Design and characterization of zero magnetic field chambers for high efficiency neutron

polarization transport

Dadisman, Ryan; Shen, Jiazhou; Feng, Hao; Crow, Lowell; Jiang, Chenyang; Wang, Tianhao; Zhang,

Yuxuan; Bilheux, Hassina; Parnell, Steven R.; More Authors

DOI

10.1016/j.nima.2019.05.092

Publication date

2019

Document Version

Accepted author manuscript

Published in

Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors

and Associated Equipment

Citation (APA)

Dadisman, R., Shen, J., Feng, H., Crow, L., Jiang, C., Wang, T., Zhang, Y., Bilheux, H., Parnell, S. R., &

More Authors (2019). Design and characterization of zero magnetic field chambers for high efficiency

neutron polarization transport. Nuclear Instruments and Methods in Physics Research, Section A:

Accelerators, Spectrometers, Detectors and Associated Equipment, 940, 174-180.

https://doi.org/10.1016/j.nima.2019.05.092

Important note

To cite this publication, please use the final published version (if applicable).

Please check the document version above.

Copyright

Other than for strictly personal use, it is not permitted to download, forward or distribute the text or part of it, without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license such as Creative Commons. Takedown policy

Please contact us and provide details if you believe this document breaches copyrights. We will remove access to the work immediately and investigate your claim.

This work is downloaded from Delft University of Technology.

(2)

Please cite this article as: R. Dadisman, J. Shen, H. Feng et al., Design and characterization of zero magnetic field chambers for high efficiency neutron polarization transport, Nuclear Inst. and Methods in Physics Research, A (2019), https://doi.org/10.1016/j.nima.2019.05.092.

Contents lists available atScienceDirect

Nuclear Inst. and Methods in Physics Research, A

journal homepage:www.elsevier.com/locate/nima

Design and characterization of zero magnetic field chambers for high

efficiency neutron polarization transport

Ryan

Dadisman

a

,

Jiazhou

Shen

b

,

Hao

Feng

b

,

Lowell

Crow

a

,

Chenyang

Jiang

a

,

Tianhao

Wang

a

,

Yuxuan

Zhang

a

,

Hassina

Bilheux

a

,

Steven R.

Parnell

c

,

Roger

Pynn

a,b

,

Fankang

Li

a,∗

aNeutron Sciences Directorate, Oak Ridge National Laboratory, Oak Ridge, TN 37830, USA bCenter for Exploration of Energy and Matter, Indiana University, Bloomington, IN 47408, USA cFaculty of Applied Sciences, Delft University of Technology, Mekelweg 15, Delft, JB 2629, Netherlands

A R T I C L E

I N F O

Keywords:

Magnetic Wollaston prisms Larmor labeling

Large-angle neutron scattering Small-angle neutron scattering Zero magnetic field Polarization transport

A B S T R A C T

Several methods of polarized neutron scattering call for a zero magnetic field (ZF) region to reduce magnetic field integral aberrations while preserving the neutron polarization. Though the design for large angle neutron scattering has been presented in various places, the design characterization and tuning has not been discussed before. In this report, the tuning procedure will be discussed with both neutron polarization transport method and utilization of fluxgate magnetometers. As a tuning procedure, polarized neutrons are sensitive to any local field distortions along all trajectories within the beam, but the process is slow. With fluxgates, the entire beam region cannot be accessed simultaneously, but very fast and precise measurements can be made in accessible regions of interest. Consequently, we would like to benchmark the usage of fluxgates as a fast tuning probe compared with polarization measurements made with neutrons. Polarization transport results for tuned ZF chambers, up to 2.25 m in length, are presented.

1. Introduction 1

Methods which utilize the precession of the neutron spin in a well

2

designed magnetic field enable the investigations of material

struc-3

tures or dynamics with a resolution beyond the conventional neutron

4

scattering techniques. Mezei proposed the first technique of this type,

5

neutron spin echo (NSE) [1], which was further developed into neutron

6

resonance spin echo (NRSE) [2], modulated intensity with zero effort

7

(MIEZE) [3], spin echo small angle neutron scattering (SESANS) [4] and

8

spin echo modulated small angle neutron scattering (SEMSANS) [5].

9

NRSE uses radio frequency (RF) flippers to measure the excitations of

10

quasiparticles, for example, phonon or magnon and MIEZE only uses

11

one arm of NRSE to allow more relaxed sample environment. Similar to

12

the relation between NRSE and MIEZE, SESANS and SEMSANS employ

13

magnetic Wollaston prisms to measure the inter particle correlations of

14

materials in a extended length scale.

15

Since all of these methods label the change in either neutron energy

16

(E) or momentum transfer (Q) into a change in Larmor phase, i.e., the

17

accumulated angle of neutron spin precession, they may be referred as

18

neutron Larmor labeling techniques. To maximize the resolution of a

19

small change in either E or Q, it is critical to maximize the achievable

20

Larmor phase, 𝛷 ∝ 𝐹 𝐼 ×𝜆, where 𝐹 𝐼 is the magnetic field integral along

21

the neutron path and 𝜆 is the neutron wavelength. From this it follows

22

that maximizing the Larmor phase requires maximizing the magnetic

23

∗ Corresponding author.

