• Nie Znaleziono Wyników

Przemieszczenie się elektrody rozrusznika serca powodujące brak skutecznej stymulacji komorowej

N/A
N/A
Protected

Academic year: 2022

Share "Przemieszczenie się elektrody rozrusznika serca powodujące brak skutecznej stymulacji komorowej"

Copied!
3
0
0

Pełen tekst

(1)

www.fc.viamedica.pl

Folia Cardiologica 2014 tom 9, nr 2, strony 179–181 Copyright © 2014 Via Medica

ISSN 2353–7752

praCa KazuIStyCzNa

179 Address for correspondence: lek. Anna Rydlewska, Klinika Elektrokardiologii, Instytut Kardiologii, Collegium Medicum, Uniwersytet Jagielloński, Szpital Jana Pawła II, ul. Prądnicka 80, 30–611 Kraków, e-mail: annarydlewska@op.pl

Spontaneous pacemaker distal lead tip migration resulting in lack of ventricular stimulation

Przemieszczenie się elektrody rozrusznika serca powodujące brak skutecznej stymulacji komorowej

Anna Rydlewska

1

, Wojciech Suślik

2

, Barbara Małecka

1

, Andrzej Ząbek

1

, Jacek Lelakowski

1

1Department of Electrocardiology, Institute of Cardiology, Jagiellonian University Medical College, John Paul II Hospital, Kraków, Poland

2Department of Cardiac and Vascular Diseases, Institute of Cardiology, Jagiellonian University Medical College, John Paul II Hospital, Kraków, Poland

Abstract

We present a case of spontaneous distal lead tip migration resulting in lack of ventricular stimulation diagnosed over a year after implantation. Ventricular lead tip was dislodged through tricuspid valve and fixed itself to right atrium wall.

Key words: heart stimulation, pacemaker follow-up, stimulation disturbances

(Folia Cardiologica 2014; 9, 2: 179–181)

Case report

88-year-old male with a history of ischemic cerebral in- farction, arterial hypertension and type 2 diabetes mellitus was admitted to hospital after transient ischemic attack for further neurological diagnostics. ECG at admission showed ineffective ventricular stimulation (Fig. 1).

A DDDR pacemaker with active fixation leads was implanted 14 months previously for sick sinus syndrome.

The procedure was complicated by pneumothorax. After the implantation patient did not present for a follow-up visit.

Telemetric pacemaker control confirmed lack of ventricular stimulation. It also showed that both ventricular and atrial channel stimulation resulted in atrial stimulation with excellent electrical parameters but with different P wave morphology (Fig. 2). Excessive lead bending at tricuspid valve level was present in transthoracic echocardiography

(Fig. 3). Chest X-ray confirmed that both lead tips were positioned in right atrium (Fig. 4). Patient was scheduled for transvenous ventricular lead extraction.

Discussion

Pacemaker lead tip migration is a rare complication of implantation procedure [1]. Most commonly, it concerns atrial leads. The risk factors of dislodgement are: implan- tation in low volume centre, inexperienced operator (less than 25 implantations per year), chronic heart failure as an indication to implantation, using passive or atrial leads [2]. It is a complication most commonly classified as early, diagnosed in perioperative period [3]. In our case, diag- nosis was made over a year after implantation. However, most probably the displacement took place early after implantation, before the lead tip got endothelialized in the

(2)

180

Folia Cardiologica 2014, tom 9, nr 2

www.fc.viamedica.pl surrounding tissues. Another interesting aspect of the case

is that the lead tip migrated through tricuspid valve and spontaneously fixed itself to right atrial wall with excellent stimulation parameters. The only known risk factor of displacement in this case, was performing implantation in a low volume centre. The lead that got displaced was not Figure 1. ECG at admission — ineffective ventricular stimulation

Figure 2. Atrial stimulation in both ventricular and atrial channels with different P wave morphology

Figure 3. Transthoracic echocardiography — excessive lead ben- ding at tricuspid valve level

Figure 4. Chest X-ray — both lead tips visible in right atrium

(3)

www.fc.viamedica.pl 181

Anna Rydlewska et al., Ineffective ventricular stimulation only a ventricular lead, but also it had an active fixation

system. Lack of pacemaker follow-up after discharge and most of all lack of chest X-ray at discharge from hospital resulted in delay in diagnosis. An examination, which in this case was the first one to raise questions on leads position, was echocardiography. It is an examination that is unjustly often omitted in pacemaker patients, although it might give insight not only into lead position, but also right and left ventricle performance and tricuspid valve function. Luckily for the patient, he mainly demanded atrial

stimulation, but ineffective ventricular stimulation might be a potentially fatal complication. In the described case, the complication could have been diagnosed using such basic and widely available examinations as standard ECG and standard chest X-ray, none of which were performed in a pacemaker patient for over a year.

Conflict(s) of interest

None declared.

Streszczenie

Zaprezentowano przypadek dyslokacji końcówki elektrody komorowej zdiagnozowanej ponad rok po implantacji roz- rusznika serca. Elektroda komorowa przemieściła się przez zastawkę trójdzielną i zaklinowała się samoistnie na wolnej ścianie prawego przedsionka.

Słowa kluczowe: stymulacja serca, zaburzenia stymulacji, kontrola rozrusznika

(Folia Cardiologica 2014; 9, 2: 179–181)

References

1. Aggarwal R.K., Connelly D.T., Ray S.G. et al. Early complications of permanent pacemaker implantation: no difference between dual and single chamber systems. Br. Heart J. 1995; 73: 571–575.

2. Kirkfeldt R.E., Johansen J.B., Nohr E.A. et al. Pneumothorax in cardiac pacing: a population based cohort study of 28860 Danish patients.

Europace 2012; 14: 1132–1138.

3. Szczeklik A., Tendera M. (ed.). Kardiologia. Wydawnictwo Medycyna Praktyczna, Kraków 2009: 473–481.

Cytaty

Powiązane dokumenty

In FIGURE 1 on page 1228, “HR” should have read “OR” and, consequently, the abbreviation list below the figure should contain “OR – odds ratio” instead of “HR –

The data used in the analysis included the following: (1) patients’ demographic baseline characteristics (date of birth, age at primary DDD implantation, age at ad- ditional

Muhammet Dural, Eskisehir Osmangazi University, Cardiology Department, Büyükdere Mahallesi, 26040 Odunpazarı/Eskişehir, 26040 Eskisehir, Turkey, e-mail:

Dual-chamber ICD implantation via a persistent left superior vena cava — use of an innominate vein for the placement of a right ventricular.. cardioverter-defibrillator

Nowadays approximately 90% of patients being considered for MCS receive a left ventricular assist device (LVAD) as a bridge to transplantation or destination therapy for those

Heart perforation by pro-MRI right ventricular lead in a 26-year-old woman.. Perforacja serca przez elektrodę komorową pro-MRI u

A 46-year-old Caucasian male patient with brady-tachycardia syndrome and a presentation of sinus bradycardia with oc- casional, atypical atrial flutter episodes had received the DDD

Poszukując mechanizmu powstania niedo- mykalności zastawki trójdzielnej, stwierdzono przyleganie jej płatka przegrodowego do przegrody międzykomorowej, za- równo w skurczu,