample is displaced be­

tween tests - this can be difficult for small samples.

It is interesting to note that the results obtained from the Shore instruments with a dwell time of 3 se­

conds differ from those obtained using 1 second. There­

fore, for Shore instruments, although different timings are unimportant for comparative work, it is important that the timing is accurate and repeatable. The time required (35 seconds specified by the standard) for an IRHD test places the IRHD instruments at a disadvan­

tage. However, work by Wallace indicates that this time can be reduced by a predictive technique, taking it into direct competition with the Shore timing, i.e. 3 seconds.

It has been observed that skilled operators can ‘pre­

dict’ the hardness of a sample much earlier in the meas­

urement cycle. For this to be the case, there has to be a common family of hardness curves for a majority of

sample types and ambient measurement conditions.

This would allow a general equation to be derived, which in turn would lead to a specific, predictable curve for any one measurement. This curve could then be used to predict the 30 second value at a much earlier point in the measurement cycle.

6. E xperim ental proce­

dure for predicting sam­

ple hardness

Both micro hardness and dead load instruments were used to perform tests on a variety of samples, in the hardness range 30-95 IRHD. The results were re­

corded every 0.5 seconds, giving 60 stored data points for each test, which were then logged on a PC. The standard load times of 5 and 30 seconds were used.

All data were plotted and a random selection of measurements was chosen for a variety of samples and temperatures. The data were normalised and superim­

posed to test the assumption of the similarity of the curves’ shape. The normalised data were then used to fit a general equation of the form:

y = a + b tc, where c is a constant (1)

Each sample will have a unique solution to equa­

tion (1). For each sample, two equations in the form of equation (1) can be created, using two t (time) values ( t i and t2) with their corresponding y (hardness) values (yi

and y i ) . These can then be solved simultaneously to

provide unique values of a and b for a particular sam­

ple. These values can then be substituted into equation (1) to give a curve for that sample.

The general equation was fitted at different t values to see which two values of t gave the best fit for the majority of samples. Although it was desirable to fit as small a t value as possible, it was also important for the equation to fit as large a number of samples as possible.

This ensured a high degree of accuracy whilst giving the greatest possible time reduction. The equation, along with the chosen t values, was re-checked using a selection of data.

For each sample, the equation and two y values at given t values, can be used to produce a ‘predicted 30 second’ value. This was compared for each sample to the actual value measured after the full 30 seconds du­

ring the same test on the same sample. This gave a measurement of prediction accuracy.

The primary load provides a datum point from which to measure the indentation depth. The standard specifies 5 seconds; investigations of this time were carried out.

7. Results

The investigation was carried out on both Wallace Dead Load and Micro Hardness Instruments.

TOM 6 maj - czerwiec 2002 r. Sta& totnenfy nr 3

(7)

As each sample was tested to the standard 30 se­

cond load time, it was possible to calculate a predicted 30 second value using the previously agreed ti and t2 values with their corresponding yi and y2 values. This was after only 6 seconds (t2) of secondary load time.

This enabled the actual 30 second values and the pre­

dicted 30 second values to be compared. Out of the 380 samples measured, 366 of the predicted values were in agreement with the actual 30 second values to within 0.5 IRHD. The samples within this agreement included samples from the entire hardness range (30-90 IRHD) as well as those of non-standard thickness and those that were tested at varying temperatures. The 14 sam­

ples that did not agree to within 0.5 IRHD did, however, agree to within 1 IRHD. This was found to be as a result of the sample not being completely flat whilst being tested.

The compounds that were tested are listed in table 1 and covers the full range of hardnesses tested, as well as including the non-standard samples used. Each of the samples was also tested more than once to ensure pre­

diction repeatability.

Table 1. Samples Tested

Sample Type Micro Dead Load

Chloro compound

y y

CR compounds

y y

EPDM compounds

y y

EPM compounds

y

FKM compounds

y

H. W. Wallace calibration blocks

y y

H. W. Wallace samples

y y

HR compounds

y

NR compounds

y y

NBR compounds

y y

Nitrile/PVC compound

y

Polyurethane compounds

y y

Q compounds

y y

Thiokol compound

y

In figure 12, a selection of samples of varying hard­

nesses are plotted, indicating their actual 30 second point as compared to their predicted 30 second value. It can be seen that there is good agreement between the predicted and actual endpoints over the entire hardness range, 30-95 IRHD.

A typical graph of actual and predicted data can be seen in figure 13. The sample shown here is neoprene, and it demonstrates the predictability of the indentation curve with respect to time. The graph shows the stan­

dard 30 second data (illustrated in fig. 13 from 2.5 se­

conds for clarity of the curves shape) along with a pre­

dicted curve, which begins at 6 seconds and culminates

Figure 12. Graph demonstrating the agreement be­

tween the actual and predicted 30 second endpoints

Figure 13. Graph showing data collected from a sample of neoprene, with its predicted curve from 6 seconds

Investigations were carried out to determine the effect of reducing the primary load time. It was found that reducing the standard 5 second primary load time to 1 second made no significant difference to the end­

point.

8. Discussion

This method provides a significant reduction in the test cycle time in situations where it is not necessary to adhere strictly to the standard, for example, in-house comparative testing. A time reduction may be benefi­

cial to users for two reasons - the same number of tests can be carried out more quickly or more tests can be earned out in the same time.