E-mail address: fankangli@hotmail.com(F. Li).

field, path length of neutron, or neutron wavelength. To achieve a high 24

signal to noise ratio for a given setup, the Larmor phase aberrations 25

need to be minimized across the beam. To maximize the resolution 26

while maintaining a high signal to noise ratio, one approach is to intro- 27

duce the usage of zero magnetic field (ZF) chamber to separate the spin 28

manipulation components far apart from each other, where the neutron 29

polarization vector will be preserved when passing through this region 30

without accumulating additional Larmor phase. This is especially true 31

for the techniques involving resonance radio frequency flippers, such 32

as NRSE or MIEZE, where an effective precession is produced inside 33

the ZF chamber such that the accumulated phase is proportional to 34

the time neutrons spend between and the angular frequency of the 35

flippers. This is also true for methods using static magnetic fields, such 36

as SESANS or SEMSANS, where the gradient of the field integral inside 37

the spin manipulation devices matters for the Larmor phase. It is true 38

that people can also use well designed uniform magnetic guide field to 39

preserve the neutron polarization [6]. But careful shielding of this field 40

is critical at the magnetic field boundary where one does not need it, 41

for example, a 𝜋∕2 spin flipper. 42

Another situation where a ZF chamber is preferable originates from 43

the fact that a uniform Larmor phase across the sample is essential, 44

especially for large angle scattering on a sample with finite size, i.e., a 45

diffraction setup. Introducing ZF chambers in the regions where spin 46

https://doi.org/10.1016/j.nima.2019.05.092

Received 21 December 2018; Received in revised form 28 May 2019; Accepted 29 May 2019 Available online xxxx

0168-9002/© 2019 Published by Elsevier B.V.

(3)

NIMA: 62251

Please cite this article as: R. Dadisman, J. Shen, H. Feng et al., Design and characterization of zero magnetic field chambers for high efficiency neutron polarization transport, Nuclear Inst. and Methods in Physics Research, A (2019), https://doi.org/10.1016/j.nima.2019.05.092.

R. Dadisman, J. Shen, H. Feng et al. Nuclear Inst. and Methods in Physics Research, A xxx (xxxx) xxx

precession is not desired is one approach to achieve this, though care

1

must be taken to preserve the neutron polarization vector.

2

One of the recent developments in Larmor labeling methods is

3

the introduction of superconducting magnetic Wollaston prisms [7].

4

Such devices are composed of pairs of triangular shaped magnetic field

5

regions using superconducting coils. These adjacent regions of field are

6

opposite in direction and both transverse to the beam direction, and

7

by using superconducting films at the interfaces, the magnetic field is

8

very uniform and well defined within the regions. The angle between

9

the interface of the opposite fields and the beam direction introduces a

10

splitting of the spin states at the interface in a beam which is polarized

11

perpendicular to the fields. Wollaston prisms can be applied for large

12

angle neutron scattering, including neutron Larmor diffraction [8,9]

13

and inelastic neutron scattering spectroscopy [10]. In such applications,

14

a sample with finite size will lead to a variation of the neutron path

15

length introducing Larmor phase aberrations in a uniform magnetic

16

field. A ZF chamber with high efficiency polarization transport is

essen-17

tial in the sample region, i.e., between the precession devices, so that

18

the Larmor phase accumulation is controlled purely by the precession

19

devices. Such prisms have also been used to implement SEMSANS for

20

small angle scattering, in which a long separation between the spin

21

flippers yields longer spin echo lengths and the span should have ZF to

22

prevent additional phase accumulation. In this report, we discuss the

23

design and characterization of the ZF chamber for both large and small

24

angle applications, and discuss the benefits of using magnetometry as a

25

fast method of optimizing compensation coils for ZF chambers, which

26

could also be used for other techniques.

27

2. The ZF chamber for large angle neutron scattering technique 28

2.1. The design of the ZF chamber

29

The ZF chamber used in MuPAD [11], a spherical neutron

polarime-30

ter at the Paul Scherrer Institut (PSI) triple axis spectrometer, and

31

NRSE of FRM-II [12–14] consists of three components: a cylindrical

32

ZF chamber with a gap in the middle at the sample height; and two

33

rectangular mu metal boxes with beam port open ends to magnetically

34

shield the gaps between the central ZF chamber and the precession

35

devices. As shown in Fig. 1, the design presented here is similar to

36

those used for MuPAD and NRSE. Differently, the gap between the

37

top and bottom mu metal cylinder is closed by an inner concentric

38

mu metal cylinder. Two beam ports on the sides of the cylinder are

39

provided to allow the beam to pass through, as shown inFig. 1(a). The

40

inner cylinder is configured such that the two beam ports will remain

41

open when rotating the scattering ARM-II. The two mu metal boxes

42

surrounding the Wollaston prisms are then connected to the two beam

43

ports to optimize the shielding efficiency. As a rule of thumb, the size

44

of the ports needs to be optimized; a larger port will have a lower

45

shielding factor, while a smaller port will limit the beam size.

46

An open ended cylindrical shield is most effective at attenuating

47

the components of the magnetic field which are transverse to its

48

longitudinal axis. Consequently, the longitudinal component of the

49

ambient magnetic field, i.e., the component normal to the top and

50

bottom, will be the dominant stray field to enter the sample region,

51

as shown inFig. 1(a) and2. Rather than closing the open ends to fully

52

shield the sample environment, compensation coils placed on the top

53

and bottom provide more flexibility to cancel the stray fields entering

54

the region while allowing access for sample environment equipment,

55

e.g., a cryostat. For the two ports where the beam is passing through,

56

compensation coils are also used here to further reduce the stray fields

57

along the beam. These features are similar to what has previously been

58

used in MuPAD [11] and NRSE [12–14].

59

This concentric shell design was simulated using MagNet [15] finite

60

element analysis software. The zero-angle configuration was simulated,

61

i.e., the left and right ports are directly aligned and this will result in

62

a hole in the shield shown inFig. 1. This represents the worst possible

63

Fig. 1. (a) The schematic drawing of the setup used for demonstrating the ZF chamber:

the superconducting magnetic Wollaston prisms, mu metal shielding, and compensation coils are represented by green, red, and yellow lines, respectively. The magnetic field generated by the top and bottom compensation coils have been indicated by arrows. Schematic of the inner and outer mu cylinder. For clarity, the mu boxes surrounding prisms have been omitted. (c) Image of the physical setup. Mu metal boxes are used to shield the flux from getting though the two ports and the cylindrical ZF chamber is used to shield the stray field at the sample position. The diameter of the cylindrical ZF chamber is 30 cm. The size of the beam ports is 4 cm × 4 cm. The thickness of the mu metal is 1.5 mm. (For interpretation of the references to color in this figure legend, the reader is referred to the web version of this article.)

(4)

Please cite this article as: R. Dadisman, J. Shen, H. Feng et al., Design and characterization of zero magnetic field chambers for high efficiency neutron polarization transport, Nuclear Inst. and Methods in Physics Research, A (2019), https://doi.org/10.1016/j.nima.2019.05.092.

Fig. 2. MagNet simulation results for the ZF chamber, with the color indicating the magnetic field strength in Gauss and the arrows indicating the direction in the respective

planes. The neutron beam passes through the ports from left to right in each image. (a) the side view (b) the top view of the middle plane. The field is substantially attenuated, except at the openings which can be further reduced by compensation coils.

(5)

NIMA: 62251

Please cite this article as: R. Dadisman, J. Shen, H. Feng et al., Design and characterization of zero magnetic field chambers for high efficiency neutron polarization transport, Nuclear Inst. and Methods in Physics Research, A (2019), https://doi.org/10.1016/j.nima.2019.05.092.

R. Dadisman, J. Shen, H. Feng et al. Nuclear Inst. and Methods in Physics Research, A xxx (xxxx) xxx

configuration for the ZF chamber because at larger angles the gap

1

will be eliminated due to the rotation of the inner mu metal cylinder.

2

Exterior Helmholtz coils are used to simulate the expected ambient

3

background field, which is ∼10 Gauss at the sample region caused

4

by the magnetized goniometer of the beamline. Simulation results, as

5

given inFig. 2, shows that any magnetic flux that is transverse to the

6

ZF chamber will be shielded and the residual component parallel to

7

the openings can be canceled using compensation coils. As shown in

8

Fig. 2(b), the hole produced by the concentric shields not overlapping

9

at low angles is not compensated, but it has a minimal effect on the

10

magnetic field at the beam and sample locations. The overlapping

11

concentric shell design will further reduce the flux leakage through this

12

hole substantially, resulting in very good field attenuation along the

13

beam line.

14

2.2. The tuning and compensation coils of the ZF chamber

15

There are two methods we can use to tune the compensation coils:

16

using a fluxgate magnetometer to minimize the field readings; and

17

maximizing the polarization efficiency with a polarized neutron beam.

18

The reason to compare the two methods is due to the time required to

19

perform these two methods, which will be discussed later in this report.

20

Since there are four coils involved, a clear strategy is important, which

21

will be presented here. For both methods, the first step is always to

22

tune the top and bottom coils, after which their settings will be fixed

23

for the following tuning. It is important to tune the top and bottom coils

24

first since they will also contribute the most stray field to the central

25

sample region and will also produce stray fields inside the two ports.

26

For the two ports, there are three sources of stray field: from the top and

27

bottom compensation coils, shown inFig. 1(a); external ambient stray

28

field; and the neighboring superconducting Wollaston prisms. Please

29

note that the field here might be high (a couple of 𝐺) due to the flux

30

concentration of the surrounding mu metal. The first two contributions

31

are constant but the contribution from the prism scales with the current

32

within the prism’s coils. Consequently the compensation coils inside the

33

two ports need to be tuned for each prism current setting.

34

For this experiment, the fluxgate magnetometer tuning procedure is

35

simply to tune the current inside the coils such that the stray field is less

36

than 0.01G measured at a fixed spot inside each port, using a Bartington

37

Mag-01H magnetometer. Any stray magnetic fields inside the ports

38

will interact with the neutron polarization vectors thus leading to low

39

polarization transfer efficiency when neutrons are passing through the

40

whole setup inFig. 1. The tuning procedure using neutron polarization

41

in this case is to optimize the polarization efficiency by scanning the

42

compensation current inside the left and right coils, saving time by

43

using the top and bottom compensation coil settings found via

mag-44

netometry. Additionally, rather than measuring the full polarization

45

with two spin states measurements, only one spin state was used.

46

When the polarization transport is maximized, it would yield a global

47

maximum or minimum neutron count per unit time, depending on the

48

experimental setup.

49

The experiment was conducted on HB-2D beam line [16] at the

50

High Flux Isotope Reactor (HFIR) of Oak Ridge National Laboratory

51

using 4 Å neutrons. To measure the polarization transfer efficiency, the

52

apparatus was set up with the fields inside the two nutators aligned

53

parallel to the magnetic field vectors inside the prisms (𝜋 − 𝜋 mode).