When a new sample type is used, the prediction validity can be proved by the user. As well as running one test to the reduced time period and producing a pre­

dicted endpoint, a test can be carried out to the full 30 seconds. In this way, a comparison can be drawn as to the reliability of the prediction.

The equation and mathematics used to fit the curve, and hence predict the hardness value, is incorporated as

S&zd&Mueny nr 3 maj - czerwiec 2002 r. TOM 6

(8)

śe gumy _____________

an additional extra in to the newly launched Wallace Data Collection Software.

9. Conclusion

This paper has taken a historical look at the IRHD and Shore Hardness measurement instruments as well as discussing and emphasising the fundamental diffe­

rences between the most common instruments used for rubber and elastomer hardness characterisation. The different various instruments exhibit advantages and disadvantages with certain sample types. IRHD instru­

ments are preferred for non-destructive testing and the Micro IRHD is generally a better choice for testing curved surfaces. The Shore A instrument is preferable for testing non-standard thickness samples and when shorter test cycle times are required. Accurate and re­

peatable timing is critical to allow Shore A and M in­

struments to provide consistent and comparable results.

The standard 35 second IRHD test time (5 second primary load and 30 second secondary load time) for a rubber sample can be significantly reduced by fitting a general equation of the form given in Equation (1).

The equation can be solved for each sample to produce an accurate prediction of the ‘final’ value with a high confidence level. Furthermore, the standard 5 second primary load time can be reduced to 1 second.

The prediction accuracy required determines the minimum test time. To ensure an agreement between the predicted value and the actual value to within 0.5 IRHD, the total test time can be reduced to 7 seconds (1 second primary time and 6 seconds secondary load time). For less accuracy, such as agreement within 1 IRHD between the predicted and actual values, the total test times can be reduced to 5 seconds. This time can be reduced still further but with less accuracy between the actual and predicted endpoints.

References

1. Bassi A. C., Casa F. & Mendici R., Polymer Testing 7, 165 (1987)

2. Gurney H. P, India Rubber Journal 497 (1921) 3. A New Rubber Hardness Tester ', The Rubber Age

29, 242 (1939-40)

4. Scott J. R., Transactions I. R. I. 11, 224 (1935) 5. Scott J. R. & Newton R. G., Journal of Rubber

Research 9, 91 (1940)

6. Daynes H. A. & Scott J. R., Journal of Rubber Research 12, 94 (1943)

7. Scott J. R., Journal of Rubber Research 17, 145 (1948)

8. Newton R. G., Journal of Rubber Research 17, 178 (1948)

9. Scott J. R., Transactions I. R. I. 27 (5), 249 (1951) 10. Scott J. R. & Soden A. L, Proceedings of the Inter­

national Rubber Conference, Washington, Paper 22, pp. 170-176(1959)

11. Price K., Progress of Rubber Technology 42, 59 (1979)

12. Warner J.C. & Jerdonek J.A., European Rubber Journal & Urethanes Today 162, 11 (1980) 13. Chang W. V. & Sun S.C., Rubber Chemistry and

Technology 64, 202 (1991)

14. „Rubber and Plastics Testing', Klucklow, pp. 153- 162, 1963

„Physical Testing of Rubbers”, J. R. Scott, pp. 91- 110, 1965

15. Brown R. P. & Soekarnein A., Polymer Testing 10, 117(1991)

16. Briscoe B.J. & Sebastian K.S., Rubber Chemistry and Technology 66, 827 (1993)

Informujemy, że od dnia 0 1 .0 7 .2 0 0 1 Instytut Przemysłu Gumowego „Stomil”

w Piastowie reprezentuje w Polsce firmę:

H . W . W A L L A C E S . C o . L t d

producenta urządzeń i aparatury badawczo-pomiarowej takiej jak:

• twardościomierze

• reometry (wulkametry) H.W. Wallace S. Co. Ltd

, 172 St. James’s Road

9ę Croydon, England CR9 2HR

• Inne Tel.: 44f0) 20 8686 4954

Fax.: 44 (0) 20 8688 4441

E-mail: hwwallace@compuserve.com Web: www.hwwallace.co.uk

TOM 6 maj - czerwiec 2002 r. SC aatw tienfy nr 3

Cytaty

Powiązane dokumenty

The assessment of the impact of lighting on pedes- trian safety indicates that pedestrian crossings are a particularly important element of road infrastructure in this

Przedstawiona stra- tegia dawkowania jest tym istotniejsza w praktyce, gdy w co- raz większej populacji pacjentów stwierdza się dodatkowe czynniki zmieniające

Trudno nie zgodzić się z Autorem monografii, który podkreśla, że główną rolę w procesie literackim odgrywają maski jako mistyfikacje, zaś odpowiednie spoj- rzenie na

Losonczi [9] proved the stability of the Hosszú equation in the class of real functions defined on the set of all reals and posed the problem of the stability of this equation in

Celem przeprowadzonych przeze mnie badań jest analiza języka mediów i ukazanie technik stosowanych przez dziennikarzy w celu wywołania określonego wrażenia odbiorcy a przez

The aim of experimental part was to determine the effect of the added amount of HNT additives into LDPE material on the change of hardness that mechanical

5 and 6 Ottoman vs Christian traces of paint and coatings: searching for evidence to solve issues of interpretation, Tower of the Winds (photo credit: Greek Ministry of Culture