54

In this situation, only the ZF chamber contributes to the variation

55

of the polarization transfer when we are tuning the compensation

56

coils. The compensation tuning with neutrons was performed with each

57

coil separately, in which the coil not being tuned was turned off. A

58

comparison of the compensation current inside the two coils found via

59

neutron measurement and magnetometry has been plotted inFig. 3for

60

different currents inside the prisms. Please note that the field inside the

61

prisms has been set such that they are all parallel to each other; the

62

field inside the ARM-I and ARM-II prisms are opposite to each other

63

with the same field intensity. Clearly, for both methods, the trend of

64

Fig. 3. The tuning curve of the left and right compensation coils on the two ports in

Fig. 1as a function of current inside the two prisms. The black and red lines denote the left and right compensation coils respectively. (For interpretation of the references to color in this figure legend, the reader is referred to the web version of this article.)

the tuning curve agrees with each other, though the optimum current 65

in the compensation coils obtained by neutron polarization is always 66

lower than that of the one by fluxgate. The reason is that the fluxgate, 67

inserted into the port from the ZF chamber side, is ∼ 2.5 cm away from 68

the prism due to the vacuum chamber surrounding it, which prevents 69

accurate minimization of the field just outside the prism. 70

To understand how the tuning will contribute to our actual neutron 71

measurements, the polarization efficiency is measured by setting the 72

nutator field perpendicular to the field inside the prisms (𝜋∕2 − 𝜋∕2 73

mode). In this mode, the neutron polarization vector will start to 74

precess around the magnetic field vectors inside the prisms. Therefore 75

the neutron will accumulate a Larmor phase inside ARM-I which will 76

be mostly canceled in ARM-II as in neutron spin echo [17], since they 77

always have the same magnitude but opposite directed magnetic field. 78

The polarization efficiency is obtained by sweeping the current inside 79

ARM-II about the current set in ARM-I, for both methods of tuning. 80

To demonstrate the necessity of optimizing the compensation coils for 81

each prism current setting, we performed an ‘‘off-tune’’ measurement 82

by fixing only the right compensation coil constant at the optimum for 83

2 A inside the neighboring prism, then tested it with different prism 84

settings. As shown in Fig. 4, the results given by either fluxgate or 85

neutrons are very close, so that we can use the fluxgate method as a 86

fast way of tuning the compensation coil in the two ports. However, 87

when the right compensation coil is off-tune for high prism current, 88

some depolarization occurs indicating that the compensation must be 89

exactly performed for each prism setting. It should be noted that the 90

measurement inFig. 4does not represent the worse situation, where 91

the left compensation coil is still optimized to each prism current, and 92

the prisms is only operating in a low current far below its routine 93

operational current (50 A). Additionally, a NRSE setup with a ZF tube 94

in each arm and a ZF chamber around the sample would have 6 port 95

compensation coils, so that even a small loss of polarization at each 96

port will compound to substantial depolarization. The necessary act of 97

re-tuning the compensation coils when the prism settings are changed 98

provides further support for the faster method using magnetometry. 99

2.3. Characterization of the ZF chamber with an off-tune compensation coil 100

So now the question becomes what happens if the compensation coil 101

is not completely on-tune and how much stray field can be tolerated. 102

To investigate this, the setup shown inFig. 1(a) was used with the field 103

of the first nutator aligned parallel to the prism field and the second 104

nutator aligned perpendicular to the prism field (𝜋 − 𝜋∕2 mode). In this 105

(6)

Please cite this article as: R. Dadisman, J. Shen, H. Feng et al., Design and characterization of zero magnetic field chambers for high efficiency neutron polarization transport, Nuclear Inst. and Methods in Physics Research, A (2019), https://doi.org/10.1016/j.nima.2019.05.092.

Fig. 4. The polarization efficiency, as a function of prism current, obtained with the

tuning current optimized using neutrons (red circle) and a fluxgate magnetometer (black square). The measurement with the right compensation coil off-tune (blue diamond) is presented for comparison. For these three measurements, the left compen-sation coil is always optimized for each prism current. The damping of the polarization with the compensation coil on tune (red circle and black square) is due to the residual magnetic field of the iron pole pieces, which will introduce inhomogeneous field integral across the beam. This can be fixed and improved by properly degaussing all the iron pole pieces inside the device.

Fig. 5. The oscillations obtained by scanning the current inside ARM-II with the nutator

fields orthogonal to each other, so that nominally no polarization should be transmitted. The observed increase in oscillation amplitude with increasing field inside the ZF chamber indicates more depolarization in the region. Data fitted to a simple sinusoid function to determine the polarization.

case, if the ZF chamber is perfect, we expect to observe no polarization 1

at the detector since there is no projection from a vertically orientated 2

polarization vector (defined by the first nutator) into a horizontal plane 3

(defined by the second nutator). In this way, we are able to measure 4

how much of the neutron polarization vector has been kicked out of 5

its principal axis into the horizontal plane, i.e., depolarized, due to 6

remnant ambient magnetic fields. 7

To measure the polarization efficiency of the principal precession 8

axis, the current inside of ARM-II was scanned, as shown inFig. 5, 9

with various settings of stray field inside the right port. As measured 10

by the fluxgate, when there is no stray field present, there are still 11

some measurable oscillations. Again, this means the polarization vector 12

of some of the neutrons is kicked out of its principal precession axis, 13

defined by the first nutator, when going through the ZF chamber. As 14

the stray field increases, the oscillations become clearer, which will 15

introduce more depolarization. But with a stray field of 0.3G shown 16

in the middle plot ofFig. 5, the polarization is almost the same as that 17

of the ZF (top plot ofFig. 5). So in subsequent experiments involving 18

the ZF chamber, it is routinely operated with a stray field of less than 19

0.1G. 20

3. The ZF chamber for low angle neutron scattering technique 21

The SEMSANS method employs two magnetic Wollaston prisms 22

separated by a ZF region, as discussed in [18,19], which generates 23

spatial intensity modulations on the detector, due to the gradient of 24

magnetic field integrals across the devices. The intensity modulation 25

can be used to measure the real-space density autocorrelation function 26

of samples and for dark field imaging. To achieve a high resolution, an 27

intensity modulation with a small period is always favorable, which can 28

be achieved, for example, by separating the two prisms far apart. This 29

(7)

NIMA: 62251

Please cite this article as: R. Dadisman, J. Shen, H. Feng et al., Design and characterization of zero magnetic field chambers for high efficiency neutron polarization transport, Nuclear Inst. and Methods in Physics Research, A (2019), https://doi.org/10.1016/j.nima.2019.05.092.

R. Dadisman, J. Shen, H. Feng et al. Nuclear Inst. and Methods in Physics Research, A xxx (xxxx) xxx

is where a ZF region is required to minimize unwanted phase

accumu-1

lation and to preserve the neutron polarization vectors by minimizing

2

the magnetic field along the path. We have also tried using two pairs

3

of Helmholtz coils between two prisms to transport polarization with

4

an additional magnetic field stepper in between to provide a field flip.

5

But it requires additional effort to properly shield or compensate the

6

stray field projected out by the Helmholtz coil.

7

For SEMSANS, the two Wollaston prisms are located at distances

8

𝐿1 and 𝐿2 from the detector, where 𝐿1 > 𝐿2, with the sample and

9

analyzer after the precession devices. As discussed in detail in [18,19],

10

to minimize the aberrations and maximize the contrast of the intensity

11

modulation on the detector due to the effect of beam divergence, the

12

magnetic fields in the prisms must satisfy,

13

𝐵1𝐿1= 𝐵2𝐿2 (1)

14

The spin echo length is proportional to the difference between the

15

magnetic fields, 𝛿𝑧∝ (𝐵2−𝐵1) = 𝐵2(1−𝐿2∕𝐿1). From this it is clear that

16

a long distance between the precession devices, i.e., a long ZF tube, will

17

allow for longer spin echo lengths.

18

3.1. The design of the ZF tube

19

Since the sample, where scattering occurs, is located after the

sec-20

ond prism, the ZF region can be provided by a straight, closed magnetic

21

shielding ZF tube. In this case, to achieve a higher shielding factor, a

22

two-layer mu metal cylindrical tube bolted to the windows of the mu

23

metal surrounding the prisms is capable of reducing the background

24

magnetic fields by several orders of magnitude. This is the general

25

design which was optimized for this case, the apparatus shown inFig. 7.

26

Analytical solutions for the optimum spacing of two ‘‘infinitely’’ long,

27

cylindrical magnetic shielding concentric shells exist and predict an

28

optimum ratio of the outer to inner radius of router≈

3 rinner [20]. The

29

expense and size of the magnetic shielding materials justifies a study

30

of the flatness of this optimization to determine a point of diminishing

31

returns, and so simulations were implemented using COMSOL [21] and

32

MagNet [15] to study the shielding as a function of the spacing between

33

the inner and outer tubes in the ‘‘short’’ and ‘‘infinitely long’’ limits. A

34

2D planar model, equivalent to the ‘‘infinitely’’ long limit, was made

35

in COMSOL, and agrees with the analytical model. A 3D model of a

36

0.5 m long shield was made using MagNet, and indicated the optimum

37

was found at a slightly higher ratio than the analytical solution for an

38

external field transverse to the longitudinal axis of the cylinders. Aside

39

from the different geometries, each simulation software also utilized

40

slightly different B–H curves to determine the permeability of mu

41

metal, which changes the overall field attenuation. Collectively, these

42

demonstrate that, for the inner diameter ZF tube under consideration,

43

the optimization does not change dramatically for a very long or very

44

short ZF tube.

45

The first datum on Fig. 6 is for the case where the outer layer

46

touches the inner layer. As the gap increases, there is a very sharp

in-47

crease in the magnetic field attenuation. With router∕rinner= 1.25, 99.7%

48

of the maximum attenuation gain with router∕rinner =

3is achieved,

49

using an outer layer which is 28% smaller. From the perspective of

50

portability and cost, the ratio router∕rinner = 1.25 is a value point for an

51

effective shield. As in the case of Section2, the open ends of the ZF

52

tube require compensation coils to minimize the stray axial field and

53

the fringe fields from the prisms.

54

3.2. Polarization transport measurements of the ZF tube in SEMSANS

55

The experiment was conducted at beamline CG-1D [22] at the HFIR.

56

2 Å neutrons were polarized using a V-cavity supermirror upstream

57

of the Wollaston prisms. As shown inFig. 7, instead of using a single

58

prism as an individual unit, a dual prism was used, which is equivalent

59

to a single prism with current doubled due to the doubling of the

60

magnetic field integral [7]. A 2.25 m long ZF tube connected the two

61

Fig. 6. Comparison of COMSOL and MagNet simulation results for magnetic field

attenuation of a two-layer ZF tube demonstrates minimal improvement for router∕rinner>

1.25.

prisms, and is constructed from 2-layers of 1.5 mm thick mu metal, 62

radially separated by 10 mm and with an inner diameter of 75 mm. This 63

corresponds to router∕rinner ≈1.27, near the ratio at which simulation 64

indicates further attenuation gains are minimal. The tube flanges were 65

directly bolted to the mu metal shielding of the prisms, as shown in 66

Fig. 7. The flange overlaps with the prism shielding to reduce the 67

magnetic field penetration into the otherwise open ends of the tubes, 68

and a 20-turn compensation coil was wound on each end to further 69

compensate the fringe fields. Downstream of the second prism was 70

a 3He spin analyzer and CCD detector, where are not shown in the 71

picture. 72

In this case, the interior of the ZF tube cannot be accessed with a 73

fluxgate, therefore, polarized neutrons are used as a probe to tune the 74

compensation coils, as described before. For future routine operation, 75

it would be advantageous to have fluxgates built into the design, which 76

would save several hours from the compensation coil tuning procedure. 77

However, as the Wollaston prisms are designed with flexibility to be 78

used in a variety of setups, they do not have this built-in feature. The 79

two arms were set to the same magnetic field and the two nutators 80

aligned to produce a field parallel to the field inside the two arms. The 81

left compensation coil was tuned by measuring the neutron polarization 82

efficiency with the right coil turned off; the right coil was tuned by the 83

same procedure, but with the left coil set at its optimum setting.Fig. 8 84

shows the optimization results for the left and right compensation coils. 85

An interesting feature is that one of the compensation coils affects the 86

polarization less than the other. There are two reasons for this: the right 87

coil was tuned with the left coil off-tune, then the left coil was tuned 88

with the right coil optimized; and there is a current imbalance inherent 89

in the SEMSANS method as shown in Eq.(1), such that the field inside 90

the right prism and thus the stray field requiring compensation is larger, 91

a factor of three in this setup. At the optimized current settings 𝐼1 = 92

0.4 A and 𝐼2= −3.5 A, the measured polarization was 79.4% without 93

optimizing the experimental setup. For comparison, a measurement of 94

the beam polarization using guide fields, rather than the ZF tube and 95

Wollaston prisms apparatus, yielded a polarization efficiency of 89%. 96

There is only 10% loss in polarization, which is acceptable considering 97

a ZF tube of 2.25 m. 98

It is necessary to also determine whether any Larmor phase aber- 99

rations are generated through the ZF tube by measuring the spatial 100

modulations of the polarization at the detector. After the compensation 101

coils were tuned, the apparatus was set up for SEMSANS by setting the 102

nutator fields perpendicular to the field inside the prisms (𝜋∕2 − 𝜋∕2 103

mode). To maximize the contrast of the intensity modulations on the 104

detector, the current inside the two arms were tuned to achieve 𝐼1𝐿1= 105

𝐼2𝐿2, as in Eq.(1). Misalignment of prisms, guide fields, and detectors 106

(8)

Please cite this article as: R. Dadisman, J. Shen, H. Feng et al., Design and characterization of zero magnetic field chambers for high efficiency neutron polarization transport, Nuclear Inst. and Methods in Physics Research, A (2019), https://doi.org/10.1016/j.nima.2019.05.092.

Fig. 7. Schematic and image of the SEMSANS setup at the CG-1D beamline. The

superconducting magnetic Wollaston prisms, mu metal shielding, and compensation coils are represented by green, red, and yellow lines, respectively.

Fig. 8. The tuning curve of the left and right compensation coils by measuring the

polarization transfer efficiency through the SEMSANS apparatus.

Fig. 9. Polarization fringes result using the SEMSANS setup demonstrates high

polarization transport efficiency through the ZF tube.

can cause a tilt of the modulation relative to the camera’s pixel axes 1

and is corrected with a pixel shifting algorithm. The background is 2

subtracted from the raw images, and the polarization calculated per 3

pixel yielding a polarization map as shown inFig. 9. Good uniformity 4

is observed along the vertical axis of the polarization, indicating that 5

minimal phase aberrations are generated inside the ZF tube. Using 2 Å 6

neutrons, we measured a spatial period of 303 μm for prism currents 7

𝐼1 = 14.59 A and 𝐼2 = 4.59 A. Scaling this to the highest allowed 8

current for the device (50 A), this same setup can achieve 88 μm. With 9

cold neutrons, the fringe period can be further reduced. 10

4. Summary 11

We have presented the design of ZF chambers and discussed the 12

operation and characterization of such chambers in combination with 13

superconducting Wollaston prisms applied to both large angle and 14

small angle polarized neutron scattering. Such techniques have the 15

potential to be utilized for other polarized neutron scattering methods. 16

We have demonstrated that a well designed ZF chamber is capable of 17

achieving high efficiency polarization transport over long distances, 18

2.25 m in this work, which is important for high resolution neutron 19

Larmor labeling techniques. For an apparatus in which the magnetic 20

field flux is not well confined or the boundary not well defined, a ZF 21

chamber might not be the best option [23]. Instead a homogeneous 22

(9)

NIMA: 62251

Please cite this article as: R. Dadisman, J. Shen, H. Feng et al., Design and characterization of zero magnetic field chambers for high efficiency neutron polarization transport, Nuclear Inst. and Methods in Physics Research, A (2019), https://doi.org/10.1016/j.nima.2019.05.092.

R. Dadisman, J. Shen, H. Feng et al. Nuclear Inst. and Methods in Physics Research, A xxx (xxxx) xxx

guide field can be used to maintain polarization, with a flipper placed

1

in the center such that phase aberrations caused by the guide field

2

inhomogeneity may be minimized [6].

3

We have also presented a comparison of the compensation coil

4

tuning methods for ZF chambers, using magnetometry and neutron

5

polarization. Both methods are shown to produce equivalent

polar-6

ization transport results, which shows the effectiveness of the using

7

fluxgate magnetometer as the tuning tool. Between the two methods,

8

the amount of time required to perform the tuning is the key difference

9

between fluxgate and neutron polarization methods. Without using

10

complicated automated system, it takes less than 0.5 min to optimize

11

the compensation coils for a ZF chamber using the fluxgate

magnetome-12

ter for each given prism current. For a routinely operated instrument,

13

this whole procedure can be automated with for example a proportional

14

integral derivative (PID) control system. Therefore whenever a

config-15

uration change is made, the system can automatically compensate the

16

stray field of all ports simultaneously. Such a design would require

17

the magnetometers to avoid the beam area. For comparison, using

18

polarized neutrons, the measurements would take several minutes per

19

each compensation coil setting, depending on the beam intensity and

20

detector efficiency. Potentially it would take hours to complete the

21

compensation procedure for all the current settings inside the prisms.

22

And this procedure can only performed one by one, which will scale

23

up the time required for multiple ZF ports. This method is also strongly

24

dependent on the initial current used in each coil, such that the actual

25

optimum may not even be easily found using neutron polarization

26

measurements alone. So, the fluxgate magnetometer provides us with

27

another quick but effective way of tuning the ZF chamber, by which

28

the whole experiment can be much simplified.

29

Acknowledgments 30

This research used resources at the High Flux Isotope Reactor,

31

a DOE Office of Science User Facility operated by the Oak Ridge

32

National Laboratory. We would like to thank Jack Doskow with Indiana

33

University for CAD model contributions, Thomas Keller for supporting

34

the design of the ZF chamber, Matthew Pearson with ORNL for software

35

development and assistance setting up power supplies, and Tianhao

36

Wang with ORNL for providing polarization results for the V-cavity and

37

3He analyzer performance on beamline CG-1D.

38

Notice: This manuscript has been authored by UT-Battelle, LLC,

39

USA, under contract DE-AC05-00OR22725 with the US Department

40

of Energy (DOE). The US government retains and the publisher, by

41

accepting the article for publication, acknowledges that the US

govern-42

ment retains a nonexclusive, paid-up, irrevocable, worldwide license to

43

publish or reproduce the published form of this manuscript, or allow

44

others to do so, for US government purposes. DOE will provide public

45

access to these results of federally sponsored research in accordance

46

with the DOE Public Access Plan (

http://energy.gov/downloads/doe-47

public-access-plan).

48

References 49

[1] F. Mezei, Neutron spin echo: A new concept in polarized thermal neutron 50

techniques, Z. Phys. A 255 (2) (1972) 146–160, http://dx.doi.org/10.1007/ 51

BF01394523. 52

[2] R. Golub, R. Gähler, A neutron resonance spin echo spectrometer for quasi-elastic 53

and inelastic scattering, Phys. Lett. A 123 (1) (1987) 43–48,http://dx.doi.org/ 54

10.1016/0375-9601(87)90760-2. 55

[3] R. Gähler, R. Golub, T. Keller, Neutron resonance spin echo—a new tool for high 56

resolution spectroscopy, Physica B 180–181 (1992) 899–902,http://dx.doi.org/ 57

10.1016/0921-4526(92)90503-K. 58

[4] M. Rekveldt, Novel SANS instrument using neutron spin echo, Nucl. Instrum. 59

Methods Phys. Res. B 114 (3) (1996) 366–370, http://dx.doi.org/10.1016/0168-60

583X(96)00213-3. 61

[5] W.G. Bouwman, C.P. Duif, J. Plomp, A. Wiedenmann, R. Gähler, Combined 62

SANS–sesans, from 1nm to 0.1mm in one instrument, in: Proceedings of 63

the 8th International Workshop on Polarised Neutrons for Condensed Matter 64

Investigation, Physica B 406 (12) (2011) 2357–2360,http://dx.doi.org/10.1016/ 65

j.physb.2010.11.069. 66

[6] O. Uca, Spin-echo small-angle neutron scattering development, Ph.D. thesis, Delft 67

University of Technology, 2003. 68

[7] F. Li, S.R. Parnell, W.A. Hamilton, B.B. Maranville, T. Wang, R. Semerad, D.V. 69 Baxter, J.T. Cremer, R. Pynn, Superconducting magnetic wollaston prism for 70 neutron spin encoding, Rev. Sci. Instrum. 85 (5) (2014) 053303,http://dx.doi. 71

org/10.1063/1.4875984. 72

[8] F. Li, H. Feng, A.N. Thaler, S.R. Parnell, W.A. Hamilton, L. Crow, W. Yang, 73 A.B. Jones, H. Bai, M. Matsuda, D.V. Baxter, T. Keller, J.A. Fernandez-Baca, R. 74 Pynn, High resolution neutron larmor diffraction using superconducting magnetic 75

wollaston prisms, Sci. Rep. 7 (1) (2017) 865. 76

[9] F. Li, H. Feng, A.N. Thaler, S.R. Parnell, L. Crow, M. Matsuda, F. Ye, T. Kimura, 77 J.A. Fernandez-Baca, R. Pynn, New capabilities in high-resolution neutron larmor 78 diffraction at ORNL, J. Appl. Crystallogr. 51 (3) (2018) 584–590. 79 [10] F. Li, R. Pynn, A novel neutron spin echo technique for measuring phonon 80 linewidths using magnetic wollaston prisms, J. Appl. Crystallogr. 47 (6) (2014) 81

1849–1854. 82

[11] M. Janoschek, S. Klimko, R. Gähler, B. Roessli, P. Böni, Spherical neutron 83

polarimetry with mupad, Physica B 397 (1) (2007) 125–130. 84

[12] T. Keller, B. Keimer, K. Habicht, R. Golub, F. Mezei, Neutron resonance spin 85 echo-triple axis spectrometry (NRSE-TAS), in: Neutron Spin Echo Spectroscopy, 86 in: Lecture Notes in Physics, Vol. 601, Springer Berlin Heidelberg, 2003, pp. 87

74–86. 88

[13] S. Klimko, Zeta, a zero field spin echo method for very high resolution study of 89 elementary excitations and first application, Ph.D. thesis, Technische Universität 90

Berlin, 2003. 91

[14] M. Duc Le, M. Skoulatos, D.L.a. Quintero-Castro, R. Toft-Petersen, F. Groitl, K.C. 92 Rule, K. Habicht, The upgraded cold neutron three-axis spectrometer FLEXX at 93

BER II at HZB, Neutron News 25 (2) (2014) 19–22. 94

[15] MagNet©, Infolytica Corp. 95

[16] L. Crow, W.A. Hamilton, J.K. Zhao, J.L. Robertson, The HB-2d polarized neutron 96 development beamline at the high flux isotope reactor, J. Phys. Conf. Ser. 746 97

(1) (2016) 012010. 98

(10)

Please cite this article as: R. Dadisman, J. Shen, H. Feng et al., Design and characterization of zero magnetic field chambers for high efficiency neutron polarization transport, Nuclear Inst. and Methods in Physics Research, A (2019), https://doi.org/10.1016/j.nima.2019.05.092.

[17] F. Mezei, Neutron spin echo and polarized neutrons, in: Inelastic Neutron 1

Scattering, IAEA, Vienna, 1978, p. 125. 2

[18] F. Li, S.R. Parnell, H. Bai, W. Yang, W.A. Hamilton, B.B. Maranville, R. Ashkar, 3

D.V. Baxter, J.T. Cremer, R. Pynn, Spin echo modulated small-angle neutron 4

scattering using superconducting magnetic wollaston prisms, J. Appl. Crystallogr. 5

49 (2016) 55–63,http://dx.doi.org/10.1107/S1600576715021573. 6

[19] F. Li, S.R. Parnell, R. Dalgliesh, A. Washington, J. Plomp, R. Pynn, Data 7

correction of intensity modulated small angle scattering, Scientific Reports 9 (1) 8

(2019)http://dx.doi.org/10.1038/s41598-019-44493-9. 9

[20] V.V. Yashchuk, S.-K. Lee, E. Paperno, Magnetic shielding, in: Optical Magne-10

tometry, Cambridge University Press, 2013, pp. 225–248,http://dx.doi.org/10. 11

1017/CBO9780511846380.013. 12

[21] COMSOL Multiphysics©, version 5.4, COMSOL Inc., URLwww.comsol.com. 13

[22] L. Santodonato, H. Bilheux, B. Bailey, J. Bilheux, P. Nguyen, A. Tremsin, 14

D. Selby, L. Walker, The CG-1D neutron imaging beamline at the oak ridge 15

national laboratory high flux isotope reactor, in: Proceedings of the 10th 16

World Conference on Neutron Radiography (WCNR-10) Grindelwald, Switzerland 17

October 5–10, 2014, Physics Procedia 69 (2015) 104–108,http://dx.doi.org/10. 18

1016/j.phpro.2015.07.015. 19

[23] J. Plomp, Spin-echo development for a time-of-flight neutron reflectometer, Ph.D. 20

thesis, Delft University of Technology, 2009. 21

Cytaty

Powiązane dokumenty

Nie lekceważąc motywacji o charakterze politycznym i prestiżowym, należy w tym tekście uwypuklić szczególną rolę poufnych, tajnych narad i spotkań, w końcu też „wymiany

invoked from Ruby scripts, analyze Ruby source code, locate grid object classes, their instances and operations, check operation arguments and then, using these data, find

Humans react to lan- guage in both these senses as to an aesthetic and cognitive value” (Markowski, Puzynina, 2001: 54). If this statement was true, each language and each of

Grande Dame Guignol, Gothic fiction, Savage Intruder, Hollywood, dystopia https://orcid.org/0000-0002-5514-1287... * Department of Canadian, Intermedial and Postcolonial

— zogniskowane wywiady grupowe (FGI), w których uczestniczyli miesz- kańcy Wrocławia zaangażowani w tworzenie ewaluowanych instalacji bądź biorący udział w otwarciach

Kontekst sprawcy lub biorcy tworzy taką różną rzeczywistość, podobnie jak kontekst sytuacyjny lub kontekst wcześniejszego doświadczenia kieruje percep- cją

Using a simple optical cavity consisting of a gold mirror and a SiO 2 layer, the absorption ef ficiency can be boosted to over 97%: this result is confirmed experimentally by

Latawiec, Pracownia Konserwacji Zbiorów Specjalnych, w: Zbiory specjalne Biblioteki Uniwer- syteckiej Katolickiego Uniwersytetu Lubelskiego, red... towskiego 39 , wówczas